The Anritsu Spectrum Master MS2720T is a high-performance handheld spectrum analyzer designed for field technicians and engineers. Offering continuous frequency coverage up to 43 GHz, it delivers benchtop-grade measurements in a portable, ruggedized form factor, featuring an 8.4-inch touchscreen and optional tracking generators up to 20 GHz.

Specifications

General

ParameterValue
InterfacesUSB Type A (x2), USB Type Mini-B, 10/100 Base-T Ethernet (LAN) RJ45
Display8.4-inch SVGA Touchscreen, 800 x 600 resolution
Internal Memory CapacityInternal: > 40,000 traces, > 40,000 screens, > 10,000 setups
EMC StandardsEN 61326-1, Class A
Safety StandardsEN 61010-1

Physical

ParameterValue
Power SupplyAC/DC Adapter: 100 to 240 VAC, 50/60 Hz; External DC Input: 12 to 15 VDC
Operating Temperature-10 °C to +55 °C
Dimensions (W×H×D)315 mm × 211 mm × 77 mm
Weight4.4 to 4.9 kg

Frequency Specifications

ParameterValue
Frequency Range (MS2720T-0709)9 kHz to 9 GHz
Frequency Range (MS2720T-0713)9 kHz to 13 GHz
Frequency Range (MS2720T-0720)9 kHz to 20 GHz
Frequency Range (MS2720T-0732)9 kHz to 32 GHz
Frequency Range (MS2720T-0743)9 kHz to 43 GHz
Frequency Tuning Resolution1 Hz
Frequency Span Range10 Hz to max frequency of model, plus 0 Hz (zero span)
Internal Frequency Reference Aging±1.0 ppm/10 years
Internal Frequency Reference Accuracy±0.12 ppm (after 20 min warm-up) / ±0.05 ppm (with GPS ON)

Resolution and Video Bandwidths (RBW/VBW)

ParameterValue
Resolution Bandwidth (RBW) Range1 Hz to 10 MHz
Resolution Bandwidth Step Type1-3 sequence (1 Hz, 3 Hz, 10 Hz, 30 Hz...)
Resolution Bandwidth Accuracy±10 % (1 Hz to 10 MHz, -3 dB)
Video Bandwidth (VBW) Range1 Hz to 10 MHz
Video Bandwidth Step Type1-3 sequence
Zero Span Bandwidth OptionsSupports up to 10 MHz RBW & VBW

Sweep and Triggering

ParameterValue
Sweep ModesPerformance, Fast, No FFT, Burst Detect
Burst Detect Capture DurationCaptures bursts as short as 200 µs
Sweep Trigger SourcesFree Run, Video, External, Gated, GPS
Trigger SettingsHysteresis, Hold-off, Delay

Amplitude and Level Specifications

ParameterValue
Measurement RangeDANL to +30 dBm
Maximum Safe Input Level (Continuous Wave, >= 10 dB Attenuation)+30 dBm Peak, ±50 VDC
Maximum Safe Input Level (Continuous Wave, Preamp ON)+10 dBm Peak, ±50 VDC
Display Range1 to 15 dB/div in 1 dB steps, ten divisions displayed
Input Attenuation Range0 to 65 dB
Input Attenuation Step Size5.0 dB
Reference Level Range-150 dBm to +30 dBm
Level Display Units (Log)dBm, dBV, dBmV, dBµV
Level Display Units (Linear)nV, µV, mV, V, kV, nW, µW, mW, W, kW

Spectral Purity and Dynamic Range

ParameterValue
Dynamic Range>106 dB minimum at 2.4 GHz (2/3 TOI-DANL in 1 Hz RBW)
SSB Phase Noise at 1 GHz (10 kHz offset)–112 dBc/Hz
SSB Phase Noise (9 GHz model, 100 kHz offset)-110 dBc/Hz maximum (-115 dBc/Hz typical)

Tracking Generator (Option)

ParameterValue
Tracking Generator Frequency Range100 kHz to 9, 13, or 20 GHz (option dependent)
Tracking Generator Power Step Size0.1 dB

Measurement Capabilities and Software Suites

ParameterValue
Standard Smart MeasurementsField Strength, Occupied Bandwidth (OBW), Channel Power, ACPR, AM/FM/SSB Demodulation, Carrier-to-Interference (C/I), Emission Mask
Channel Power MeasurementMeasures the total power in a specified bandwidth
AM/FM/SSB DemodulationAM, wide/narrow FM, upper/lower SSB (audio output only)
Optional Cellular/Wireless Technology StandardsLTE (FDD & TDD), GSM/GPRS/EDGE, W-CDMA/HSPA+, TD-SCDMA/HSPA+, CDMA/EV-DO
LTE Measurements BandwidthUp to 20 MHz
Interference Analyzer OptionSpectrogram, Signal Strength Meter, Interference Mapping
High Accuracy Power MeterSupported (Option)
IF Output Option30 MHz wide Zero-Span IF Output at 140 MHz center frequency

Port and Interface Connections

ParameterValue
RF Input Connector (9, 13, 20 GHz models)Type N, female, 50 Ω
RF Input Connector (32, 43 GHz models)Ruggedized Type K, male, 50 Ω
External Reference Input Frequency10 MHz
External Trigger Input ConnectorBNC, female
External Audio Headphone Jack3.5 mm 3-conductor headset jack
GPS Antenna Input ConnectorSMA, female (with GPS option)

Display and User Interface

ParameterValue
Display Options for ReadabilityNormal, black and white, night vision, or high contrast
Number of Traces6
Trace Detection ModesPeak, RMS, Negative, Sample, Normal
Marker Functions6 markers, delta markers, noise markers, frequency counter

Power and Battery Management

ParameterValue
Battery Operating TimeUp to 3 hours typical

Physical and Ergonomic Design

ParameterValue
Weight (9, 13, 20 GHz models)4.4 kg (9.7 lbs) including battery
Weight (32, 43 GHz models)4.9 kg (10.8 lbs) including battery

Environmental and Compliance

ParameterValue
Storage Temperature Range-40 °C to +75 °C
Recommended Calibration Cycle12 months

Performance

ParameterValue
Third-Order Intercept (TOI)+20 dBm typical (at 2.4 GHz, Preamp Off)
Absolute Amplitude Accuracy±0.3 dB
Internal Frequency Reference Temperature Stability±0.02 ppm (0 °C to 50 °C)
Quasi-Peak DetectorStandard (with 200 Hz, 9 kHz, 120 kHz CISPR resolution bandwidths)

Inputs & Outputs

ParameterValue
Maximum RF Input DC Voltage±50 VDC
USB Interfaces2 x USB Type A (Host), 1 x USB Mini-B (Device)
Ethernet Interface10/100 Base-T (RJ-45)
External Reference Input Level-10 dBm to +10 dBm
RF Input Impedance50 Ω
RF Input VSWR< 1.5:1 typical (9 GHz and 13 GHz models, ≥ 10 dB attenuation); < 1.8:1 typical (20 GHz, 32 GHz, and 43 GHz models, ≥ 10 dB attenuation)
Tracking Generator ConnectorN female (9, 13, 20 GHz models with Option 809/813/820)
External Trigger Input LevelTTL threshold (minimum high level ≥ 1.4 V, maximum low level ≤ 0.4 V)
GPS Antenna Voltage3.3 V or 5.0 V (software selectable for active antennas)

Tracking Generator Option

ParameterValue
Tracking Generator Output Power Range-50 dBm to 0 dBm (Option 809/813/820)

IF Output Option

ParameterValue
IF Output Center Frequency140 MHz (Option 89)
IF Output Bandwidth30 MHz (Option 89)

GPS Option

ParameterValue
GPS Frequency Accuracy±25 ppb (with GPS on and locked, Option 31)

Display

ParameterValue
Display Resolution800 × 600 pixels (SVGA)

Power

ParameterValue
Power Consumption26 W typical (spectrum analyzer only, battery charging off); 35 W typical (with Tracking Generator, battery charging off)
Battery TypeLithium-ion

Environmental

ParameterValue
Maximum Operating Altitude4,600 m (15,000 ft)
Operating Humidity95% non-condensing
Environmental ClassMIL-PRF-28800F Class 2

Receiver & Front End

ParameterValue
Tracking PreselectorStandard on all models (9 kHz to maximum frequency of the model)

Bandwidth

ParameterValue
Resolution Bandwidth Shape Factor< 5:1 (60 dB / 3 dB bandwidth)

Sweep Functions

ParameterValue
Gated Sweep OptionOption 0090
Gated Sweep SetupGate Delay: 1 µs to 10 s; Gate Length: 1 µs to 10 s
Zero Span Sweep Time Range10 µs to 600 s

Frequency

ParameterValue
Preamplifier Frequency Range100 kHz to 9/13/20/32/43 GHz (model dependent)

Technology

Superheterodyne Spectrum Analysis

Applications

RF/Microwave Signal Analysis, Spectrum Monitoring, Interference Hunting, Cellular Network Testing