The Keysight Agilent 11970U is a high-performance waveguide harmonic mixer designed to extend the frequency range of spectrum analyzers to the U-band (40 to 60 GHz). It employs a dual-diode design that achieves exceptionally flat frequency response and low conversion loss without requiring external DC bias or tuning stubs. This makes both manual and automated millimeter-wave measurements simple, precise, and highly reliable.

Specifications

Physical

ParameterValue
Power SupplyNone (passive device, no external DC bias required)
Operating Temperature0 to 55 °C
Dimensions (W×H×D)70 mm × 47 mm × 103 mm
Weight0.14 kg
Shipping Weight0.9 kg

Frequency & Band Specifications

ParameterValue
Waveguide Band DesignationU Band
RF Input Frequency Range40 to 60 GHz
LO Input Frequency Range3.0 to 6.1 GHz
IF Output Frequency RangeDC to 1300 MHz
Harmonic Number (N)10-
Waveguide DesignationWR-19

Electrical & Performance Parameters

ParameterValue
Conversion Loss (Maximum)28 dB
Conversion Loss Flatness Across Band±1.9 dB
Maximum RF Input Power (CW Damage Level)+20 dBm (100 mW)
Maximum Peak Pulse Power+24 dBm (250 mW) with pulse width <1 µs
Optimum LO Drive Level+14 to +18 dBm

Port Impedance & VSWR

ParameterValue
Typical RF Input SWR<2.2:1
Nominal Port Impedance50 Ω (LO and IF ports)

Physical Interfaces & Connectors

ParameterValue
RF Input Connector TypeWR-19 Waveguide
Waveguide Flange TypeUG-383/U-M circular flange
LO Input Connector TypeSMA female
IF Output Connector TypeSMA female

DC Bias & Control Requirements

ParameterValue
External DC Bias RequirementsNone (unbiased dual-diode design)
Tuning AdjustmentsNone required (no tuning stubs)

Environmental & Reliability

ParameterValue
Storage Temperature Range-40 to +75 °C

Compliance & Quality Standards

ParameterValue
Calibration TraceabilityIndividually calibrated conversion loss chart attached to housing

General & Compatibility

ParameterValue
Compatible Spectrum AnalyzersKeysight E4440A/46A/48A PSA, E4407B ESA-E, 8560 E-series, 8566B (with 11975A amplifier), 70000 MMS (with 70907A/B interface)

Performance

ParameterValue
Typical Calibration Uncertainty±2.0 dB
Calibration Step Size1 GHz
Maximum LO Input Power+20 dBm

Environmental

ParameterValue
Environmental StandardsMIL-T-28800C, Type III, Class 3, Style C (except operating temperature)

Compliance

ParameterValue
MIL-T-28800C ComplianceType III, Class 3, Style C

Calibration

ParameterValue
Recommended Calibration Cycle1 year

Technology

Harmonic Mixer

Applications

Millimeter-wave spectrum analysis and frequency extension

Looking for the Keysight Technologies (Agilent HP) 11970U?

Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.

Selling this kind of equipment? Sell with us