The Fluke 5790A is a highly accurate automated AC measurement standard designed for calibration and metrology applications. Combining the precision of a thermal transfer standard with the ease of use of a digital voltmeter, it provides exceptional AC voltage measurement accuracy, wideband capability, and robust system integration.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488.2), RS-232-C |
| Display | Vacuum Fluorescent Display (VFD) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 V / 120 V / 220 V / 240 V AC (±10%), 47 Hz to 63 Hz, 120 VA |
| Operating Temperature | 0 to 50 °C |
| Dimensions (W×H×D) | 432 mm × 178 mm × 473 mm |
| Weight | 24 kg (53 lb) |
AC Voltage Measurement Performance
| Parameter | Value |
|---|---|
| AC Voltage Ranges | 2.2 mV, 7 mV, 22 mV, 70 mV, 220 mV, 700 mV, 2.2 V, 7 V, 22 V, 70 V, 220 V, 700 V, 1000 V |
| Frequency Range | 10 Hz to 1 MHz |
| Best 24-Hour AC Voltage Uncertainty | ±24 ppm |
| Best 1-Year AC Voltage Uncertainty | ±110 ppm |
| Absolute Display Resolution | 6.5 digits |
| Inputs 1 & 2 Impedance | 1 MΩ shunted by 140 pF |
| Input 3 Impedance | 1 MΩ shunted by 150 pF |
| Maximum Input Voltage | 1000 V RMS |
| Maximum Peak Input Voltage | 1414 V Peak |
AC/DC Transfer Difference & Shunt Support
| Parameter | Value |
|---|---|
| AC/DC Transfer Resolution | 1 ppm (0.0001%) |
| AC/DC Transfer Repeatability | ±1 ppm |
| Compatible External Current Shunts | Fluke A40 and A40A series shunts |
| External Shunt Input Voltage Range | Up to 2.2 V (via Input 3 / Aux Input) |
Analog & Guard Connections
| Parameter | Value |
|---|---|
| Input 1 Connector Type | Type N female |
| Input 2 Connector Type | Type N female |
| Input 3 Connector Type | Dual binding posts |
| Guard Terminal Configurations | Binding post, switchable internally or externally |
| Isolation Input to Guard | >10^10 Ω, <150 pF |
| Isolation Guard to Chassis | >10^9 Ω, <0.1 µF |
| Chassis Ground Terminal | Rear-panel binding post |
Wideband Voltage Performance (Optional)
| Parameter | Value |
|---|---|
| Wideband Voltage Ranges | 2.2 mV, 7 mV, 22 mV, 70 mV, 220 mV, 700 mV, 2.2 V, 7 V (with Option 5790A-03) |
| Wideband Frequency Range | 10 Hz to 30 MHz (with Option 5790A-03) |
| Wideband Best 1-Year Uncertainty | ±0.25% to ±0.5% (frequency dependent) |
| Wideband Input Connector Type | Type N female (Input 4) |
| Wideband Input Impedance | 50 Ω |
Frequency Measurement
| Parameter | Value |
|---|---|
| Frequency Measurement Range | 10 Hz to 1 MHz (up to 30 MHz with Wideband Option) |
| Frequency Measurement Accuracy | ±0.01% (±100 ppm) |
| Frequency Measurement Resolution | 6 digits |
Metrological Reliability & Reference
| Parameter | Value |
|---|---|
| Internal Reference Type | Solid-state zener reference standard |
| Warm-up Time to Rated Accuracy | 30 minutes |
| Recommended Calibration Cycle | 1 year (or 90 days for higher accuracy requirements) |
Environmental Specifications
| Parameter | Value |
|---|---|
| Calibration Temperature Range | 18 °C to 28 °C |
| Storage Temperature Range | -40 °C to 75 °C |
| Relative Humidity Limits (Operating) | Max 80% non-condensing up to 30 °C, 75% up to 40 °C, 45% up to 50 °C |
| Operating Altitude | Up to 3,050 meters (10,000 feet) |
| Non-Operating Altitude | Up to 12,200 meters (40,000 feet) |
| Shock and Vibration Compliance | Complies with MIL-T-28800D Type III, Class 5, Style E |
Physical & Mechanical Characteristics
| Parameter | Value |
|---|---|
| Enclosure Style | Metal outer chassis with optional rack mount capability |
Compliance & Safety
| Parameter | Value |
|---|---|
| Safety Compliance | IEC 1010-1, ANSI/ISA-S82.01, CAN/CSA-C22.2 No. 1010.1-92 |
| Overvoltage Category | CAT II |
| EMC Compliance | EN 55011 Class A, EN 50082-1 |
Performance
| Parameter | Value |
|---|---|
| Crest Factor | 6:1 at full scale, up to 10:1 at downscale inputs |
Inputs
| Parameter | Value |
|---|---|
| Input 1 Capacitance | < 150 pF |
| Input 2 Capacitance | < 150 pF |
| Wideband Input Maximum Continuous Power | 1 W |
| Wideband Input Maximum Voltage | 7 V RMS (10 V peak) |
Outputs
| Parameter | Value |
|---|---|
| Analog Output Voltage Range | 0 V to 3 V DC (proportional to input) |
Power Requirements
| Parameter | Value |
|---|---|
| Power Consumption | 120 VA (maximum) |
| Line Frequency Range | 47 Hz to 63 Hz |
| Mains Voltage Selection | 90 V to 132 V AC / 180 V to 264 V AC, manually switchable |
Options
| Parameter | Value |
|---|---|
| Hardware Options | Option 5790A-03 Wideband AC Voltage (10 Hz to 30 MHz) |
Wideband Option (Option 5790A-03)
| Parameter | Value |
|---|---|
| Wideband Input VSWR | < 1.01 (10 Hz to 1 MHz), < 1.02 (1 MHz to 10 MHz), < 1.05 (10 MHz to 30 MHz) |
DC Voltage Capability
| Parameter | Value |
|---|---|
| DC Voltage Measurement Ranges | 2.2 mV to 1000 V (in 13 ranges) |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Total Number of Input Channels | 4 (Input 1, Input 2, Input 3, and Wideband Option Input) |
Measurement Technology
| Parameter | Value |
|---|---|
| RMS Sensor Type | Patented solid-state RMS sensor |
Technology
Automated AC/DC Thermal Transfer Standard
Applications
AC Voltage Calibration and Metrology Standard