The Keysight (Agilent) 11613B is a dedicated transfer standard specifically designed for the calibration of Agilent 8757D and 8757E scalar network analyzers. It features a precise 27.778 kHz source, an array of precision attenuators, and dedicated verification and adjustment software designed to update internal calibration parameters directly within the analyzer's nonvolatile memory.

Specifications

General

ParameterValue
Interfaces9-pin ADC interface

General & Metrological Overview

ParameterValue
Calibrator TypeDedicated transfer standard
Compatible AnalyzersAgilent/Keysight 8757D, Agilent/Keysight 8757E
Connector Types9-pin ADC interface, four 5-pin detector connectors
Internal Components27.778 kHz source, precision attenuators
Verification SoftwareIncluded (verifies and adjusts internal analyzer calibration parameters)

Output Specifications

ParameterValue
Signal Symmetry50 ±2% duty cycle
Signal Frequency27,778 ±12 Hz
Peak-to-Peak Voltage10.6 ±0.5 V
Rise Time (10% to 90%)3.4 ±0.4 µs
OvershootNegligible (with high impedance load)

Technology

Analog Calibration Standard

Applications

Calibration of Agilent 8757D/E Scalar Network Analyzers