The Giga-tronics 6062A is a synthesized RF signal generator covering the 100 kHz to 2.1 GHz frequency range. Developed originally under the Fluke brand and subsequently supported by Giga-tronics, it delivers high spectral purity, indirect synthesis, and a full suite of modulation capabilities including AM, FM, phase modulation, and fast-rise pulse modulation as a standard feature, making it highly suited for pulsed communications, radar receiver testing, and automated test environments.

Specifications

Frequency

ParameterValue
Minimum Frequency100 kHz
Maximum Frequency2.1 GHz
Frequency Resolution10 Hz (0.1 to 1050 MHz), 20 Hz (1050 to 2100 MHz)

Output

ParameterValue
Maximum Output Power+13 dBm
Output Power Range-137 dBm to +13 dBm
Phase Noise<-115 dBc/Hz @ 20 kHz offset, 1 GHz carrier

Modulation

ParameterValue
ModulationAM, FM, Phase Modulation (PM), Pulse Modulation
Sweep ModesFrequency Sweep, Amplitude (Step) Sweep
Standard Internal Modulation Frequencies400 Hz and 1000 Hz
Option 130 Modulation Frequency Range10 Hz to 99.9 kHz (option dependent)
Option 130 Modulation Frequency Resolution10 Hz (10 Hz to 9.99 kHz), 100 Hz (10.0 kHz to 99.9 kHz) (option dependent)
AM Resolution1%

General

ParameterValue
Reference Oscillator10 MHz Internal Crystal
InterfacesIEEE-488 (GPIB)
DisplayLCD

Physical

ParameterValue
Power Supply100 V / 120 V / 220 V / 240 V AC (±10%), 47 Hz to 63 Hz
Dimensions (W×H×D)432 mm x 133 mm x 533 mm
Weight14.5 kg

Frequency Specifications

ParameterValue
Frequency Switching Speed< 100 ms
Internal Reference Aging Rate< 1.5 x 10^-8 per day
External Reference Input Frequency1, 2, 5, or 10 MHz
External Reference Input Amplitude0.2 V to 5 V p-p

Amplitude Specifications

ParameterValue
Minimum Output Power-137 dBm
Amplitude Resolution0.1 dB
Amplitude Accuracy±1.5 dB (levels above -127 dBm)
Output Impedance50 Ohms
Output Voltage SWR< 1.5:1 (for output levels < 0 dBm)
Maximum Reverse Power ProtectionUp to 50 W (from 50 Ohm source)

Spectral Purity

ParameterValue
Harmonics<-30 dBc (levels < +13 dBm)
Non-Harmonics (Spurious)<-60 dBc (at carrier frequencies up to 1050 MHz)
Sub-Harmonics<-50 dBc (above 1050 MHz)
Residual FM< 12 Hz RMS (0.3 to 3 kHz BW, @ 1 GHz)

Analog Modulation

ParameterValue
FM Maximum Deviation800 kHz (for carriers 1050 to 2100 MHz)
FM Rate Range20 Hz to 100 kHz
FM Distortion< 1% (at 1 kHz rate)
AM Depth Range0% to 99%
AM Rate Range20 Hz to 50 kHz
AM Distortion< 1.5% (at 30% depth, 1 kHz rate)
PM Maximum Deviation10 radians (for carriers 1050 to 2100 MHz)
PM Rate Range20 Hz to 10 kHz
Pulse Modulation On/Off Ratio> 80 dB
Pulse Modulation Rise/Fall Time< 15 ns (typically < 10 ns)
Pulse Modulation Minimum Pulse Width50 ns

System Connectivity and Interfaces

ParameterValue
RF Output ConnectorType N female
External Reference Input ConnectorBNC female
Memory RegistersUp to 50 non-volatile instrument set-ups

Physical and Environmental Specifications

ParameterValue
Operating Temperature Range0 to 50 °C
Storage Temperature Range-40 to 71 °C
Rack Unit (RU) Height3U

Input & Output

ParameterValue
External Modulation Input ConnectorBNC female (front panel)
External Pulse Input ConnectorBNC female (front panel)
Reference Output ConnectorBNC female (rear panel)
Sweep Output ConnectorBNC female (rear panel)

Performance

ParameterValue
Sweep Output Voltage Range0 to 10 V
Warm-up Time30 minutes

Power

ParameterValue
Power Consumption180 VA max
AC Input Voltages100 V / 120 V / 220 V / 240 V AC (±10%)
AC Input Frequency47 Hz to 440 Hz

Compliance & Standards

ParameterValue
RF Electromagnetic CompatibilityMeets MIL-STD-461B

Environmental

ParameterValue
Relative Humidity95% RH max at +30 °C, 75% RH max at +40 °C, 45% RH max at +50 °C
Operating Altitude3050 m (10,000 ft)
Non-Operating Altitude12200 m (40,000 ft)

Compliance

ParameterValue
Vibration and Shock ComplianceMeets MIL-T-28800C Class 5
Safety ComplianceDesigned to comply with IEC 348 and UL 1244

Reference

ParameterValue
Reference Output Frequency10 MHz

Interfaces

ParameterValue
GPIB IEEE-488 Capability SubsetSH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, E1

Technology

Indirect Synthesis

Applications

Receiver testing, pulsed communication, radar system testing, ATE applications