The Advantest Q8383 is a high-performance optical spectrum analyzer (OSA) designed for analyzing optical devices and light sources. It features a broad wavelength range of 550 nm to 1750 nm, supporting both single-mode and multimode fibers with a dynamic range of 55 dB and high level accuracy.

Specifications

General

ParameterValue
InterfacesGPIB
Display6.5-inch color LCD

Physical

ParameterValue
Power Supply100 to 120 VAC / 220 to 240 VAC, 50/60 Hz, 150 VA or less
Operating Temperature10 to 40 °C
Dimensions (W×H×D)424 mm × 177 mm × 450 mm
Weight15

Wavelength Specifications

ParameterValue
Wavelength Range350 nm to 1750 nm
Minimum Wavelength350 nm
Maximum Wavelength1750 nm

Resolution and Bandwidth

ParameterValue
Resolution Bandwidth (FWHM)0.1 nm to 5 nm
Minimum Resolution Bandwidth0.1 nm
Maximum Resolution Bandwidth5 nm

Optical Power and Amplitude

ParameterValue
Power Level Accuracy±0.4 dB

Optical Interfaces

ParameterValue
Compatible Fiber TypesSingle-mode, Multimode

External Connectivity and Control

ParameterValue
GPIB (IEEE-488) InterfaceStandard

Wavelength

ParameterValue
Wavelength Accuracy±0.5 nm
Wavelength Reproducibility±0.05 nm

Measurement Power

ParameterValue
Level Linearity±0.5 dB
Polarization Dependency±0.1 dB (at 1.3 µm / 1.55 µm)

Inputs & Outputs

ParameterValue
Optical Input ConnectorFC-PC (standard)

Data Storage

ParameterValue
Storage Drive3.5-inch floppy disk drive (2DD/2HD)

Measurement Capabilities

ParameterValue
Sweep ModesSingle, Continuous, Average
Number of Sampling Points51 to 1001 points

Operating Conditions

ParameterValue
Warm-up Time1 hour

Environmental

ParameterValue
Operating Humidity Range85% RH or less (non-condensing)
Storage Temperature Range-20 to +60 °C

Hardware Options

ParameterValue
Built-in PrinterThermal printer

Technology

Optical Spectrum Analysis

Applications

Fiber optic testing and component spectral evaluation