The Aeroflex 3920 is a high-performance radio test set designed for professional engineering, production, and field service applications. This comprehensive platform supports analog PMR (AM, FM, SSB) measurements and advanced digital testing for modern PMR standards including P25, TETRA, DMR, dPMR, NXDN, and HPD. It features high-accuracy RF and audio baseband generators, a 5 MHz channel spectrum analyzer, and a 4 MHz dual-channel oscilloscope.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488.2), LAN (Ethernet), USB Host, USB Device, RS-232, VGA Monitor Output |
| Display | 6.4-inch Color TFT LCD, VGA resolution (640 x 480) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 to 240 VAC, 50 to 60 Hz / 400 Hz, 150 W maximum |
| Operating Temperature | 0 to +50 °C |
| Dimensions (W×H×D) | 356 mm × 197 mm × 520 mm |
| Weight | 16.5 kg |
RF Signal Generator
| Parameter | Value |
|---|---|
| RF Generator Frequency Range | 10 MHz to 2.7 GHz |
| RF Generator Frequency Resolution | 1 Hz |
| RF Generator Frequency Accuracy | Same as internal reference |
| RF Generator Output Level Range (T/R Port) | -130 dBm to -40 dBm |
| RF Generator Output Level Range (Duplex Port) | -130 dBm to -10 dBm |
| RF Generator Output Level Resolution | 0.1 dB |
| RF Generator Output Level Accuracy | ±1.5 dB (typically < ±1.0 dB) |
| RF Generator Harmonics | <-30 dBc |
| RF Generator Non-Harmonics | <-50 dBc (for offsets > 10 kHz) |
| RF Generator Phase Noise | <-95 dBc/Hz (at 20 kHz offset, 1 GHz carrier) |
| FM Deviation Range | 0 to 150 kHz |
| FM Deviation Resolution | 1 Hz |
| FM Deviation Accuracy | ±3% (plus residual FM) |
| AM Modulation Depth Range | 0% to 90% |
| AM Modulation Resolution | 0.1% |
| AM Modulation Accuracy | ±5% of setting (plus residual AM) |
RF Receiver
| Parameter | Value |
|---|---|
| RF Receiver Frequency Range | 10 MHz to 2.7 GHz |
| RF Receiver Input Impedance | 50 Ω nominal |
| RF Receiver Input VSWR (T/R Port) | <1.25:1 (up to 1 GHz) |
| RF Receiver Input VSWR (Antenna Port) | <1.5:1 |
| Maximum Input Power (T/R Port, Continuous) | 50 W |
| Maximum Input Power (T/R Port, Short-term) | 125 W (with duty cycle, 30 seconds max) |
| RF Power Measurement Range (T/R Port) | 100 mW to 125 W |
| RF Power Measurement Accuracy | ±10% (±0.4 dB) |
| RF Power Measurement Resolution | 0.01 dB or 3 digits |
| FM Deviation Measurement Range | 0 to 150 kHz |
| FM Deviation Measurement Accuracy | ±3% (plus residual) |
| AM Depth Measurement Range | 0% to 90% |
| AM Depth Measurement Accuracy | ±5% (plus residual) |
| Demodulation Types | AM, FM, SSB (USB, LSB) |
Audio Generator & Analyzer
| Parameter | Value |
|---|---|
| Audio Generator Count | 3 independent audio modulators/generators |
| Audio Generator Frequency Range | 1 Hz to 20 kHz |
| Audio Generator Frequency Resolution | 0.1 Hz |
| Audio Generator Output Level Range | 0.1 mV to 5 V RMS (into 600 Ω) |
| Audio Generator Output Distortion (THD) | <0.5% (at 1 kHz, 1 V RMS) |
| Audio Analyzer Frequency Range | 10 Hz to 20 kHz |
| Audio Analyzer Level Range | 10 mV to 30 V RMS |
| Distortion Measurement Range (THD) | 0% to 100% |
| Distortion Measurement Accuracy | ±0.5% distortion (±10% of reading) |
| SINAD Measurement Range | 0 to 40 dB |
| SINAD Resolution | 0.1 dB |
| Audio Filters | 300 Hz LP, 300 Hz HP, 3 kHz LP, 15 kHz LP, C-message, CCITT psophometric |
| DTMF Generator/Decoder | Supported |
| DCS Generator/Decoder | Supported |
Digital Radio Test Capabilities
| Parameter | Value |
|---|---|
| Supported Digital Radio Technologies | P25, TETRA, DMR, dPMR, NXDN, HPD (option dependent) |
| TETRA Test Option | Supported (including MS and BS test modes) |
| APCO P25 Test Option | Supported (Phase 1 Conventional/Trunked and Phase 2) |
| DMR Test Option | Supported (TIER II & TIER III) |
| NXDN / dPMR Test Option | Supported |
Spectrum Analyzer & Oscilloscope
| Parameter | Value |
|---|---|
| Spectrum Analyzer Frequency Range | 10 MHz to 2.7 GHz |
| Spectrum Analyzer Resolution Bandwidth (RBW) | 100 Hz to 1 MHz (in 1-3-10 sequence) |
| Spectrum Analyzer Video Bandwidth (VBW) | 10 Hz to 1 MHz |
| Oscilloscope Channels | 1 |
| Oscilloscope Bandwidth | DC to 50 kHz |
Reference Frequency
| Parameter | Value |
|---|---|
| Internal Reference Frequency | 10 MHz |
| Standard Reference Temperature Stability | ±0.1 ppm |
| Standard Reference Aging Rate | ±0.1 ppm per year |
| External Reference Input Frequency | 10 MHz (nominal) |
Environmental & General
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to +71 °C |
| Remote Control Language | SCPI compliant command set |
Power Requirements
| Parameter | Value |
|---|---|
| AC Input Voltage | 100 to 240 VAC |
| AC Input Frequency | 50 to 60 Hz |
| Power Consumption | 150 W (typical) |
Environmental
| Parameter | Value |
|---|---|
| Environmental Compliance | MIL-PRF-28800F Class 3 |
| Operating Humidity | Up to 95% RH (non-condensing) |
| Operating Altitude | Up to 3,048 m (10,000 ft) |
| Shock & Vibration Compliance | Meets MIL-PRF-28800F Class 3 |
Internal Reference
| Parameter | Value |
|---|---|
| Warm-up Time | 15 minutes (standard reference) |
Audio Analyzer
| Parameter | Value |
|---|---|
| Audio Frequency Counter Resolution | 0.1 Hz |
Reference Timebase Options
| Parameter | Value |
|---|---|
| High Stability OCXO Temperature Stability | ±0.01 ppm (0 °C to 50 °C, Option 10) |
| High Stability OCXO Aging Rate | ±0.1 ppm/year (±0.005 ppm/day after 30 days of continuous operation, Option 10) |
RF Generator - Phase Modulation
| Parameter | Value |
|---|---|
| Phase Modulation (PM) Generator Deviation Range | 0 to 10 rad |
| Phase Modulation (PM) Generator Deviation Resolution | 0.001 rad |
| Phase Modulation (PM) Generator Deviation Accuracy | ±5% (at 1 kHz) |
RF Receiver - Phase Modulation Analyzer
| Parameter | Value |
|---|---|
| Phase Modulation (PM) Analyzer Deviation Range | 0 to 10 rad |
| Phase Modulation (PM) Analyzer Deviation Accuracy | ±5% of reading ±0.05 rad (at 1 kHz) |
Audio Generator
| Parameter | Value |
|---|---|
| Audio Generator Output Impedance | <2 Ω |
Oscilloscope
| Parameter | Value |
|---|---|
| Oscilloscope Vertical Sensitivity Range | 10 mV/div to 10 V/div (in 1-2-5 sequence) |
| Oscilloscope Timebase Range | 50 µs/div to 100 ms/div (in 1-2-5 sequence) |
| Oscilloscope Maximum Input Voltage | 30 V RMS |
RF Generator
| Parameter | Value |
|---|---|
| RF Generator Output VSWR (Duplex Port) | <1.25:1 (<1 GHz), <1.30:1 (<2.7 GHz) |
Technology
RF and PMR (Private Mobile Radio) Testing
Applications
Parametric testing of analog and digital radio systems (TETRA, P25, DMR, dPMR, NXDN)