Keysight Technologies (Agilent HP) E4407B vs Keysight Technologies (Agilent HP) 8563E

This technical comparison reviews the Keysight (Agilent HP) E4407B and the Keysight (Agilent HP) 8563E spectrum analyzers. Both instruments support measurement capabilities up to 26.5 GHz, but they serve different roles in performance and field ruggedness. The E4407B belongs to the ESA-E Series as a mid-performance benchtop unit, while the 8563E is a high-performance portable analyzer built for harsher environments. Engineers can use this analysis of their shared and distinct characteristics to choose the ideal instrument for their RF and microwave testing requirements.

Standard Frequency Range
9 kHz to 26.5 GHz (Both models)
Phase Noise @ 10 kHz Offset
E4407B: ≤ -90 dBc/Hz | 8563E: ≤ -113 dBc/Hz
Operating Temperature
E4407B: 0 to 55 °C | 8563E: -10 to +55 °C
RF Input Connector
APC-3.5 mm male (Both models)
Keysight Technologies (Agilent HP) E4407BKeysight Technologies (Agilent HP) 8563E

Specifications

Specification
Keysight Technologies (Agilent HP)E4407B
Keysight Technologies (Agilent HP)8563E
General
Calibration Cycle1 year—
EMC/EMI ComplianceCISPR Pub 11 Group 1 Class A—
Interfaces—GPIB (IEEE-488)
Display—CRT
Standard Measurements—Adjacent channel power, channel power, carrier power, occupied bandwidth, time-gated measurements
Warm-up Time—5 minutes (for frequency reference stability)
Physical
Power Supply90 to 132 V (47-440 Hz) / 195 to 250 V (47-66 Hz); < 300 W—
Operating Temperature(°C)0 to 55-10 to +55
Dimensions (W×H×D)(mm)416 × 222 × 409—
Weight(kg)17.1—
Frequency Specifications
Frequency Range9 kHz to 26.5 GHz (100 Hz to 26.5 GHz with Option UKB)—
Phase Noise @ 10 kHz Offset≤ -90 dBc/Hz (1 GHz carrier)—
Sweep Time Range1 ms to 4000 s (Span > 0 Hz), 50 µs to 4000 s (Zero Span)—
Frequency Range (Standard)—9 kHz to 26.5 GHz
Frequency Range (Option 006)—30 Hz to 26.5 GHz
Preselected Frequency Range—2.75 GHz to 26.5 GHz
Frequency Band Breakdown—0 to 2.9 GHz, 2.75 to 6.46 GHz, 5.86 to 13.0 GHz, 12.4 to 19.7 GHz, 19.1 to 26.5 GHz
Frequency Reference Aging Rate—±1 × 10⁻⁷ / year
Frequency Reference Temperature Stability—±1 × 10⁻⁸ (-10 to +55 °C)
Span Range—0 Hz (Zero Span), 100 Hz to 26.5 GHz
Sweep Time Range (Zero Span)—50 µs to 6000 s
Sweep Time Range (Span > 0 Hz)—50 ms to 100 s
Sweep Time Accuracy—±10%
Resolution Bandwidth (RBW) Range—1 Hz to 2 MHz (1, 3, 10 sequence)
Digital Resolution Bandwidths—1, 3, 10, 30, 100 Hz
Video Bandwidth (VBW) Range—1 Hz to 3 MHz (1, 3, 10 sequence)
Frequency Counter Resolution—1 Hz
Amplitude Specifications
Absolute Amplitude Accuracy±0.4 dB overall—
Maximum Safe Continuous CW Power+30 dBm (1 W)—
Input Attenuator Range0 to 65 dB in 5 dB steps0 to 70 dB
Amplitude Measurement Range—DANL to +30 dBm
Maximum Safe Input (Continuous Power)—+30 dBm (1 W)
Maximum Safe Input (DC Voltage)—0 Vdc
Displayed Average Noise Level (DANL)—≤ -149 dBm (10 MHz to 2.9 GHz, 1 Hz RBW)
1 dB Gain Compression Point—-5 dBm (10 MHz to 2.9 GHz)
Display Range (Log Scale)—10 to 100 dB (1 to 10 dB/div)
Display Range (Linear Scale)—10 divisions
Reference Level Range—-119.9 to +30 dBm
Input Attenuator Step Size—10 dB
Inputs and Outputs
Front Panel RF Input ConnectorAPC-3.5 mm (male)—
RF Input Impedance50 Ω50 Ω (nominal)
RF Input Connector—APC-3.5 mm (male)
External Reference Input—10 MHz
Sweep Output—0 to +10 V (nominal)
Video Output—0 to +1 V (nominal)
General and Environmental
Warm-up Time5 minutes to guaranteed measurement accuracy—
Storage Temperature Range-40 to +75 °C—
Frequency
Span Range0 Hz (zero span), 100 Hz to 26.5 GHz—
Timebase Aging Rate (Standard)± 2 × 10^-6 / year—
Timebase Aging Rate (Option 1D5)± 1 × 10^-7 / year—
Inputs & Outputs
Timebase Reference Input/Output10 MHz—
Calibrator Signal—300 MHz, -10 dBm
Calibration Output Connector—BNC (female)
Probe Power—+15 Vdc, -12.6 Vdc (150 mA max)
10 MHz Reference Output—> 0 dBm into 50 Ω, BNC (female)
External Trigger Input—TTL levels, rear panel BNC (female)
Earphone Output—1/8 inch (3.5 mm) front-panel jack
Amplitude
Amplitude Reference Signal50 MHz, -20 dBm—
Reference Level Range-149.9 dBm to +30 dBm—
Amplitude UnitsdBm, dBmV, dBµV, V, WdBm, dBmV, dBµV, Volts, Watts
Standard DetectorsNormal, positive peak, sample, negative peak—
Log Scale Fidelity± 0.3 dB (0 to -60 dB from reference level)—
Residual Responses<-90 dBm (0 dB attenuation, 50 Ω load)—
Amplitude Detectors—Normal (Rosenfell), Positive Peak, Negative Peak, Sample
Options & Enhancements
Tracking Generator Frequency (Option 1DN)9 kHz to 3.0 GHz—
Frequency Performance
Phase Noise @ 100 kHz Offset≤ -113 dBc/Hz (typical ≤ -117 dBc/Hz)—
Phase Noise @ 1 MHz Offset≤ -119 dBc/Hz (typical ≤ -127 dBc/Hz)—
Sweep Specifications
Sweep Points401—
Options & Configurations
Time-Gated Spectrum AnalysisAvailable via Option 1D6—
DSP and Fast ADCAvailable via Option B7D—
Dynamic Range and Spectral Purity
Phase Noise (at 10 kHz offset)—≤ -113 dBc/Hz (at 1 GHz)
Phase Noise (at 30 kHz offset)—≤ -113 dBc/Hz (at 1 GHz)
Third Order Intercept (TOI)—> +15 dBm (10 MHz to 2.9 GHz)
Residual Responses—≤ -90 dBm (≥ 200 kHz, 0 dB attenuation)
Environmental and Compliance
Shock and Vibration Profile—MIL-T-28800 rugged
Sweep & Trigger
Sweep Trigger Modes—Free run, line, video, external
Features
Audio Demodulation—AM / FM (with built-in speaker and headphone jack)
Amplitude & Resolution
RBW Selectivity (Analog Filters)—< 15:1 (60 dB / 3 dB ratio for 300 Hz to 2 MHz)
RBW Selectivity (Digital Filters)—< 5:1 (60 dB / 3 dB ratio for 1 Hz to 100 Hz)
Log Scale Divisions—1, 2, 5, and 10 dB/div
Options
Option 001—Second IF Output (310.7 MHz), rear panel BNC
Environmental & Physical
Environmental Profile—Meets MIL-T-28800 Type III, Class 3, Style C
Compliance
EMI / RFI Compliance—Meets CISPR 11 and MIL-STD-461C

RF Performance and Phase Noise

The main performance differentiator between these two instruments is their phase noise capability. The Keysight 8563E provides high-performance tuning with a phase noise specification of ≤ -113 dBc/Hz at a 10 kHz offset from a 1 GHz carrier. In comparison, the E4407B is rated at ≤ -90 dBc/Hz under the same offset conditions.

Amplitude limits and reference levels also vary. The E4407B supports a wider reference level range of -149.9 dBm to +30 dBm, whereas the 8563E offers a range of -119.9 to +30 dBm. Both instruments handle a maximum safe continuous CW input power of +30 dBm (1 W).

Frequency Coverage and Resolution Bandwidth

While both units default to a standard frequency range of 9 kHz to 26.5 GHz, they address low-frequency options differently. The 8563E can be configured with Option 006 to extend its lower range down to 30 Hz. The E4407B requires Option UKB to reach down to 100 Hz. Both units offer zero span up to a full 26.5 GHz span range.

For resolution bandwidth, the 8563E features an analog filter range from 1 Hz to 2 MHz and digital resolution bandwidths down to 1 Hz, delivering tight selectivity for resolving closely spaced signals.

Attenuator Control and Mechanical Design

The E4407B features an input attenuator range of 0 to 65 dB adjustable in fine 5 dB steps. The 8563E offers a slightly broader attenuation range of 0 to 70 dB with a step size of 10 dB.

Regarding environmental ruggedness, the 8563E is designed to meet MIL-T-28800 Type III, Class 3, Style C specifications. This standard gives the 8563E a broader operating temperature window of -10 to +55 °C, compared to the 0 to 55 °C limit of the E4407B. Additionally, the E4407B is physically larger and weighs 17.1 kg.

Frequently asked questions

Which analyzer is better suited for harsh field environments?
The Keysight 8563E is designed for harsh conditions, built to MIL-T-28800 specifications, and operates in lower temperatures down to -10 °C.
Can either instrument generate tracking signals?
The E4407B can be equipped with Option 1DN to provide a tracking generator frequency range from 9 kHz to 3.0 GHz.