The Advantest R3265A is a high-performance microwave EMI-adapted portable spectrum analyzer designed for a frequency range of 100 Hz to 8 GHz. It features a high input sensitivity of -145 dBm, standard GPIB and RS-232C interfaces, and a built-in BASIC controller for automated test routines. Its resolution bandwidth ranges down to 10 Hz, and its fast sweep capabilities down to 5 µs/div make it highly suitable for detailed electromagnetic interference measurements and RF analysis.

Specifications

General

ParameterValue
DisplayColor CRT

Physical

ParameterValue
Power Supply100 V to 120 V AC / 220 V to 240 V AC, 50/60 Hz
Operating Temperature0 to 50 °C
Dimensions (W×H×D)350 mm × 177 mm × 450 mm
Weight22 kg

Frequency Specifications

ParameterValue
Frequency Range100 Hz to 8 GHz

Bandwidth (RBW/VBW)

ParameterValue
Resolution Bandwidth (RBW) Range10 Hz to 3 MHz
EMI Bandwidths (6 dB)200 Hz, 9 kHz, 120 kHz, 1 MHz
Video Bandwidth (VBW) Range1 Hz to 3 MHz
RBW Steps1, 3 sequence
VBW Steps1, 3 sequence

Amplitude & Measurement Range

ParameterValue
Maximum Input Sensitivity (DANL)-135 dBm
Maximum Safe RF Input Power+30 dBm (1 W)
Maximum DC Input Voltage0 V

Inputs, Outputs & Connectivity

ParameterValue
RF Input Impedance50 Ω
RF Input Connector TypeN-type (female)

Sweep & Trigger Specifications

ParameterValue
Fastest Sweep Rate5 µs/div
Sweep Time Range (Zero Span)50 µs to 1000 s
Sweep Time Range (Span > 0 Hz)20 ms to 1000 s

Display & User Interface

ParameterValue
BASIC ControllerEquipped as standard
Memory Card InterfaceMemory card function

Frequency

ParameterValue
Frequency Span Range2 kHz - 8 GHz, plus Zero Span
Residual FM< 20 Hz p-p / 100 ms

Frequency Reference

ParameterValue
Internal Reference Oscillator Stability±2 × 10^-8/day, ±1 × 10^-7/year

Performance

ParameterValue
SSB Phase Noise≤ -100 dBc/Hz @ 10 kHz offset

Amplitude

ParameterValue
Amplitude Measurement Range-135 dBm to +30 dBm
Input Attenuator Range0 to 70 dB in 10 dB steps
Calibration Output Signal20 MHz, -20 dBm ±0.3 dB
Frequency Response (100 Hz to 3.2 GHz)±1.0 dB
Frequency Response (3.2 GHz to 8 GHz)±1.5 dB
1 dB Gain Compression>-10 dBm (for frequencies > 10 MHz)
Log Scale Fidelity±1.0 dB / 0 to -90 dB, ±0.2 dB / 1 dB

Display & Sweep

ParameterValue
Log Display Scale10, 5, 2, 1 dB/div
Linear Display Scale10% of reference level/div
Trigger ModesFree Run, Line, Video, External, Single

EMI Measurement

ParameterValue
Quasi-Peak Detector Bandwidths200 Hz, 9 kHz, 120 kHz, 1 MHz (CISPR compliant)

Spurious Responses

ParameterValue
Residual Responses<-100 dBm (1 MHz to 3.2 GHz, no input signal, 50 Ω load)
Second Harmonic Distortion<-70 dBc (10 MHz to 3.2 GHz, mixer level -30 dBm)
Third-Order Intermodulation Distortion<-75 dBc (10 MHz to 8 GHz, mixer level -30 dBm)

Inputs & Outputs

ParameterValue
External Reference Input Frequency10 MHz
External Reference Input Level-10 dBm to +10 dBm

Technology

Superheterodyne Spectrum Analyzer

Applications

Microwave EMI-Adapted Spectrum Analysis

Looking for the Advantest R3265A?

Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.

Selling this kind of equipment? Sell with us