The Fluke Norma 5000 is a modular, high-precision power analyzer designed for complex power electronics testing. It supports up to six power phases and features simultaneous parallel acquisition of all phases. With bandwidths up to 10 MHz and sample rates up to 1 MHz (configuration dependent), it delivers class-leading accuracy for any waveform, frequency, or phase shift. Built-in functions include harmonic analysis up to the 40th order, FFT, vector diagram display, and DSO mode, alongside an internal printer.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | RS232, IEEE488, Ethernet, USB 2.0 (by choice/option) |
| Display | 144 mm (5.7 in) color display |
Physical
| Parameter | Value |
|---|---|
| Operating Temperature | +5 to +35 |
| Dimensions (W×H×D) | 427 x 150 x 310 mm |
| Weight | 7 |
General Characteristics
| Parameter | Value |
|---|---|
| Number of Power Phases | 1 to 6 (configurable) |
| Measurement Bandwidth | DC to 3 MHz / 10 MHz (configuration dependent) |
| Sample Rate | 102 kHz / 341 kHz / 1 MHz (configuration dependent) |
| Measurement Architecture | Simultaneous parallel acquisition on all phases; galvanically separated inputs |
Power and Energy Measurement
| Parameter | Value |
|---|---|
| Average Time | 15 ms to 3600 s (freely selectable) |
Frequency and Harmonic Analysis
| Parameter | Value |
|---|---|
| FFT Features | FFT analysis included |
Display and Data Processing
| Parameter | Value |
|---|---|
| Internal Storage Capacity | 4 MB RAM for sample or average data |
| Internal Printer | Optional |
| Waveform Display Capability | DSO (Scope) mode included |
| Vector/Phasor Display Capability | Vector diagram mode included |
| Recorder Function | Included |
Motor and Mechanical Evaluation (Optional)
| Parameter | Value |
|---|---|
| Motor and Mechanical Evaluation | Torque and speed measurement via external sensors (requires PI1 Process Interface option) |
Interfaces and Connectivity
| Parameter | Value |
|---|---|
| Analog Outputs (D/A) | 4 analog outputs (with PI1 Process Interface option) |
Inputs
| Parameter | Value |
|---|---|
| Voltage Input Ranges | 0.3 V, 1 V, 3 V, 10 V, 30 V, 100 V, 300 V, 1000 V |
| Current Input Ranges (10 A Module Option) | 30 mA, 100 mA, 300 mA, 1 A, 3 A, 10 A |
| External Shunt / BNC Input Ranges | 30 mV, 100 mV, 300 mV, 1 V, 3 V, 10 V |
Environmental
| Parameter | Value |
|---|---|
| Storage Temperature | -20 °C to +50 °C |
Compliance
| Parameter | Value |
|---|---|
| Safety Rating / Measurement Category | 1000 V CAT II / 600 V CAT III |
Technology
Precision Power Analysis
Applications
Power Electronics, Motor Drives, Lighting, and Transformer Testing