The Keysight (Agilent/HP) 3458A is an industry-standard 8½-digit precision benchtop digital multimeter, renowned for its exceptional accuracy, speed, and flexibility. Used extensively in calibration laboratories and high-throughput production test environments, it offers measurement speeds up to 100,000 readings per second (at 4.5 digits) and an outstanding DC voltage transfer accuracy of 0.05 ppm.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488)
Warm-up Time4 hours (to meet full specifications)

Physical

ParameterValue
Power Supply100 / 120 / 220 / 240 V AC ±10%, 48-66 Hz or 360-420 Hz
Operating Temperature0 to 55
Dimensions (W×H×D)425.5 × 88.9 × 502.9
Weight12

General Performance

ParameterValue
Maximum Resolution8 ½ digits

DC Voltage Specifications

ParameterValue
DC Voltage Ranges100 mV, 1 V, 10 V, 100 V, 1000 V
Input Resistance> 10 GΩ (0.1 V to 10 V ranges), 10 MΩ ±1% (100 V to 1000 V ranges)
Transfer Accuracy0.1 ppm
1-Year Accuracy8 ppm (4 ppm with option 002)

AC Voltage Specifications

ParameterValue
AC Voltage Ranges10 mV, 100 mV, 1 V, 10 V, 100 V, 1000 V
Frequency Bandwidth1 Hz to 10 MHz
Measurement TechniqueAnalog RMS, Synchronous Sub-sampled, Random Sampled

DC Current Specifications

ParameterValue
DC Current Ranges100 nA, 1 µA, 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A

AC Current Specifications

ParameterValue
AC Current Ranges100 µA, 1 mA, 10 mA, 100 mA, 1 A

Resistance Specifications

ParameterValue
Resistance Ranges10 Ω, 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ, 1 GΩ
Measurement Configurations2-Wire, 4-Wire
Offset Compensation CapabilityYes

Frequency and Period Specifications

ParameterValue
Frequency Range1 Hz to 10 MHz
Period Range100 ns to 1 s

Measurement Speeds and Characteristics

ParameterValue
Maximum Readings per Second100,000 readings/sec (at 4 ½ digits)

Triggering and Memory

ParameterValue
Standard Reading Memory20 KB
Optional Reading Memory Expansion148 KB (Option 001)

Math and Advanced Functions

ParameterValue
Supported Math FunctionsNull, Scale, Offset, Filter, Statistics (Min/Max/Avg/StdDev), Limit Testing

Environmental and Calibration

ParameterValue
Auto-Calibration (ACAL) CapabilityYes (Internal comprehensive calibration for DCV, Ohms, and AC)

Connectivity and Interfaces

ParameterValue
Front/Rear Terminal SwitchYes

Physical and Mechanical

ParameterValue
Rack Mount Compatibility2U Full-Rack

Performance

ParameterValue
High Stability ReferenceOption 002 (4 ppm/year)

Environmental

ParameterValue
Storage Temperature-40 °C to +75 °C

Inputs & Protection

ParameterValue
Maximum Input Voltage1000 V (DC or peak AC)
Maximum Input Current1 A (DC or AC rms)

Interference Rejection

ParameterValue
Common Mode Rejection (CMR)140 dB (DC, with 1 kΩ unbalance in LO lead)
Normal Mode Rejection (NMR)> 80 dB (at 50/60 Hz ± 0.1%)

AC Voltage & Current

ParameterValue
Maximum Crest Factor5:1 at full scale

Power

ParameterValue
Power Line Frequency48 - 66 Hz or 360 - 420 Hz

Technology

Digital Multimeter

Applications

Calibration, R&D, Automated Test Equipment (ATE)