Fluke 5500A vs Fluke 5520A

This comparison evaluates the Fluke 5500A and the Fluke 5520A multi-product calibrators. Designed for calibration laboratories and test benches, both instruments serve as centralized standards for electrical instrument verification. They share a similar design chassis, identical external dimensions, and matching primary connectivity. However, the Fluke 5520A expands upon the output capabilities of the Fluke 5500A, offering wider current, resistance, and capacitance ranges, alongside higher maximum oscilloscope calibration bandwidths and dedicated options for power quality and pressure modules.

Warm-up Time
30 minutes
Dimensions (W×H×D)
432 × 178 × 473 mm
Operating Temperature Range
0 to 50 °C
Phase Resolution
0.01°
Fluke 5500AFluke 5520A

Specifications

Specification
Fluke5500A
Fluke5520A
General
InterfacesRS-232; IEEE-488 (GPIB)GPIB (IEEE-488), RS-232
Physical
Power Supply100 / 120 / 220 / 240 V AC ±10%, 47 to 63 Hz—
Operating Temperature(°C)0 to +50 °C0 °C to 50 °C
Dimensions (W×H×D)(mm)432 × 178 × 473 mm432 mm × 178 mm × 473 mm
Weight(kg)20 kg—
General Specifications
Frequency Range0.01 Hz to 2.0 MHz—
Output CapabilitiesDC/AC voltage, DC/AC current, resistance, capacitance, simulated power (phase control), thermocouples, RTDs—
Waveform GenerationMultiple waveforms and harmonics—
DC Voltage Specifications
Output Ranges0 to ±1020 V—
Temperature Calibration
Supported TypesThermocouples and RTDs—
Power and Energy (Simulated)
DC/AC Power CapabilitiesDual DC or AC outputs (two voltages, or voltage + current) with phase control—
DC Power Range109 µW to 11 kW—
AC Power Range109 µW to 11 kW—
Oscilloscope Calibration Options
Leveled Sine Wave BandwidthUp to 600 MHz (with Oscilloscope Calibration option)—
Supported SignalsLeveled sine wave, fast edge, time mark, amplitude signals—
Safety and Compliance
Calibration TraceabilityTraceable to nationally recognized standards (certificate included)—
Environmental
Storage Temperature-20 to +70 °C—
Non-Operating AltitudeUp to 12,200 m (40,000 ft)—
Shock and VibrationMIL-T-28800E, Type III, Class 5—
Performance
Warm-up Time30 minutes (to meet full specifications)—
Frequency Output Accuracy—± 2.5 ppm
Performance Specifications
DC Voltage Output Range0 to ±1020 V—
AC Voltage Output Range1 mV to 1020 V—
DC Current Output Range0 to ±11 A—
AC Current Output Range29 µA to 11 A—
Resistance Output Range0 to 329.999 MΩ—
Capacitance Output Range330 pF to 1.0999 mF—
Phase Range0 to ±179.99°—
Phase Resolution0.01°—
Options & Upgrades
Oscilloscope Calibration300 MHz (Option SC300) or 600 MHz (Option SC600)—
Power
Maximum Power Consumption600 VA—
Power Consumption—600 VA
Line Voltage Settings—100 V / 120 V / 220 V / 240 V (± 10%)
Line Frequency Range—47 Hz to 63 Hz
Compliance
Safety StandardsIEC 1010-1 (1990), ANSI/ISA S82.01-1994, CAN/CSA-C22.2 No. 1010.1-92—
DC Voltage Function
DC Voltage Range Limits—0 to ±1020 V
DC Voltage Best 1-Year Uncertainty—±11 ppm of setting
DC Current Function
DC Current Range Limits—0 to ±20.5 A
AC Voltage Function
AC Voltage Range Limits—1 mV to 1020 V
AC Voltage Frequency Range—10 Hz to 500 kHz
AC Voltage Waveforms—Sine, square, triangle, truncated sine
AC Current Function
AC Current Range Limits—29 µA to 20.5 A
AC Current Frequency Range—10 Hz to 30 kHz
Resistance Function
Resistance Range—0 to 1100 MΩ
Resistance Output Mode—Continuously variable
Capacitance Function
Capacitance Range—0.19 nF to 110 mF
Capacitance Output Mode—Continuously variable
Power and Phase Function
Phase Resolution—0.01°
Dual Output Capabilities—Two simultaneous voltage outputs or voltage and current
Frequency and Duty Cycle
Frequency Output Range—0.01 Hz to 2.0 MHz
Oscilloscope Calibration Option
Oscilloscope Calibration Bandwidth Options—Up to 300 MHz, 600 MHz, or 1.1 GHz (by option)
Power Quality Option
Power Quality Option Model—5520A-PQ
Pressure and Humidity Measurement
Pressure Module Compatibility—Fluke 700 Series
General & Environmental Specifications
Calibration Software Compatibility—MET/CAL Plus
Warm-up Time—30 minutes
Recommended Calibration Interval—1 year (12 months)
Storage Temperature Range—-20 °C to 70 °C
Operating Altitude Limit—3,050 m (10,000 ft)
Safety Compliance—Designed to comply with IEC 1010-1
Temperature
Thermocouple Simulation Resolution—0.01 °C / 0.01 °F
RTD Simulation Resolution—0.01 °C / 0.01 °F
Safety & Isolation
Maximum Guard-to-Earth Voltage—20 V peak

Electrical Sourcing Differences

While both calibrators cover core electrical parameters, their sourcing ranges vary significantly.

  • Voltage: Both systems share a DC and AC voltage limit of 0 to ±1020 V (starting from 1 mV for AC voltage).
  • Current: The Fluke 5500A delivers a DC current range of 0 to ±11 A and an AC current range of 29 µA to 11 A. The Fluke 5520A expands this range, providing up to 20.5 A for both DC and AC outputs.
  • Resistance: The Fluke 5500A sources resistance from 0 to 329.999 MΩ. The Fluke 5520A provides a wider continuously variable range of 0 to 1100 MΩ.
  • Capacitance: The Fluke 5500A has a capacitance range of 330 pF to 1.0999 mF, while the Fluke 5520A operates over a continuously variable range from 0.19 nF to 110 mF.

Oscilloscope and Functional Calibration Options

Both instruments offer optional internal hardware packages for oscilloscope calibration, but they target different bandwidth thresholds. The Fluke 5500A can be configured with either a 300 MHz option (SC300) or a 600 MHz option (SC600). The Fluke 5520A offers bandwidth packages up to 300 MHz, 600 MHz, or 1.1 GHz.

Additionally, the Fluke 5520A supports features not available on the Fluke 5500A, including compatibility with Fluke 700 Series pressure modules and compatibility with MET/CAL Plus calibration software. For power quality applications, a specialized version of the higher-end unit, the Fluke 5520A-PQ, is available.

Shared Physical and Operating Standards

The physical chassis and operational footprint of both units are virtually identical. Each model has dimensions of 432 × 178 × 473 mm, requires a 30-minute warm-up period to meet its full specifications, and operates within a temperature range of 0 °C to 50 °C. Both units feature maximum power consumption ratings of 600 VA and can handle line voltage settings from 100 V to 240 V AC (±10%) at frequencies of 47 Hz to 63 Hz. Sourced signals can be controlled via standard GPIB (IEEE-488) or RS-232 interfaces.

Frequently asked questions

What is the primary current capability difference between these two calibrators?
The Fluke 5500A has a maximum current output limit of 11 A for both AC and DC, whereas the Fluke 5520A extends this limit up to 20.5 A.
Which model supports pressure calibration modules?
The Fluke 5520A supports external Fluke 700 Series pressure modules, a capability not specified for the Fluke 5500A.
Are the resistance output modes different on these units?
Yes. The Fluke 5500A sources discrete resistance values up to 329.999 MΩ, while the Fluke 5520A offers a continuously variable resistance output mode up to 1100 MΩ.