The Fluke 5522A is a high-performance Multi-Product Calibrator designed to address a wide workload of electrical and electronic calibration. Although the product is sometimes miscategorized as a VNA calibration kit in standard catalogs, the Fluke 5522A is a primary laboratory electrical calibrator. It calibrates digital multimeters (up to 6.5 digits), oscilloscopes (up to 600 MHz or 1.1 GHz with optional modules), power meters, power quality analyzers, thermometers, and various other measurement instruments. It features robust reverse-power protection (V-Guard) to safeguard against accidental operator voltage inputs, ensuring long-term reliability.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488), RS-232
DisplayDual displays: high-contrast backlit graphic LCD for menus/settings, and a segmented vacuum fluorescent display (VFD) for primary output values

Physical

ParameterValue
Power Supply100 V to 240 V AC (±10%), 47 Hz to 63 Hz, < 300 VA
Operating Temperature0 °C to 50 °C
Dimensions (W×H×D)443 mm × 178 mm × 473 mm
Weight22.4

DC Voltage

ParameterValue
DC Voltage Range0 to ±1020 V
DC Voltage Resolution0.1 µV
DC Voltage 1-Year Absolute Uncertainty±(11 ppm of reading + 1.5 µV) on 3.3 V range
DC Voltage Ranges329.9999 mV, 3.299999 V, 32.99999 V, 329.9999 V, 1020.000 V
DC Voltage Maximum Compliance Current220 mA (on 3.3 V and 33 V ranges)
DC Voltage Source Impedance50 Ω (on 330 mV range)

AC Voltage

ParameterValue
AC Voltage Range1 mV to 1020 V
AC Voltage Frequency Range10 Hz to 500 kHz
AC Voltage 1-Year Absolute Uncertainty±(0.015% of reading + 0.002% of range)
AC Voltage WaveformsSine, square, triangle, truncated sine

DC Current

ParameterValue
DC Current Range0 to ±20.5 A
DC Current 1-Year Absolute Uncertainty±(100 ppm of reading + 0.05 µA) on 3.3 mA range
DC Current Maximum Compliance Voltage7 V (up to 330 mA)

AC Current

ParameterValue
AC Current Range29 µA to 20.5 A
AC Current Frequency Range10 Hz to 30 kHz
AC Current 1-Year Absolute Uncertainty±(0.04% of reading + 0.003% of range) at 1 kHz
AC Current Ranges29 µA to 329.999 µA, 330 µA to 3.29999 mA, 3.3 mA to 32.9999 mA, 33 mA to 329.999 mA, 330 mA to 2.99999 A, 3 A to 20.50000 A

Resistance

ParameterValue
Resistance Range0 Ω to 1.1 GΩ
Resistance StyleContinuously variable
Resistance 1-Year Absolute Uncertainty±28 ppm of reading on 10.9 kΩ range

Capacitance

ParameterValue
Capacitance Range220 pF to 110 mF
Capacitance 1-Year Absolute Uncertainty±(0.25% of reading + 10 nF) on 3.3 µF range

Temperature Calibration (Thermocouple)

ParameterValue
TC Source/Measure TypesB, C, E, J, K, L, N, R, S, T, U, BP, XK
TC Source Temperature Range-250 °C to +2316 °C (type-dependent)
TC Source Best Uncertainty±0.09 °C (Type K)

Temperature Calibration (RTD)

ParameterValue
RTD Source TypesPt385 (100 Ω, 200 Ω, 500 Ω, 1000 Ω), Pt3926, Pt3916, Ni120 (120 Ω Ni), Cu10
RTD Temperature Range-200 °C to +850 °C (type-dependent)
RTD Best Uncertainty±0.03 °C (Pt385, 100 Ω)

Phase & Power

ParameterValue
Phase Range-180.00° to +180.00°
Phase Resolution0.01°
Power Range (Simulated)Up to 20.9 kW (with voltage and current outputs simultaneously)
Power Frequency Range10 Hz to 30 kHz

Oscilloscope Calibration (Options)

ParameterValue
SC600 Option BandwidthUp to 600 MHz
SC1100 Option BandwidthUp to 1.1 GHz
Oscilloscope Calibration FunctionsAmplitude, edge, fast edge, leveled sine wave, time marker, wave generator

General Technical Specifications

ParameterValue
Warm-up Time30 minutes
Calibration Interval1 Year
Frequency Uncertainty±2.5 ppm
Reverse Power Protection (V-Guard)Up to 300 V peak on all voltage/current output terminals

Environmental

ParameterValue
Operating AltitudeUp to 3,050 m (10,000 ft)
Storage Temperature Range-20 °C to +70 °C
Relative Humidity (Operating)< 80% to 30 °C, < 70% to 40 °C, < 40% to 50 °C
Storage Altitude12,192 m (40,000 ft)

Accessories

ParameterValue
Compatible Current CoilFluke 5500A/COIL 50-Turn Current Coil (extends AC/DC current capability to 1025 A)

Frequency

ParameterValue
Frequency Source Range0.01 Hz to 2.0 MHz
Frequency Source Resolution11 digits

Power & Phase

ParameterValue
Power Factor Range0 to 1

AC Signal

ParameterValue
Harmonics GenerationUp to 50th harmonic

Temperature

ParameterValue
Thermocouple Resolution0.01 °C or 0.01 °F

Power Supply

ParameterValue
Line Frequency47 Hz to 63 Hz
Maximum Power Consumption600 VA

Oscilloscope Calibration

ParameterValue
SC300 Option Bandwidth300 MHz

Compliance

ParameterValue
EMC ComplianceIEC 61326-1

Auxiliary AC Voltage Output

ParameterValue
Auxiliary AC Voltage Range10 mV to 3.3 V
Auxiliary Voltage Phase Resolution0.01°

Thermocouple

ParameterValue
Thermocouple Types SupportedB, C, E, J, K, L, N, R, S, T, U, XK, BP

RTD

ParameterValue
RTD Types SupportedPt385 (10 Ω, 50 Ω, 100 Ω, 200 Ω, 500 Ω, 1000 Ω), Pt3926 (100 Ω), Pt3916 (100 Ω), Pt385 (10 Ω), Ni120 (120 Ω), Cu10 (10 Ω)

Technology

Electrical Calibration

Applications

Calibration of digital multimeters, oscilloscopes, power meters, and temperature indicators