The Agilent 6890N is an industry-standard gas chromatograph (GC) system featuring advanced electronic pneumatics control (EPC) and built-in network connectivity for high-performance laboratory analytical chemistry.

Specifications

General

ParameterValue
InterfacesLAN (10Base-T), RS-232-C, APG Remote, dual analog outputs (0-1 V / 0-10 V / 0-1 mV option-dependent)
DisplayVacuum fluorescent display (VFD), 2 lines x 20 characters

Physical

ParameterValue
Power Supply120 V AC (+10% / -10%), 48-66 Hz, 2250 VA max; or 230 V AC (+10% / -10%), 48-66 Hz, 2950 VA max (hardware variant dependent)
Operating Temperature15 °C to 35 °C
Dimensions (W×H×D)580 mm x 500 mm x 540 mm
Weight49 kg

System Configuration & Architecture

ParameterValue
Chromatography TypeGas Chromatography (GC)
Maximum Column CapacityUp to 2 capillary columns or packed columns
Maximum Inlets Supported2
Maximum Detectors Supported2

Column Oven Specifications

ParameterValue
Oven Operating Temperature RangeAmbient +4 °C to 450 °C
Cryogenic Cooling OptionsLiquid CO2 (down to -50 °C) or Liquid N2 (down to -80 °C)
Ambient Temperature Rejection< 0.01 °C oven temperature change per 1 °C ambient temperature change
Temperature Setpoint Resolution1 °C
Maximum Heating Rate120 °C/min
Cool-down Time450 °C to 50 °C in < 6.0 minutes (at 22 °C ambient)
Oven Temperature Ramps6 programming ramps
Oven Temperature Plateaus7 plateaus
Maximum Run Time999.99 minutes
Oven Volume13.9 L
Oven Internal Dimensions280 mm x 305 mm x 165 mm

Gas & Flow Control

ParameterValue
Pneumatics Control TypeElectronic Pneumatics Control (EPC)
Carrier Gas OptionsHelium, Hydrogen, Nitrogen, Argon/Methane
Pressure Control Range0 to 100 psi (standard); 0 to 150 psi (with optional high-pressure EPC)
Pressure Setpoint Resolution0.01 psi

Sample Introduction & Autoinjection

ParameterValue
Supported InletsSplit/Splitless (S/SL), Purged Packed (PPIP), Cool On-Column (COC), Programmable Temperature Vaporizer (PTV), Volatiles Interface (VI)
Autosampler CompatibilityAgilent 7683 Series, Agilent 7673 Series, Agilent 7693A
Maximum Vial CapacityUp to 100 vials (utilizing Agilent 7683 series autosampler with tray)
Split/Splitless Inlet Max Temperature400 °C
Split Ratio RangeUp to 7500:1 (split flow up to 1500 mL/min He/H2)

Detector Specifications & Sensitivity

ParameterValue
Detector Types SupportedFlame Ionization Detector (FID), Thermal Conductivity Detector (TCD), Micro-Electron Capture Detector (uECD), Nitrogen-Phosphorus Detector (NPD), Flame Photometric Detector (FPD), Mass Selective Detector (MSD)
FID Data Acquisition RateUp to 200 Hz
FID Linear Dynamic Range> 10^7
FID Maximum Temperature450 °C
TCD Linear Dynamic Range10^5
TCD Maximum Temperature400 °C
uECD Maximum Temperature400 °C
uECD Radioactivity SourceNickel-63 (Ni-63), < 15 mCi (< 555 MBq)
uECD Linear Dynamic Range> 5 x 10^4
NPD Maximum Temperature400 °C
NPD Dynamic Range> 10^5 N, > 10^5 P

Control, Software & Communication

ParameterValue
Control Interface SoftwareAgilent ChemStation, OpenLab, EZChrom, or compatible third-party CDS software packages
Communication ProtocolsTCP/IP, RS-232, GPIB (optional interface board)

Power & Electrical Requirements

ParameterValue
Power Consumption Peak2250 VA (120 V configuration), 2950 VA (230 V configuration)

Physical & Environmental Specifications

ParameterValue
Operating Humidity Range5% to 95% relative humidity (non-condensing)
Maximum Operating Altitude4,600 m

Compliance & Certifications

ParameterValue
Safety CertificationsCSA, CE, UL
Quality StandardISO 9001 certified manufacturing facility

Detector Specifications

ParameterValue
Nitrogen-Phosphorus Detector (NPD) Selectivity25,000:1 g N/g C, 75,000:1 g P/g C
Flame Photometric Detector (FPD) Selectivity10^5 g C/g S, 10^6 g C/g P

Inlet Specifications

ParameterValue
Split/Splitless Inlet Flow Control Range0 to 200 mL/min (N2), 0 to 1000 mL/min (He/H2)
Split/Splitless Inlet Pressure Control Range0 to 100 psi (standard), 0 to 150 psi (high-pressure option)
Purged Packed Inlet Flow Range0 to 100 mL/min
Cool On-Column Inlet Pressure Range0 to 100 psi

Environmental

ParameterValue
Storage Temperature Range-40 °C to 70 °C

Performance

ParameterValue
Retention Time Repeatability< 0.008 % or < 0.0008 min
Peak Area Repeatability< 1 % RSD

Pneumatics

ParameterValue
Atmospheric Pressure CompensationStandard (compensates for altitude and barometric changes)

Inlets

ParameterValue
Cool On-Column Inlet Maximum Temperature450 °C
Programmed Temperature Vaporizing (PTV) Inlet Maximum Temperature450 °C
Purged Packed Inlet Maximum Temperature400 °C
Volatiles Interface Maximum Temperature400 °C

Detectors

ParameterValue
Flame Photometric Detector (FPD) Maximum Temperature250 °C
Flame Photometric Detector (FPD) Dynamic Range> 10^3 (S), > 10^4 (P)

Inputs & Outputs

ParameterValue
Analog OutputsTwo channels (0-1 V, 0-10 V, or 0-1 mV selectable)

Oven

ParameterValue
Oven Temperature Stability< ±0.01 °C for each 1 °C change in ambient temperature

Technology

Gas Chromatography (GC)

Applications

Analytical Chemistry