1100 Series

Agilent*misc

The Agilent 1100 Series is a highly successful and widely deployed modular High-Performance Liquid Chromatography (HPLC) system. Known for its reliability, flexibility, and robust performance, it consists of stackable modules including pumps, autosamplers, column compartments, and various detectors, communicating via a proprietary CAN bus to form an integrated analytical platform. Its specifications represent the gold standard for routine analytical chromatography.

Specifications

General

ParameterValue
InterfacesCAN, GPIB, RS-232C, APG Remote, optional LAN (via interface card)
DisplayOptional Handheld Control Module (G1323B) or via PC software

Physical

ParameterValue
Power Supply100 - 240 VAC, 50/60 Hz (per module)
Operating Temperature4 to 55
Wetted Materials (Standard)Stainless steel, titanium, gold, sapphire, ruby, ceramic, PTFE, PEEK

System Overview

ParameterValue
System ArchitectureModular High-Performance Liquid Chromatography (HPLC) System
Maximum Operating Pressure400 bar (5880 psi)
pH Range1.0 - 12.5 (standard materials)

Solvent Delivery System (Pump Module)

ParameterValue
Pump TypeQuaternary (G1311A) / Binary (G1312A) / Isocratic (G1310A) options
Flow Rate Range0.001 - 10.0 mL/min
Flow Rate Precision≤ 0.3% RSD (typically < 0.15% RSD)
Pressure Range0 - 400 bar (0 - 5880 psi)
Composition Range0 - 100%
Integrated Vacuum Degasser4 channels (G1322A standard degasser)
Degasser Internal Volume~12 mL per channel (G1322A)

Sample Management (Autosampler Module)

ParameterValue
Injection Volume Range0.1 - 100 µL (standard autosampler)
Injection Precision (RSD)< 0.25% RSD (from 5 to 100 µL)
Sample Capacity100 × 2 mL vials (standard tray)
Needle Wash FunctionStandard integrated

Column Compartment

ParameterValue
Temperature Control Range10 °C below ambient to 80 °C (G1316A)
Temperature Stability± 0.15 °C
Temperature Accuracy± 0.8 °C
Heat-up Time5 minutes (from ambient to 40 °C)
Cool-down Time10 minutes (from 40 °C to 20 °C)
Column CapacityUp to three 30 cm columns
Internal Volume3 µL (left heat exchanger), 6 µL (right heat exchanger)

Detection (Optical / UV-Vis / DAD)

ParameterValue
Detector TypeDiode Array Detector (G1315B)
Light Source (DAD)Deuterium and Tungsten lamps
Wavelength Range (DAD)190 - 950 nm
Wavelength Accuracy± 1 nm
Slit Width (DAD)Programmable: 1, 2, 4, 8, 16 nm
Number of Diodes (DAD)1024

Detection (Optical / UV-Vis / VWD)

ParameterValue
Light Source (VWD)Deuterium lamp
Wavelength Range (VWD)190 - 600 nm

Control, Software, and Interfaces

ParameterValue
Supported Control SoftwareAgilent ChemStation, OpenLAB CDS
Instrument Control ProtocolsCAN bus (internal module communication)

Environmental and Regulatory Compliance

ParameterValue
Non-operating Temperature Range-40 to 70 °C
Operating Humidity< 95% at 25 - 40 °C (non-condensing)
Maximum Operating Altitude2000 m (6562 ft)

Physical and Electrical

ParameterValue
Line Voltage100 - 240 VAC, ± 10%
Line Frequency50/60 Hz, ± 5%

Control & Software

ParameterValue
Hardware ControlOptional Handheld Control Module (G1323A/B) for local operation, diagnostics, and method editing

Safety & Diagnostics

ParameterValue
Leak DetectionBuilt-in leak sensors in all modules with safe leak handling and automatic system shutdown
Early Maintenance Feedback (EMF)Continuous tracking of instrument usage (e.g., pump seal wear, lamp burn time) with user-set limits
System DiagnosticsExtensive internal diagnostics, error detection, and continuous monitoring with local/remote logging

Sample Handling

ParameterValue
Supported Sample ContainersStandard 2 mL vials; 100 µL / 250 µL microvials; micro-inserts

Compliance

ParameterValue
GLP/GMP FeaturesBuilt-in system suitability, IQ/OQ protocols, electronic logs
Regulatory CertificationsCE, CSA, UL

Communications

ParameterValue
Inter-module CommunicationCAN (Controller Area Network) bus

Technology

High-Performance Liquid Chromatography (HPLC)

Applications

Analytical Chemistry