The Keysight Technologies (formerly Agilent) 34410A is a high-performance 6½ digit digital multimeter designed for rapid triggering and high-throughput system applications. Building on the industry-standard 34401A, the 34410A delivers up to 10,000 readings per second at 5½ digits, features standard USB, LAN, and GPIB connectivity, and supports advanced dual-display measurements, temperature, and capacitance testing.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488.2), USB 2.0 (USB-TMC), LAN (10/100 Base-T, LXI Class C compliant) |
| Display | Dual line vacuum fluorescent display (VFD) |
| Warm-up Time | 1 hour to full specifications |
| Real-Time Clock | Battery-backed real-time clock |
Physical
| Parameter | Value |
|---|---|
| Operating Temperature | 0 °C to 55 °C |
| Weight | 3.72 kg |
DC Voltage Specifications
| Parameter | Value |
|---|---|
| DC Voltage Ranges | 100 mV, 1 V, 10 V, 100 V, 1000 V |
| DC Voltage Resolution | 100 nV (on 100 mV range) |
| Input Resistance | 10 MΩ ±1% or >10 GΩ (selectable for 100 mV, 1 V, 10 V ranges) |
| Best 1-Year DC Voltage Accuracy | ±(0.0030% of reading + 0.0005% of range) |
| Temperature Coefficient | (0.0005% of reading + 0.0001% of range) per °C (outside 18 °C - 28 °C) |
| Normal Mode Rejection Ratio (NMRR) | 60 dB for 50/60 Hz (with aperture >= 1 PLC) |
| Common Mode Rejection Ratio (CMRR) | 140 dB for DC (with 1 kΩ LO lead unbalance) |
AC Voltage Specifications
| Parameter | Value |
|---|---|
| AC Voltage Ranges | 100 mV, 1 V, 10 V, 100 V, 750 V |
| AC Voltage Resolution | 100 nV (on 100 mV range) |
| Measurement Method | AC-coupled True RMS |
| AC Frequency Bandwidth | 3 Hz to 300 kHz |
Resistance Specifications
| Parameter | Value |
|---|---|
| Resistance Ranges | 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ, 1 GΩ |
| Resistance Resolution | 100 µΩ (on 100 Ω range) |
| Input Protection | 1000 V on all ranges |
| Best 1-Year Resistance Accuracy | ±(0.010% of reading + 0.001% of range) |
DC & AC Current Specifications
| Parameter | Value |
|---|---|
| DC Current Ranges | 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A |
| AC Current Ranges | 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A |
| Current Resolution | 100 pA (on 100 µA range) |
| Best 1-Year DC Current Accuracy | ±(0.05% of reading + 0.005% of range) |
Frequency, Period & Diode/Continuity
| Parameter | Value |
|---|---|
| Frequency Range | 3 Hz to 300 kHz |
| Period Range | 3.33 µs to 333 ms |
| Diode Test Current | 1 mA (fixed) |
| Diode Open Circuit Voltage Limit | 5 V |
| Continuity Threshold | 1 Ω to 1000 Ω (adjustable) |
Capacitance & Temperature Specifications
| Parameter | Value |
|---|---|
| Capacitance Ranges | 1 nF, 10 nF, 100 nF, 1 µF, 10 µF, 100 µF, 1000 µF, 10,000 µF |
| Capacitance Best 1-Year Accuracy | ±(1.0% of reading + 0.1% of range) |
| Supported Temperature Sensors | RTD (PT100) and Thermistor (2.2 kΩ, 5 kΩ, 10 kΩ) |
System Speeds & Digitizing
| Parameter | Value |
|---|---|
| Maximum Reading Rate (to Memory) | 10,000 readings/s at 5½ digits |
| Maximum Reading Rate (to Bus) | 10,000 readings/s at 5½ digits |
| Aperture / Integration Time Options | 0.006 PLC to 100 PLC |
| Autozero Settings | Off, On, Once |
| Internal Reading Memory Buffer | 50,000 readings |
Connectivity & Software Integration
| Parameter | Value |
|---|---|
| Web Browser User Interface | Yes, built-in LAN web interface |
Physical and Power Specifications
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to 70 °C |
| Safety Certifications | IEC 61010-1, EN 61010-1, UL 61010-1, CAN/CSA-C22.2 No. 61010-1 |
DC Voltage
| Parameter | Value |
|---|---|
| DC Input Bias Current | < 30 pA at 25 °C |
| Maximum DC Input Voltage | 1000 V on all ranges |
AC Voltage
| Parameter | Value |
|---|---|
| Maximum AC Input Voltage | 750 V rms (1000 V peak) |
| AC Bandwidth Filter Settings | 3 Hz, 20 Hz, 200 Hz |
Resistance
| Parameter | Value |
|---|---|
| Offset Compensated Ohms (OCO) | Available on 100 Ω, 1 kΩ, and 10 kΩ ranges |
System & Connectivity
| Parameter | Value |
|---|---|
| LXI Compliance | LXI Class C, Version 1.0 |
| Emulation Modes | Agilent 34401A SCPI compatibility |
Power
| Parameter | Value |
|---|---|
| Power Line Frequency | 45 Hz to 66 Hz and 360 Hz to 440 Hz (automatically sensed) |
Environmental
| Parameter | Value |
|---|---|
| Pollution Degree | Pollution Degree 2 |
Measurement Technology
| Parameter | Value |
|---|---|
| Frequency Measurement Method | Reciprocal counting technique |
Resistance Measurement
| Parameter | Value |
|---|---|
| Maximum 4-Wire Lead Resistance | 10% of range per lead (100 Ω, 1 kΩ ranges), 1 kΩ per lead (other ranges) |
Front / Rear Panel Features
| Parameter | Value |
|---|---|
| Input Terminals Selection | Front / Rear switchable from front panel |
Software & Connectivity
| Parameter | Value |
|---|---|
| USB Connection Class | USB-TMC |
| Ethernet Protocols | VXI-11, Sockets |
| LAN Speed | 10/100 Base-T |
AC Current
| Parameter | Value |
|---|---|
| AC Current Bandwidth | 3 Hz - 10 kHz |