DSX2-8000

Fluke*misc

The Fluke Networks DSX2-8000 CableAnalyzer is a premier copper cabling certification solution designed for installers and network owners. Part of the Versiv cabling certification product family, it provides rapid, precise testing up to Category 8 and Class I/II standards with Level 2G / Level VI accuracy. It features dual-band Wi-Fi, the LinkWare Live cloud testing workflow platform, and advanced diagnostics like High Definition Time Domain Crosstalk (HDTDX) and Reflectometry (HDTDR) to accelerate project execution and troubleshooting.

Specifications

General

ParameterValue
InterfacesUSB 2.0 Micro-B, USB 2.0 Type A, RJ-45 Ethernet (10/100 BASE-TX), Wi-Fi (IEEE 802.11 a/b/g/n dual-band)
Display5.7 in (14.5 cm) capacitive LCD touchscreen with 800 x 480 resolution

Physical

ParameterValue
Power Supply7.2 V Lithium-ion battery pack; External AC adapter: 100 V to 240 VAC, 50-60 Hz input, 15 VDC, 2 A output
Operating Temperature0 °C to 45 °C
Dimensions (W×H×D)133.3 mm × 66.7 mm × 279.4 mm (with DSX-8000 module and battery installed)
Weight1.28 kg

Core Signal and Electrical Performance

ParameterValue
Frequency Range1 MHz to 2000 MHz
ANSI/TIA Accuracy LevelLevel 2G (ANSI/TIA-1152-A)
IEC Accuracy LevelLevel VI (IEC 61935-1)
Supported Cable TypesShielded and unshielded twisted pair (UTP, ScTP, FTP, SSTP)
Input Overvoltage ProtectionProtected against continuous telco voltages and 100 mA over-current

Measurement Capabilities and Timing

ParameterValue
Autotest Speed - Category 6A8 seconds
Autotest Speed - Category 816 seconds
Maximum Cable Length (Propagation Delay)150 m (490 ft)
Length Resolution0.1 m
Propagation Delay Resolution1 ns
Delay Skew Resolution1 ns
DC Loop Resistance Range0 Ω to 540 Ω
DC Loop Resistance Resolution0.1 Ω
DC Resistance Unbalance Resolution0.1 Ω
NVP Range50% to 99.9%
Diagnostics / TroubleshootingHDTDX (High Definition Time Domain Crosstalk), HDTDR (High Definition Time Domain Reflectometry)

Connectivity, Ports, and Interfaces

ParameterValue
USB Ports1 x Type A (Host), 1 x Micro-B (Device)
Auxiliary Interface Ports3.5 mm stereo headset jack, RJ-45 Ethernet port

User Interface and Software Control

ParameterValue
Internal Storage Capacity (with graphs)5,000 results
Internal Storage Capacity (without graphs)12,000 results
Cloud IntegrationLinkWare Live cloud service integration via Wi-Fi or Ethernet
PC Analysis SoftwareLinkWare PC software
Supported LanguagesEnglish, French, German, Spanish, Portuguese, Italian, Japanese, Simplified Chinese, Traditional Chinese, Korean, Russian

Power and Electrical Specifications

ParameterValue
Battery ChemistryLithium-ion
Typical Battery Runtime8 hours
Battery Recharge Time4 hours (from 10% to 90% charge, with tester powered off)

Physical and Mechanical Attributes

ParameterValue
Enclosure DesignRugged overmold design for heavy-duty field environments

Environmental and Compliance Standards

ParameterValue
Storage Temperature Range-20 °C to +60 °C
Relative Humidity Range0% to 90% (non-condensing)
Maximum Operating Altitude4,000 m (13,123 ft)
Maximum Storage Altitude4,600 m (15,000 ft)
Calibration Interval1 year (recommended)
Vibration Standard ComplianceMeets MIL-PRF-28800F Class 3 requirements

Configurations, Options, and Accessories

ParameterValue
TIA Cabling StandardsCategory 3, 4, 5, 5e, 6, 6A, 8 per TIA-568.2-D
ISO/IEC Cabling StandardsClass C, D, E, Ea, F, Fa, I/II per ISO/IEC 11801

Measurements

ParameterValue
DC Loop Resistance Accuracy± (1 Ω + 1 %)
DC Resistance Unbalance Accuracy± (0.15 Ω + 1 %)
Shield Continuity Test MethodPatented AC measurement (localizes shield faults with distance)
TCL and ELTCTL Measurement SupportUp to 2000 MHz
CDNEXT and CMACRF Measurement SupportUp to 2000 MHz
Resistance Unbalance CapabilityMeasures within-pair and pair-to-pair DC resistance unbalance

Wireless Connectivity

ParameterValue
Integrated Wi-FiIEEE 802.11 a/b/g/n (2.4 GHz and 5 GHz)

Power

ParameterValue
AC Charger Input100 V AC to 240 V AC, 50 Hz to 60 Hz
AC Charger Output15 V DC, 2 A

Environmental

ParameterValue
Pollution Degree RatingPollution Degree 2
Shock and Vibration standardMeets MIL-PRF-28800F Class 3 requirements

Performance

ParameterValue
Autotest Speed - Category 5e/67 seconds

Connectivity & Communication

ParameterValue
Talkset InterfaceFull-duplex digital audio over twisted pair

Measurement Capabilities

ParameterValue
Supported Coaxial Impedance50 Ω, 75 Ω, 93 Ω (with optional adapters)
Propagation Delay Range0 ns to 4000 ns (dual-ended) / 0 ns to 4000 ns (single-ended)
Delay Skew Range0 ns to 125 ns

Display

ParameterValue
Display Resolution800 × 480 pixels

Compliance & Safety

ParameterValue
Safety ComplianceIEC 61010-1: Overvoltage Category None, Pollution Degree 2; IEC 61010-30: 30 V rms max
EMC ComplianceIEC 61326-1: Portable; CISPR 11: Group 1, Class A; FCC Part 15 Class A

Power & Battery

ParameterValue
Battery Charging Temperature Range0 °C to 40 °C

Physical & Durability

ParameterValue
Adapter Connector LifeTypical > 5,000 cycles

Technology

Copper Twisted Pair Cabling Certification

Applications

Category 5e/6/6A/8 and Class C/D/E/Ea/F/Fa/I/II cabling certification