The Agilent 7890B is a highly reliable and widely used high-performance Gas Chromatograph (GC) system. It features 5th-generation Electronic Pneumatics Control (EPC), highly precise oven temperature programming, and support for dual-inlet/multi-detector configurations. The 7890B incorporates integrated sleep/wake modes and optional helium conservation modules to reduce operating costs without compromising retention time repeatability or analytical performance.

Specifications

Physical

ParameterValue
Operating Temperature15 to 35
Dimensions (W×H×D)580 × 490 × 510
Weight49

General System Specifications

ParameterValue
Supported Inlet ChannelsUp to 2
Maximum Heated Zones6 independent heated zones (2 inlets, 2 detectors, 2 auxiliary; excluding oven)
Retention Time Repeatability< 0.0008 min (< 0.008%)

Column Oven

ParameterValue
Oven Dimensions (W×H×D)28 × 31 × 16 cm
Operating Temperature RangeAmbient +4 °C to 450 °C
Cryogenic Cooling OptionsLiquid Nitrogen (-80 °C to 450 °C) or Liquid CO2 (-40 °C to 450 °C)
Temperature Setpoint Resolution0.1 °C
Maximum Number of Temperature Ramps20 ramps (21 plateaus)
Maximum Temperature Programming RateUp to 120 °C/min (240V oven configuration)
Oven Cool-down Time4.0 minutes (450 °C to 50 °C at 22 °C ambient)
Column CapacityAccommodates up to two 105 m capillary columns

Pneumatics and Flow Control (EPC)

ParameterValue
Pressure Setpoint Resolution0.001 psi
Barometric Pressure CompensationStandard
Ambient Temperature CompensationStandard

Inlets (Injectors)

ParameterValue
Split Ratio RangeUp to 7500:1 (for Split/Splitless capillary inlet)
Available Inlet TypesSplit/Splitless (SSL), Multimode Inlet (MMI), Purged Packed (PPI), Programmable Temperature Vaporizing (PTV), Volatiles Interface (VI), Cool On-Column (COC)

Detectors

ParameterValue
Available Detector TypesFID, TCD, micro-ECD, NPD, FPD+, SCD, NCD, MS
FID Data Acquisition RateUp to 500 Hz
FID Linear Dynamic Range> 10^7
TCD Minimum Detectable Level (MDL)< 400 pg tridecane/mL
TCD Linear Dynamic Range10^5
FID Maximum Operating Temperature450 °C
TCD Maximum Operating Temperature400 °C

Autosampler and Sample Handling

ParameterValue
Compatible Liquid AutosamplersAgilent 7693A, 7650A, CTC PAL3
Headspace Sampler CompatibilityAgilent 7697A, 8697

Data Communications and Control

ParameterValue
Compatible Chromatography Data Systems (CDS)Agilent OpenLab CDS, ChemStation, EZChrom, MassHunter

Environmental Specifications

ParameterValue
Operating Humidity Range5% to 95% (non-condensing)

Power Requirements

ParameterValue
Maximum Power Consumption2250 VA (120V configuration) / 2950 VA (240V fast-heating configuration)

Pneumatics

ParameterValue
Auxiliary EPC ChannelsUp to 9 channels (via up to 3 optional modules)

Software & Features

ParameterValue
Sleep/Wake ModeSupported (user-programmable for gas and power conservation)
Early Maintenance Feedback (EMF)Supported

Accessories

ParameterValue
Capillary Flow Technology (CFT)Supported (Deans switch, splitters, purges, backflush)

Oven

ParameterValue
Oven Maximum Run Time999.99 minutes

Inlets

ParameterValue
Split/Splitless (S/SL) Inlet Maximum Operating Temperature400 °C
Multimode Inlet (MMI) Temperature Range-160 °C to 600 °C (with liquid N2 cooling)

Inputs & Outputs

ParameterValue
Analog Signal Outputs2 channels (1 mV, 1 V, 10 V scales)

System Compatibility

ParameterValue
Compatible Mass SpectrometersAgilent 5975, 5977, 7000, 7010, and 7200 Series

Safety

ParameterValue
Hydrogen SafetyOptional built-in hydrogen sensor module

Technology

Gas Chromatography (GC)

Applications

Analytical chemistry, environmental analysis, petrochemicals, pharmaceuticals, food safety, forensics, and materials testing