The Fluke 12 is a compact, reliable digital multimeter from the Fluke 10 Series, featuring VChek™ automatic measurement selection, capacitance measurement, min/max recording with relative elapsed time, and a 4000-count digital display. Built for electrical troubleshooting, it provides automatic AC/DC voltage selection and manual/automatic ranging.

Specifications

General

ParameterValue
Display4000-count LCD

Physical

ParameterValue
Power Supply9 V battery (NEDA 1604, 6F22, or 006P)
Operating Temperature-10 °C to 50 °C
Dimensions (W×H×D)70.5 mm × 142.3 mm × 34.6 mm

DC Voltage Measurement

ParameterValue
DC Voltage Range4.000 V / 40.00 V / 400.0 V / 600 V
DC Voltage Best Resolution1 mV
DC Voltage Best Accuracy±(0.9% + 1)
Overload Protection (DC)600 V rms

AC Voltage Measurement

ParameterValue
AC Voltage Range4.000 V / 40.00 V / 400.0 V / 600 V
AC Voltage Best Resolution1 mV
AC Voltage Best Accuracy±(1.9% + 3)
Measurement TypeAverage responding, RMS calibrated
Overload Protection (AC)600 V rms

Resistance Measurement

ParameterValue
Resistance Range400.0 Ω / 4.000 kΩ / 40.00 kΩ / 400.0 kΩ / 4.000 MΩ / 40.00 MΩ
Resistance Best Resolution0.1 Ω
Resistance Best Accuracy±(0.9% + 1)
Overload Protection (Resistance)600 V rms

Continuity and Diode Testing

ParameterValue
Diode Test Range2.000 V
Diode Test Resolution1 mV
Overload Protection (Continuity & Diode)600 V rms

Capacitance Measurement

ParameterValue
Capacitance Range1.000 µF / 10.00 µF / 100.0 µF / 1000 µF / 10,000 µF
Capacitance Best Resolution0.001 µF
Capacitance Best Accuracy±(1.9% + 2)
Overload Protection (Capacitance)600 V rms

Special Functions and Modes

ParameterValue
Automatic Volts/Ohms SelectionVChek™ automatic measurement selection
Ranging ModesManual and automatic ranging
Min/Max RecordingRecords highest and lowest voltage readings over 100 hours
Relative Time StampDisplays elapsed time since recording started up to 100 hours
Sleep Mode (Auto Power Off)Powers down after 45 minutes of inactivity

Safety and Compliance

ParameterValue
Maximum Working Voltage600 V AC/DC between any terminal and earth ground
Agency Approvals & CertificationsANSI/ISA-S82.01-1994, CSA C22.2 No. 1010.1-92, EN61010-1:1993

Physical and Mechanical Specifications

ParameterValue
Display Update Rate4 times per second

Power Requirements and Environmental Limits

ParameterValue
Battery Life650 hours typical with alkaline battery
Storage Temperature Range-30 °C to 60 °C
Operating Humidity Range0% to 90% (0 °C to 35 °C), 0% to 70% (35 °C to 50 °C)
Temperature Coefficient0.1 × (specified accuracy) per °C (< 18 °C or > 28 °C)
Operating AltitudeUp to 2000 m

Performance

ParameterValue
Continuity Capture Minimum Pulse Width250 µs
VChek Voltage Switch ThresholdApproximately 4.5 V AC/DC

Resistance

ParameterValue
Resistance Mode Open Circuit Voltage< 1.3 V DC (< 1.5 V DC on 40 MΩ range)
Resistance Mode Full Scale Voltage< 450 mV (up to 4 MΩ), < 1.3 V (40 MΩ)
Resistance Mode Short Circuit Current< 500 µA

Maintenance

ParameterValue
Recommended Calibration Interval1 year

Environmental

ParameterValue
Storage Altitude12000 m

Safety & Compliance

ParameterValue
Pollution Degree2
Rated Transient Overvoltage6 kV peak

Technology

Digital Multimeter

Applications

Electrical testing, troubleshooting, and general-purpose maintenance