The Fluke 5100B is a microprocessor-controlled multifunction calibrator designed to calibrate analog and digital multimeters up to 4.5 digits. It provides accurate, low-noise AC/DC voltage, AC/DC current, and resistance outputs. The calibrator can be enhanced with a 10 MHz wideband voltage option and features direct-drive outputs for Fluke power and transconductance amplifiers.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (Option 05) / RS-232 (Option 06) |
| Display | Gas-discharge digital display (neon) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 / 120 / 220 / 240 VAC (±10%), 48 Hz to 62 Hz |
| Operating Temperature | 0 °C to 50 °C |
| Dimensions (W×H×D) | 432 mm × 178 mm × 508 mm |
| Weight | 34 kg |
General & Metrological Overview
| Parameter | Value |
|---|---|
| Calibrator Type | Multifunction Calibrator |
| Target Multimeter Resolution | Up to 4.5 digits |
| Calibration Interval Recommended | 1 Year |
DC Voltage Source
| Parameter | Value |
|---|---|
| DC Voltage Range | 0 V to 1100 V |
| DC Voltage Best 1-Year Accuracy | 0.005% |
| DC Voltage Range Steps | 200 mV, 2 V, 20 V, 200 V, 1100 V |
AC Voltage Source
| Parameter | Value |
|---|---|
| AC Voltage Range | 0 V to 1100 V |
| AC Voltage Best 1-Year Accuracy | 0.05% |
| AC Voltage Range Steps | 200 mV, 2 V, 20 V, 200 V, 1100 V |
| AC Voltage Standard Frequency Range | 50 Hz to 50 kHz |
| Optional Wideband Frequency Range | 10 Hz to 10 MHz (with Option 03) |
DC Current Source
| Parameter | Value |
|---|---|
| DC Current Range | 200 µA to 2 A |
| DC Current Best 1-Year Accuracy | 0.025% |
| DC Current Range Steps | 200 µA, 2 mA, 20 mA, 200 mA, 2 A |
AC Current Source
| Parameter | Value |
|---|---|
| AC Current Range | 200 µA to 2 A |
| AC Current Best 1-Year Accuracy | 0.07% |
| AC Current Range Steps | 200 µA, 2 mA, 20 mA, 200 mA, 2 A |
| AC Current Frequency Range | 50 Hz to 1 kHz |
Resistance Simulation
| Parameter | Value |
|---|---|
| Resistance Range | 1 Ω to 10 MΩ |
| Resistance Best 1-Year Accuracy | 0.005% |
| Resistance Step Values | 1 Ω, 10 Ω, 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ |
System Control & Connectivity
| Parameter | Value |
|---|---|
| Wideband Option Compatibility | Option 03 (Low-noise, flat 10 MHz wideband AC source) |
| Direct Drive Compatibility | Fluke 5220A 20A Transconductance Amplifier & Fluke 5205A Precision Power Amplifier |
Environmental
| Parameter | Value |
|---|---|
| Relative Humidity | <80% RH (0 °C to 35 °C), <70% RH (35 °C to 50 °C) |
Performance
| Parameter | Value |
|---|---|
| Warm-up Time | 1 hour |
Electrical Safety
| Parameter | Value |
|---|---|
| Maximum Guard-to-Chassis Voltage | 1000 V dc or peak ac |
AC Outputs
| Parameter | Value |
|---|---|
| Frequency Resolution | 3 digits |
| Frequency Accuracy | ±0.1% |
Operation
| Parameter | Value |
|---|---|
| Sensing Capability | 2-wire or 4-wire external sensing |
Technology
Analog-to-digital metrology standard
Applications
Calibration of analog and digital multimeters up to 4.5 digits