The Keysight (Agilent HP) 3335A is a high-precision Synthesizer/Level Generator designed for telecommunications testing, calibration, and precise R&D applications. Operating across a frequency range of 200 Hz to 81 MHz, it is celebrated for its unmatched level flatness of ±0.05 dB, superb frequency resolution, and standard HP-IB system compatibility.

Specifications

General

ParameterValue
InterfacesHP-IB (IEEE-488)
Display7-segment red LED digital display for frequency, amplitude, and parameters
Warm-up Time20 minutes (standard), 24 hours (Option 001)

Physical

ParameterValue
Operating Temperature0 to 55 °C
Dimensions (W×H×D)425.5 mm × 132.6 mm × 497.8 mm
Weight19 kg

Frequency Specifications

ParameterValue
Frequency Range200 Hz to 80.999 999 999 MHz
Frequency Resolution0.001 Hz (1 mHz)
Frequency Reference Aging Rate< 1 × 10^-8 per day (standard); < 5 × 10^-10 per day (with high-stability Option 001)
Phase Offset AdjustabilityAdjustable in 0.1° increments

Amplitude and Level Control

ParameterValue
Output Level Range (50 Ω)-86.98 dBm to +13.01 dBm
Output Level Range (75 Ω)-86.96 dBm to +13.01 dBm
Amplitude Resolution0.01 dB
Output ImpedanceDual front panel outputs: 50 Ω and 75 Ω (individually selectable)

Spectral Purity

ParameterValue
Harmonics< -40 dBc (10 MHz to 80 MHz); < -45 dBc (200 Hz to 10 MHz)
Non-Harmonic Spurious< -70 dBc
SSB Phase Noise (10 kHz offset, 80 MHz carrier)<-105 dBc/Hz
SSB Phase Noise (10 kHz offset, 10 MHz carrier)<-112 dBc/Hz

Sweep and List Modes

ParameterValue
Sweep TypesSingle, Auto, and Manual step sweep

Inputs and Outputs

ParameterValue
RF Output Connector TypeDual BNC (one 50 Ω, one 75 Ω) on front panel
External Reference Input FrequencyAccepts 1, 2, 5, or 10 MHz references (internal system locks to 40 MHz)
Internal Reference Output Frequency10 MHz

Physical Characteristics

ParameterValue
Form FactorStandard HP full-width benchtop instrument with rackmount capability
Rack Unit Height3U

Performance

ParameterValue
Amplitude Flatness (50 Ω)±0.15 dB (10 kHz to 80 MHz, relative to 100 kHz)
Amplitude Flatness (75 Ω)±0.15 dB (10 kHz to 80 MHz, relative to 100 kHz)
Phase Increment0.1°
Option 002 Balanced Output Impedances124 Ω, 135 Ω, 150 Ω, 600 Ω
Option 002 Balanced Output Frequency Ranges124 Ω: 10 kHz to 10 MHz; 135 Ω & 150 Ω: 10 kHz to 2 MHz; 600 Ω: 10 kHz to 150 kHz
Frequency Reference Temperature Stability (Standard)< ±1 × 10^-6/°C
Frequency Reference Temperature Stability (Option 001)±2 × 10^-10/°C (0 °C to 55 °C)

Environmental

ParameterValue
Storage Temperature Range-40 °C to +75 °C

Inputs & Outputs

ParameterValue
Rear Panel Analog Sweep OutputsX-Drive (0 to 10 V) and Z-Blanking/Pen Lift

Frequency Reference

ParameterValue
Option 001 Reference Aging Rate< 5 × 10^-10 per day (after 10 days warm-up)
External Reference Input Level0 dBm to +15 dBm
External Reference Input Impedance50 Ω
Internal Reference Output Level (1 MHz / 10 MHz)≥ 0 dBm into 50 Ω

RF Output Characteristics

ParameterValue
Return Loss (50/75 Ω Unbalanced, 200 Hz - 10 MHz)≥ 30 dB
Return Loss (50/75 Ω Unbalanced, 10 MHz - 50 MHz)≥ 26 dB
Return Loss (50/75 Ω Unbalanced, 50 MHz - 80 MHz)≥ 23 dB (50 MHz - 81 MHz)

Sweep Characteristics

ParameterValue
Sweep Steps Range10 to 999 steps
Sweep Time per Step Range10 ms to 99.9 s

Option 002 Balanced Output

ParameterValue
Option 002 Balanced Output Level Range (124/135/150 Ω)-86.98 dBm to +13.01 dBm
Option 002 Balanced Output Level Range (600 Ω)-83.98 dBm to +16.01 dBm
Option 002 Balanced Output Flatness (124/135/150 Ω)±0.15 dB (10 kHz to 2 MHz)
Option 002 Balanced Output Flatness (600 Ω)±0.15 dB (50 Hz to 50 kHz)
Option 002 Balance (124/135/150 Ω)> 40 dB (10 kHz to 2 MHz)
Option 002 Balance (600 Ω)> 40 dB (50 Hz to 50 kHz)

Technology

Fractional-N Synthesis

Applications

Telecommunications testing, calibration, precise level-selective measurements