The Anritsu MG3697C is a high-performance, broadband microwave signal generator within the MG3690C series. It covers frequencies from 2 GHz to 67 GHz in its standard configuration, and is operational up to 70 GHz. It offers exceptional phase noise performance, extensive analog modulation capabilities, fast frequency switching, and is highly configurable with options extending its frequency range down to 0.1 Hz.
Specifications
Frequency
| Parameter | Value |
|---|---|
| Minimum Frequency | 2 GHz (down to 0.1 Hz with Option 22) |
| Maximum Frequency | 67 GHz (operational to 70 GHz) |
| Frequency Resolution | 0.01 Hz (standard) / 0.1 mHz (with Option 10) |
Output
| Parameter | Value |
|---|---|
| Maximum Output Power | +13 dBm (frequency dependent; up to +19 dBm with Option 15) |
| Phase Noise | -100 dBc/Hz at 10 kHz offset (10 GHz carrier, standard) |
Modulation
| Parameter | Value |
|---|---|
| Modulation | AM, FM, Phase (ΙM), Pulse (internal & external) |
| Sweep Modes | Step Sweep, Analog Sweep, Manual Sweep |
| Frequency Modulation Maximum Deviation | 10 MHz (100 MHz in Wide FM mode, Option 12 dependent) |
| Phase Modulation Maximum Deviation | 10 rad (400 rad in Wide ΦM mode, Option 12 dependent) |
General
| Parameter | Value |
|---|---|
| Reference Oscillator | 10 MHz internal reference |
| Interfaces | GPIB (IEEE-488.2), LAN (10/100 Base-T), RS-232 |
| Display | Backlit graphic LCD front panel |
| Save/Recall Memory States | 10 front panel setups |
| Warm-up Time | 1 hour |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 90 to 264 VAC, 47 to 440 Hz |
| Dimensions (W×H×D) | 429 mm W × 133 mm H × 597 mm D |
| Weight | 21 kg |
Frequency Specifications
| Parameter | Value |
|---|---|
| Internal Reference Aging Rate | < 2 x 10^-9 per day (after 30 days warm-up) |
| Internal Reference Temperature Stability | < ±2 x 10^-8 over 0 °C to 55 °C |
| External Reference Input Frequency | 10 MHz (±10 Hz) |
| External Reference Input Amplitude | 0.1 Vp-p to 6 Vp-p, 50 Ω impedance |
Amplitude Specifications
| Parameter | Value |
|---|---|
| Amplitude Resolution | 0.01 dB |
| Output Impedance | 50 Ω (nominal) |
| Attenuator Range | 0 to 110 dB in 10 dB steps (with Option 2) |
Spectral Purity
| Parameter | Value |
|---|---|
| Harmonics | <-30 dBc (2 to 20 GHz), <-20 dBc (20 to 67 GHz) |
Analog Modulation
| Parameter | Value |
|---|---|
| Amplitude Modulation Depth Range | 0% to 90% (linear) |
| Pulse Modulation On/Off Ratio | > 80 dB |
| Pulse Modulation Minimum Pulse Width | 100 ns (leveled), 10 ns (unleveled) |
| Pulse Modulation Rise/Fall Time | < 10 ns (typical) |
| Pulse Generator Repetition Frequency | 10 mHz to 1 MHz |
Sweep & Operating Modes
| Parameter | Value |
|---|---|
| Dwell Time Range | 1 ms to 99 sec |
| Trigger Modes | Auto, External, Single, GPIB |
Connectivity & Interfaces
| Parameter | Value |
|---|---|
| RF Output Connector Type | V female (1.85 mm) |
Physical & Environmental
| Parameter | Value |
|---|---|
| Operating Temperature Range | 0 °C to 50 °C |
| Storage Temperature Range | -40 °C to +75 °C |
| Relative Humidity Range | 5% to 95% non-condensing |
| Rack Unit Height | 3U |
Power Requirements
| Parameter | Value |
|---|---|
| Power Consumption | < 350 VA |
Hardware & Software Options
| Parameter | Value |
|---|---|
| Option 2 | Option 2C = Step Attenuator (adds 90 dB attenuator) |
| Option 3 | Ultra-Low Phase Noise Performance |
| Option 4 | Low Frequency Extension (10 MHz to 2 GHz) |
| Option 22 | Low Frequency Extension (0.1 Hz to 2 GHz) |
| Option 15 | High Power Output |
Sweep
| Parameter | Value |
|---|---|
| List Sweep Points | Up to 10,000 points |
Environmental
| Parameter | Value |
|---|---|
| Maximum Operating Altitude | 4,600 m |
Interfaces
| Parameter | Value |
|---|---|
| GPIB Emulation Languages | SCPI, Anritsu MG3690A/B, Wiltron 6700B, HP 8340/8341, HP 8360, HP 8673, HP 8663 |
Compliance
| Parameter | Value |
|---|---|
| Safety Compliance | EN 61010-1 |
| EMC Compliance | EN 61326-1, EN 55011 |
Options
| Parameter | Value |
|---|---|
| Option 5 | Analog Sweep Capability |
| Option 12 | Frequency and Phase Modulation (FM/ΦM) |
| Option 13 | Pulse Modulation |
| Option 14 | Amplitude Modulation (AM) |
| Option 16 | High Stability Timebase |
| Option 17 | Delete Front Panel |
| Option 20 | Scan Modulation |
| Option 28 | Option 28 = Analog Modulation (combines Options 12, 13, and 14) |
| Option 9 Rear Panel RF Output | Moves the RF output connector to the rear panel |
| Option 11 Internal LF Generator | Adds an internal LF generator (0.1 Hz to 1 MHz sine, 0.1 Hz to 100 kHz square, triangle, ramp, noise) |
Reference Oscillator Options
| Parameter | Value |
|---|---|
| Option 7 Rubidium Aging Rate | < 5 x 10^-11 per day (after 30 days warm-up) |
| Option 7 Rubidium Temperature Stability | < ±2 x 10^-10 over 0 °C to 50 °C |
| Option 7 Rubidium Warm-up Time | < 7.5 minutes to lock, < 10 minutes to meet temperature stability |
RF Output Performance
| Parameter | Value |
|---|---|
| Source SWR | < 1.4 (< 2 GHz), < 2.0 (2 to 40 GHz), < 2.3 (40 to 70 GHz) typical |
Pulse Modulation
| Parameter | Value |
|---|---|
| Pulse Width Resolution | 10 ns |
| Pulse Delay Resolution | 10 ns |
| Internal Pulse Generator Modes | Free-run, Delayed, Gated, Triggered, Doublet, Triplet, Quadruplet |
Amplitude Performance
| Parameter | Value |
|---|---|
| Power Sweep Resolution | 0.01 dB |
Phase Modulation
| Parameter | Value |
|---|---|
| Phase Modulation Bandwidth | DC to 100 kHz (3 dB bandwidth) |
Frequency Modulation
| Parameter | Value |
|---|---|
| DC FM Modulation Bandwidth | DC to 10 MHz (3 dB bandwidth in FM Extreme Mode) |
| AC FM Modulation Bandwidth | 100 Hz to 250 kHz (3 dB bandwidth) |
Technology
RF and Microwave Synthesized Signal Generator
Applications
LO Substitution, Clock Generation, Radar Simulation, RF Component Characterization