The Fluke 5700A is a high-precision multifunction calibrator designed to calibrate digital multimeters, voltage standards, and other precision electrical instruments. Released as a standard-setting instrument in metrology, it features direct and alternating voltage, current, and resistance sourcing, and supports Artifact Calibration which minimizes the number of external standards required to maintain full accuracy.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488), RS-232-C
Sensing SelectionInternal (2-wire) or External (4-wire)

Physical

ParameterValue
Power Supply100 V / 120 V / 220 V / 240 V AC ±10%, 47 Hz to 63 Hz, 300 VA
Operating Temperature0 °C to 50 °C
Dimensions (W×H×D)432 mm × 178 mm × 630 mm
Weight27.2

DC Voltage Specifications

ParameterValue
DC Voltage Range Limits0 to ±1100 V
DC Voltage Range Steps220 mV, 2.2 V, 22 V, 220 V, 1100 V

AC Voltage Specifications

ParameterValue
AC Voltage Range Limits220 µV to 1100 V
AC Voltage Frequency Range10 Hz to 1 MHz

DC Current Specifications

ParameterValue
DC Current Range Limits0 to ±2.2 A (11 A with 5725A Amplifier)
DC Current Range Steps220 µA, 2.2 mA, 22 mA, 220 mA, 2.2 A

AC Current Specifications

ParameterValue
AC Current Range Limits9 µA to 2.2 A (11 A with 5725A Amplifier)
AC Current Frequency Range10 Hz to 10 kHz

Resistance Specifications

ParameterValue
Resistance Nominal Values1 Ω, 1.9 Ω, 10 Ω, 19 Ω, 100 Ω, 190 Ω, 1 kΩ, 1.9 kΩ, 10 kΩ, 19 kΩ, 100 kΩ, 190 kΩ, 1 MΩ, 1.9 MΩ, 10 MΩ, 19 MΩ, 100 MΩ

Wideband AC Voltage (Optional)

ParameterValue
Wideband AC Voltage Range Limits300 µV to 3.5 V (with Option 5700A-03)
Wideband AC Voltage Frequency Range10 Hz to 30 MHz (with Option 5700A-03)
Wideband AC Voltage Output Impedance50 Ω (with Option 5700A-03)

Reference and Calibration

ParameterValue
Artifact Calibration Standards Required10 V DC reference, 1 Ω resistance standard, 10 kΩ resistance standard
Artifact Calibration CompatibilityYes

Environmental and Standards Compliance

ParameterValue
Storage Temperature Range-40 °C to 75 °C
Relative Humidity (Non-condensing)<80% (up to 30 °C), <70% (up to 40 °C), <40% (up to 50 °C)
Operating Altitude0 to 3050 m (10,000 ft)

Power Requirements

ParameterValue
Line Voltage Selection100 V / 120 V / 220 V / 240 V AC (±10%, user-selectable)
Line Frequency Range47 Hz to 63 Hz
Maximum Power Consumption300 VA
Line Voltage Settings100 V / 120 V / 220 V / 240 V AC (±10%)

Options

ParameterValue
Wideband Output OptionOption 5700A-03 (Adds 10 Hz to 30 MHz output)

Interfaces & Software

ParameterValue
Remote Programming Emulation ModesFluke 5100B, Fluke 5200A

Technology

Solid-State Calibration Standard

Applications

Calibration of high-precision digital multimeters up to 7.5 digits