The HP 11848A is the dedicated phase noise interface for the HP 3048A phase noise measurement system. It houses phase detectors, amplifiers, filters, and signal-routing switches, supporting several measurement techniques for both phase noise and AM noise across a 5 MHz to 18 GHz frequency range.

Specifications

General

ParameterValue
InterfacesHP-IB (IEEE-488)

Physical

ParameterValue
Power Supply100 / 120 / 220 / 240 VAC (+5%/-10%), 48 to 66 Hz, 120 VA max
Operating Temperature0 to 55 °C
Dimensions (W×H×D)425.5 mm × 132.6 mm × 514 mm
Weight15.5 kg

Compatibility & System Integration

ParameterValue
System CompatibilityHP 3048A Phase Noise Measurement System
Measurement SupportPhase noise and AM noise measurements

Signal & Performance Characteristics

ParameterValue
Standard Input Frequency Range5 MHz to 1.6 GHz
Extended Input Frequency Range1.2 GHz to 18 GHz (requires Option 201)
Phase Noise Offset Range0.01 Hz to 40 MHz
Left Port RF Input Amplitude (5 MHz - 1.6 GHz)+15 dBm to +23 dBm
Right Port RF Input Amplitude (5 MHz - 1.6 GHz)0 dBm to +23 dBm
Left Port Microwave Input Amplitude (1.2 GHz - 18 GHz, Option 201)+7 dBm to +10 dBm
Right Port Microwave Input Amplitude (1.2 GHz - 18 GHz, Option 201)0 dBm to +10 dBm
External Phase Detector InputProvided for measurements outside the internal frequency range
Error DetectionDetectors that sense erroneous measurement conditions
Signal RoutingNumerous internal signal-routing switches
Loop Control NetworksProcess phase detector output for voltage control of the input sources
Noise Processing NetworksProcess phase noise for measurement by a spectrum analyzer

Internal Calibration Sources

ParameterValue
Calibration Source 10 MHz A Amplitude+15 dBm
Calibration Source 10 MHz A Tuning±100 Hz
Calibration Source 10 MHz B Amplitude+6 dBm
Calibration Source 10 MHz B Tuning±1 kHz
Calibration Source 350-500 MHz Amplitude+17 dBm
Calibration Source 350-500 MHz Tuning±20 MHz
Calibration Source 400 MHz Amplitude-5 dBm
Calibration Source 400 MHz TuningFixed
Internal Calibration VerificationsFour built-in RF sources for functional checks of signal-handling circuits

System Options

ParameterValue
Option 201High frequency phase detector (1.2 GHz to 18 GHz)
Option 907Front-handle kit
Option 908Rack-flange kit
Option 909Rack-flange and front-handle combination kit

Environmental

ParameterValue
Storage Temperature-55 to +75 °C

LNA Performance

ParameterValue
LNA Frequency Range0.1 Hz to 45 MHz
LNA Gain Settings20 dB / 40 dB
LNA Input Noise Density (10 Hz)< 10 nV/√Hz
LNA Input Noise Density (100 Hz)< 4 nV/√Hz
LNA Input Noise Density (1 kHz)< 2 nV/√Hz
LNA Input Noise Density (10 kHz - 45 MHz)< 1.4 nV/√Hz

RF Inputs

ParameterValue
Maximum Safe RF Input Level+23 dBm

Tuning Output

ParameterValue
Tuning Voltage Output Range-10 V to +10 V

Applications

Phase noise and AM noise measurement

Looking for the Keysight Technologies (Agilent HP) 11848A?

Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.

Selling this kind of equipment? Sell with us