The Keysight 11713C Attenuator/Switch Driver provides GPIB, USB, and LXI/LAN control for up to four programmable attenuators and four coaxial switches (or up to 20 SPDT switches). It is designed to simplify the integration of electromechanical switches and programmable attenuators into automated RF and microwave test systems, offering selectable drive voltages and backward compatibility with the legacy HP 11713A.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488), LAN (10/100 Base-T, LXI Class C), USB 2.0 (Type-B) |
| Display | Monochrome alphanumeric LCD display |
| Cooling | Convection (fanless) |
| Number of Save/Recall Registers | 10 (registers 0 to 9) |
| Non-Volatile State Storage | Yes (up to 10 instrument states) |
| Power Status Indicator LED | Amber (Standby) / Green (On) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 to 240 VAC (±10%), 50 to 60 Hz, 100 VA maximum |
| Operating Temperature | 0 °C to 50 °C |
| Dimensions (W×H×D) | 212.6 mm × 88.0 mm × 348.3 mm |
| Weight | 3.2 kg |
Drive Capability & Control Specifications
| Parameter | Value |
|---|---|
| Number of Attenuator Control Ports | 2 independent drive ports (ATTEN X, Y) |
| Number of Switch Control Ports | 2 independent drive ports (SWITCH A, B) |
| Maximum Number of SPDT Switches Driven | Up to 20 SPDT switches |
| Maximum Number of Programmable Attenuators Driven | Up to 4 programmable attenuators |
| Supported Drive Voltages | 5 V, 15 V, 24 V, and OFF (user selectable per bank) |
| Maximum Continuous Total Current | 1.7 A (total across all channels) |
| Drive Pulse Width Range | 15 ms to 2.0 s (programmable in 1 ms steps) |
| Compatible Switch/Attenuator Technology | Electromechanical |
Connectivity & Remote Control
| Parameter | Value |
|---|---|
| GPIB Interface Standard | IEEE-488.2 |
| LAN Interface Compliance | 10/100 Base-T, LXI Class C compliant |
| USB Interface Type | USB 2.0 High Speed (Type-B) |
| Command Language Support | SCPI (Standard Commands for Programmable Instruments) |
| Legacy Code Emulation Mode | HP 11713A command set emulation |
| Web User Interface | Built-in web server for remote control and configuration via web browser |
Front Panel Controls & Interfaces
| Parameter | Value |
|---|---|
| Manual Channel Control | 20-key front panel local keypad for manual state toggling |
| Local/Remote Operation Key | Dedicated 'Local' key to return instrument to manual control |
| Power Switch Type | Front panel standby/on button |
Rear Panel Connectors & Interface Hardware
| Parameter | Value |
|---|---|
| Drive Output Option 101 Connectors | Viking 12-pin (Qty 4) |
| Drive Output Option 201 Connectors | D-sub 15-pin (Qty 4) |
| Drive Output Option 301 Connectors | Viking 12-pin (Qty 2) and D-sub 15-pin (Qty 2) |
| Chassis Ground Terminal | Rear panel binding post |
Power & Electrical Specifications
| Parameter | Value |
|---|---|
| AC Line Voltage Input Range | 100 to 240 VAC (±10%) |
| AC Line Frequency Range | 50 to 60 Hz |
| Maximum Power Consumption | 100 VA |
Physical and Environmental Specifications
| Parameter | Value |
|---|---|
| Rack Space Requirements | 2U height, half-rack width |
| Dimensions with Bumpers (W×H×D) | 261.3 mm × 103.5 mm × 378.3 mm |
| Dimensions without Bumpers (W×H×D) | 212.6 mm × 88.0 mm × 348.3 mm |
| Net Weight | 3.2 kg (7.1 lbs) |
| Shipping Weight | 6.7 kg (14.8 lbs) |
| Storage Temperature Range | -40 °C to +70 °C |
| Operating Humidity Range | Up to 95% relative humidity (RH) at 40 °C (non-condensing) |
| Operating Altitude Range | Up to 4,600 meters (15,000 feet) |
| Storage Altitude Range | Up to 15,300 meters (50,000 feet) |
Regulatory and Compliance
| Parameter | Value |
|---|---|
| Product Safety Standards | IEC 61010-1:2001 / EN 61010-1:2001 |
| EMC Emissions Standard | CISPR 11, Group 1, Class A / EN 55011 |
| EMC Immunity Standard | IEC 61326-1 / EN 61326-1 |
Compatible Models & Accessories
| Parameter | Value |
|---|---|
| Supported Attenuators (849x Series) | Keysight 8494G/H, 8495G/H, 8496G/H, 8497K |
| Supported Attenuators (8490x Series) | Keysight 84904K/L/M, 84906K/L/M, 84907K/L/M, 84908M |
| Supported Switches (N181x Series) | Keysight N1810TL, N1811TL, N1812UL |
| Supported Coaxial SPDT Switches (876x Series) | Keysight 8761A/B, 8762A/B/C/F, 8765A/B/C/D/F |
| Supported Transfer Switches | Keysight 8763A/B/C, 8764A/B/C |
| Supported Multiport Switches | Keysight 8766K, 8767K, 8768K, 8769K, U3020AS/AS02 |
| Compatible Drive Cable Model (Viking to Viking) | 25-pin D-sub to dual-Viking |
| Compatible Drive Cable Model (Viking to D-sub) | 25-pin D-sub to dual 9-pin D-sub |
| Compatible Drive Cable Model (Viking to Bare Wire) | 25-pin D-sub to 10-pair bare wire |
| Compatible Drive Cable Model (D-sub to D-sub) | Keysight 11713-60045 |
| Compatible Drive Cable Model (D-sub to Viking) | Keysight 11713-60046 |
| Compatible Drive Cable Model (D-sub to Bare Wire) | Keysight 11713-60047 |
Performance
| Parameter | Value |
|---|---|
| Peak Drive Current | 1.7 A (for ≤ 1 s) |
| Pulse Width Resolution | 1 ms |
| Default Pulse Width | 15 ms |
| Drive Voltage Tolerance | ±10% |
| Command Processing Time (GPIB) | < 1.0 ms (typical) |
| Command Processing Time (USB) | < 1.5 ms (typical) |
| Command Processing Time (LAN) | < 2.0 ms (typical) |
Protection Features
| Parameter | Value |
|---|---|
| Overcurrent Protection | Electronic current limiting |
Connectivity & Control
| Parameter | Value |
|---|---|
| LAN Interface Speed | 10/100 Base-T |
| IP Address Configuration | DHCP, Auto-IP, Static IP |
| Software Drivers | IVI-COM, IVI-C, LabVIEW |
| LXI Compliance | LXI Class C |
Regulatory & Compliance
| Parameter | Value |
|---|---|
| Overvoltage Category | CAT II |
| Pollution Degree | Pollution Degree 2 |
Drive Outputs
| Parameter | Value |
|---|---|
| Number of Control Banks | 2 (Bank A and Bank B) |
| Total Drive Lines | 20 (10 per bank) |
| Drive Voltage Selection Method | Rear-panel 3-position slide switch |
| Drive Voltage Configuration | Independent selection for each bank |
Connectivity
| Parameter | Value |
|---|---|
| USB Device Port Connector | Type-B (rear panel) |
| LAN Port Connector | RJ-45 (rear panel) |
| GPIB Port Connector | 24-pin IEEE-488 (rear panel) |
| LXI Class | LXI Class C (LXI Core 2011) |
Technology
Electromechanical Switch and Attenuator Driver
Applications
RF and Microwave Automated Test Systems