The California Instruments 5001iX is a high-performance programmable AC and DC power source from the iX Series, providing up to 5000 VA of AC power and 3750 W of DC power. Designed for precision testing, frequency conversion, and power simulation, the 5001iX features a built-in analyzer for comprehensive measurements, arbitrary waveform generation, and standard GPIB and RS-232 interfaces in a compact 4U rackmount chassis.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488.2), RS-232 |
| Display | Backlit graphic LCD |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 208-240 VAC (1-Phase or 3-Phase) or 380-415 VAC (3-Phase) ±10%, 47-63 Hz |
| Operating Temperature | 0 to 40 °C |
| Dimensions (W×H×D) | 483 mm × 178 mm × 602 mm |
| Weight | 50 |
AC Input Specifications
| Parameter | Value |
|---|---|
| Input Phase Configuration | 1-Phase or 3-Phase (factory configured) |
| Input Voltage Range | 187 to 264 VAC (208/240V model) or 342 to 456 VAC (380/415V model) |
| Input Frequency Range | 47 to 63 Hz |
AC Output Specifications
| Parameter | Value |
|---|---|
| Maximum Output Power | 5000 VA |
| AC Output Voltage Ranges | 0 to 150 VRMS / 0 to 300 VRMS |
| Maximum AC Current (Low Range) | 33.3 ARMS |
| Maximum AC Current (High Range) | 16.6 ARMS |
| Output Frequency Range | 16 Hz to 1000 Hz (up to 5000 Hz with Option HF) |
| Frequency Resolution | 0.01 Hz (16.00 to 99.99 Hz), 0.1 Hz (100.0 to 999.9 Hz), 1 Hz (1000 to 5000 Hz with Option HF) |
| Frequency Accuracy | ±0.01% |
| Voltage Resolution | 0.1 V |
| Voltage Programming Accuracy | ±0.2% of full scale |
| Output Phase Configuration | 1-Phase |
| Crest Factor | 3:1 |
| Voltage Line Regulation | ±0.1% for ±10% line change |
| Total Harmonic Distortion (THD) | < 0.5% (45 to 66 Hz), < 1.0% (66 to 500 Hz), < 2.0% (500 to 1000 Hz) |
DC Output Specifications
| Parameter | Value |
|---|---|
| Maximum DC Power | 3750 W |
| DC Output Voltage Ranges | 0 to 200 VDC / 0 to 400 VDC |
| Maximum DC Current (Low Range) | 18.75 ADC |
| Maximum DC Current (High Range) | 9.37 ADC |
| DC Voltage Load Regulation | ±0.1% with remote sense |
| DC Voltage Line Regulation | ±0.1% for ±10% line change |
Measurement & Metrology
| Parameter | Value |
|---|---|
| Voltage Measurement Range (RMS) | 0 to 300 VRMS |
| Voltage Measurement Accuracy | ±(0.1% of reading + 0.2% of range) |
| Voltage Measurement Resolution | 0.1 V |
| Current Measurement Range (RMS) | 0 to 33.3 ARMS |
| Current Measurement Accuracy | ±(0.2% of reading + 0.4% of range) |
| Current Measurement Resolution | 0.01 A |
| Real Power Measurement Range | 0 to 5000 W |
| Real Power Measurement Accuracy | ±(0.3% of reading + 0.5% of range) |
| Apparent Power Measurement Range | 0 to 5000 VA |
| Power Factor Measurement Range | 0.00 to 1.00 |
| Harmonic Distortion Measurement (THD) | Up to 50th harmonic |
Transient & Waveform Programming
| Parameter | Value |
|---|---|
| Standard Waveform Types | Sine, Square, Clipped Sine |
| Arbitrary Waveform Memory Depth | 1024 points per waveform |
| Transient Programming Modes | List, Step, Pulse |
| Number of Arbitrary Waveforms | Up to 16 user-defined waveforms |
Protection & Safety
| Parameter | Value |
|---|---|
| Over-Voltage Protection (OVP) Range | Programmable shutdown |
| Over-Current Protection (OCP) Range | Programmable (current limit or shutdown) |
| Over-Temperature Protection (OTP) | Automatic shutdown |
System Control & Connectivity
| Parameter | Value |
|---|---|
| Remote Sense Terminals | Yes (rear panel) |
| External Trigger Input / Output | Yes (BNC connectors) |
Environmental Specifications
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to 85 °C |
| Cooling Mechanism | Forced air, side-to-rear |
Physical & Regulatory
| Parameter | Value |
|---|---|
| Enclosure Type / Rack Units (U) | Rackmount, 4U |
| Safety Compliance Certifications | CE Mark |
Output Specifications
| Parameter | Value |
|---|---|
| Peak AC Current (Low Range) | 112 A peak |
| Peak AC Current (High Range) | 56 A peak |
| DC Output Offset Voltage | < 50 mV |
Power Specifications
| Parameter | Value |
|---|---|
| Maximum Input Current (208 V AC Input) | 24 A RMS per phase |
| Maximum Input Current (400 V AC Input) | 13 A RMS per phase |
Measurement Specifications
| Parameter | Value |
|---|---|
| Reactive Power (VAR) Measurement Range | 0 to 5000 VAR |
| Phase Angle Measurement Range | 0.0° to 359.9° |
Arbitrary Waveform Generator
| Parameter | Value |
|---|---|
| Waveform Generation Method | Direct Digital Synthesis (DDS) |
| Waveform DAC Resolution | 12-bit |
| Waveform Point Memory | 1024 points per cycle |
Programmable Impedance (Option -I)
| Parameter | Value |
|---|---|
| Output Impedance Rs Programming Range (Option -I) | 0 to 2000 mΩ |
| Output Impedance Ls Programming Range (Option -I) | 0 to 2000 µH |
| Output Impedance Programming Resolution (Option -I) | 1 mΩ / 1 µH |
Controller
| Parameter | Value |
|---|---|
| Programming Language | SCPI |
Input
| Parameter | Value |
|---|---|
| Input Protection | Internal circuit breaker |
Technology
PWM Switcher
Applications
AC/DC Power Simulation, Compliance Testing, Frequency Conversion