5790A

FlukeAC

The Fluke 5790A is a highly accurate automated AC measurement standard designed for calibration and metrology applications. Combining the precision of a thermal transfer standard with the ease of use of a digital voltmeter, it provides exceptional AC voltage measurement accuracy, wideband capability, and robust system integration.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488.2), RS-232-C
DisplayVacuum Fluorescent Display (VFD)

Physical

ParameterValue
Power Supply100 V / 120 V / 220 V / 240 V AC (±10%), 47 Hz to 63 Hz, 120 VA
Operating Temperature0 to 50 °C
Dimensions (W×H×D)432 mm × 178 mm × 473 mm
Weight24 kg (53 lb)

AC Voltage Measurement Performance

ParameterValue
AC Voltage Ranges2.2 mV, 7 mV, 22 mV, 70 mV, 220 mV, 700 mV, 2.2 V, 7 V, 22 V, 70 V, 220 V, 700 V, 1000 V
Frequency Range10 Hz to 1 MHz
Best 24-Hour AC Voltage Uncertainty±24 ppm
Best 1-Year AC Voltage Uncertainty±110 ppm
Absolute Display Resolution6.5 digits
Inputs 1 & 2 Impedance1 MΩ shunted by 140 pF
Input 3 Impedance1 MΩ shunted by 150 pF
Maximum Input Voltage1000 V RMS
Maximum Peak Input Voltage1414 V Peak

AC/DC Transfer Difference & Shunt Support

ParameterValue
AC/DC Transfer Resolution1 ppm (0.0001%)
AC/DC Transfer Repeatability±1 ppm
Compatible External Current ShuntsFluke A40 and A40A series shunts
External Shunt Input Voltage RangeUp to 2.2 V (via Input 3 / Aux Input)

Analog & Guard Connections

ParameterValue
Input 1 Connector TypeType N female
Input 2 Connector TypeType N female
Input 3 Connector TypeDual binding posts
Guard Terminal ConfigurationsBinding post, switchable internally or externally
Isolation Input to Guard>10^10 Ω, <150 pF
Isolation Guard to Chassis>10^9 Ω, <0.1 µF
Chassis Ground TerminalRear-panel binding post

Wideband Voltage Performance (Optional)

ParameterValue
Wideband Voltage Ranges2.2 mV, 7 mV, 22 mV, 70 mV, 220 mV, 700 mV, 2.2 V, 7 V (with Option 5790A-03)
Wideband Frequency Range10 Hz to 30 MHz (with Option 5790A-03)
Wideband Best 1-Year Uncertainty±0.25% to ±0.5% (frequency dependent)
Wideband Input Connector TypeType N female (Input 4)
Wideband Input Impedance50 Ω

Frequency Measurement

ParameterValue
Frequency Measurement Range10 Hz to 1 MHz (up to 30 MHz with Wideband Option)
Frequency Measurement Accuracy±0.01% (±100 ppm)
Frequency Measurement Resolution6 digits

Metrological Reliability & Reference

ParameterValue
Internal Reference TypeSolid-state zener reference standard
Warm-up Time to Rated Accuracy30 minutes
Recommended Calibration Cycle1 year (or 90 days for higher accuracy requirements)

Environmental Specifications

ParameterValue
Calibration Temperature Range18 °C to 28 °C
Storage Temperature Range-40 °C to 75 °C
Relative Humidity Limits (Operating)Max 80% non-condensing up to 30 °C, 75% up to 40 °C, 45% up to 50 °C
Operating AltitudeUp to 3,050 meters (10,000 feet)
Non-Operating AltitudeUp to 12,200 meters (40,000 feet)
Shock and Vibration ComplianceComplies with MIL-T-28800D Type III, Class 5, Style E

Physical & Mechanical Characteristics

ParameterValue
Enclosure StyleMetal outer chassis with optional rack mount capability

Compliance & Safety

ParameterValue
Safety ComplianceIEC 1010-1, ANSI/ISA-S82.01, CAN/CSA-C22.2 No. 1010.1-92
Overvoltage CategoryCAT II
EMC ComplianceEN 55011 Class A, EN 50082-1

Performance

ParameterValue
Crest Factor6:1 at full scale, up to 10:1 at downscale inputs

Inputs

ParameterValue
Input 1 Capacitance< 150 pF
Input 2 Capacitance< 150 pF
Wideband Input Maximum Continuous Power1 W
Wideband Input Maximum Voltage7 V RMS (10 V peak)

Outputs

ParameterValue
Analog Output Voltage Range0 V to 3 V DC (proportional to input)

Power Requirements

ParameterValue
Power Consumption120 VA (maximum)
Line Frequency Range47 Hz to 63 Hz
Mains Voltage Selection90 V to 132 V AC / 180 V to 264 V AC, manually switchable

Options

ParameterValue
Hardware OptionsOption 5790A-03 Wideband AC Voltage (10 Hz to 30 MHz)

Wideband Option (Option 5790A-03)

ParameterValue
Wideband Input VSWR< 1.01 (10 Hz to 1 MHz), < 1.02 (1 MHz to 10 MHz), < 1.05 (10 MHz to 30 MHz)

DC Voltage Capability

ParameterValue
DC Voltage Measurement Ranges2.2 mV to 1000 V (in 13 ranges)

Inputs & Outputs

ParameterValue
Total Number of Input Channels4 (Input 1, Input 2, Input 3, and Wideband Option Input)

Measurement Technology

ParameterValue
RMS Sensor TypePatented solid-state RMS sensor

Technology

Automated AC/DC Thermal Transfer Standard

Applications

AC Voltage Calibration and Metrology Standard