The Keysight/Agilent/HP 3478A is a highly reliable 5.5-digit benchtop and system digital multimeter equipped with a standard HP-IB (GPIB) interface. It is designed for precise measurements of DC and AC voltage, DC and AC current, and 2-wire or 4-wire resistance, featuring electronic calibration and switchable front/rear input terminals.

Specifications

General

ParameterValue
InterfacesHP-IB (IEEE-488.1)
Display12-character Liquid Crystal Display (LCD) with integrated annunciators
Ranging ModesAuto and Manual
Power-on Self TestYes

Physical

ParameterValue
Operating Temperature0 to 55 °C
Dimensions (W×H×D)212.3 mm W × 88.9 mm H × 275.6 mm D
Weight3.0 kg

DC Voltage

ParameterValue
DCV Ranges30 mV, 300 mV, 3 V, 30 V, 300 V
DCV Maximum Resolution100 nV (on 30 mV range)
DCV Input Resistance (≤ 3 V Ranges)> 10 GΩ
DCV Input Resistance (30 V and 300 V Ranges)10 MΩ ± 1%
DCV Maximum Input Voltage300 V DC or peak AC (HI to LO)
Input Bias Current< 20 pA at 25 °C

AC Voltage

ParameterValue
ACV Ranges300 mV, 3 V, 30 V, 300 V
ACV Measurement TypeTrue RMS (AC-coupled)
ACV Frequency Range20 Hz to 100 kHz
ACV Input Impedance1 MΩ ± 1% shunted by < 60 pF
ACV Maximum Input Voltage300 V RMS, 450 V peak
ACV Maximum Resolution1 μV (on 300 mV range)
ACV Crest Factor4:1 at full scale

DC Current

ParameterValue
DCI Ranges300 mA, 3 A
DCI Maximum Resolution1 μA (on 300 mA range)
DCI Protection Fuse3 A, 250 V (rear panel)
DCI Maximum Continuous Input3 A
DCI Burden Voltage (300 mA Range)< 0.3 V
DCI Burden Voltage (3 A Range)< 1.0 V

AC Current

ParameterValue
ACI Ranges300 mA, 3 A
ACI Measurement TypeTrue RMS (AC-coupled)
ACI Maximum Resolution1 μA (on 300 mA range)
ACI Frequency Range20 Hz to 20 kHz
ACI Protection Fuse3 A, 250 V
ACI Burden Voltage (300 mA Range)< 0.3 V
ACI Burden Voltage (3 A Range)< 1.0 V

Resistance

ParameterValue
Ohms Measurement Types2-wire and 4-wire ohms
Ohms Ranges30 Ω, 300 Ω, 3 kΩ, 30 kΩ, 300 kΩ, 3 MΩ, 30 MΩ
Ohms Maximum Resolution100 μΩ (on 30 Ω range)
Ohms Test Current (30 Ω, 300 Ω Ranges)1 mA
Ohms Test Current (3 kΩ Range)1 mA
Ohms Test Current (30 kΩ Range)100 μA
Ohms Test Current (300 kΩ Range)10 μA
Ohms Test Current (3 MΩ Range)1 μA
Ohms Test Current (30 MΩ Range)100 nA
Ohms Input ProtectionProtected to 250 V RMS
Max. Lead Resistance (4-wire ohms)100 Ω in any lead (10 Ω on 30 Ω range)
Ohms Open Circuit Voltage< 6.5 V

Measurement Rates and Performance

ParameterValue
Display Resolution Modes3.5, 4.5, or 5.5 digits
Maximum Reading Rate (3.5 digits)71 readings per second (at 60 Hz)
Maximum Reading Rate (4.5 digits)33 readings per second (at 60 Hz)
Maximum Reading Rate (5.5 digits)1.4 readings per second (at 60 Hz)
Integration Time / Power Line Cycles (5.5 digits)10 NPLC
Integration Time / Power Line Cycles (4.5 digits)1 NPLC
Integration Time / Power Line Cycles (3.5 digits)0.1 NPLC
A/D Converter TechnologyMulti-slope II integrating A/D

Noise and Rejection

ParameterValue
Normal Mode Rejection Ratio (NMRR)> 80 dB at 50/60 Hz ±0.1% (in 5.5-digit mode)
Common Mode Rejection Ratio (CMRR) DC> 140 dB at DC (with 1 kΩ imbalance in LO)
Common Mode Rejection Ratio (CMRR) AC> 120 dB at 50/60 Hz (with 1 kΩ imbalance in LO)

System and Interface Configuration

ParameterValue
HP-IB Interface StandardIEEE-488.1 compliant
HP-IB Capability CodesSH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E1
HP-IB Address SelectionRear-panel DIP switch-selectable
Calibration MethodElectronic, from front panel or over HP-IB (no internal hardware adjustments)
Calibration Security LockRear panel hardware slide switch
Maximum Common Mode Voltage500 V peak (LO to chassis ground)

Math and Control Functions

ParameterValue
Front/Rear Input SwitchingNo rear input terminals / Front panel inputs only
Trigger ModesInternal, External, Hold, Fast (bypasses settling delay)

Physical and Environmental

ParameterValue
Storage Temperature Range-40 °C to +75 °C
Warm-up Time1 hour to rated specifications
Relative Humidity LimitsUp to 95% RH at 40 °C
Form FactorHalf-rack width benchtop / system cabinet

Power Requirements

ParameterValue
AC Input Frequency Range48 Hz to 66 Hz
AC Input Voltage Settings100 V / 120 V / 220 V / 240 V AC ±10%
Power Consumption15 VA maximum
Line Frequency48 Hz to 66 Hz
Line Fuse0.15 A @ 115 V / 0.08 A @ 230 V (slow-blow)

DC Voltage Accuracy

ParameterValue
DCV 1-Year Accuracy (3 V Range)±(0.0035% of reading + 2 counts) (at 5.5 digits, 23°C ± 5°C)
DCV 1-Year Accuracy (30 V / 300 V Ranges)±(0.004% of reading + 2 counts) / ±(0.0045% of reading + 2 counts) (at 5.5 digits, 23°C ± 5°C)
DCV 1-Year Accuracy (300 mV Range)±(0.005% of reading + 4 counts) (at 5.5 digits, 23°C ± 5°C)

AC Voltage Accuracy

ParameterValue
ACV 1-Year Accuracy (45 Hz - 20 kHz)±(0.16% of reading + 110 counts) (3 V to 300 V ranges, at 5.5 digits, 23°C ± 5°C)

Resistance Accuracy

ParameterValue
Ohms 1-Year Accuracy (3 kΩ to 300 kΩ Ranges)±(0.011% of reading + 2 counts) (at 5.5 digits, 23°C ± 5°C)
Ohms 1-Year Accuracy (30 Ω Range)±(0.012% of reading + 35 counts) (with Null, at 5.5 digits, 23°C ± 5°C)

DC Current Accuracy

ParameterValue
DCI 1-Year Accuracy (300 mA Range)±(0.15% of reading + 20 counts) (at 5.5 digits, 23°C ± 5°C)
DCI 1-Year Accuracy (3 A Range)±(0.15% of reading + 10 counts) (at 5.5 digits, 23°C ± 5°C)

AC Current Accuracy

ParameterValue
ACI 1-Year Accuracy (45 Hz - 20 kHz)±(0.8% of reading + 120 counts) (300 mA range, at 5.5 digits, 23°C ± 5°C)

Input Protection

ParameterValue
DCV/ACV Input Protection300 V rms or 450 V peak continuous on all ranges

Triggering

ParameterValue
External Trigger Input LevelTTL compatible (negative-edge triggered)
External Trigger Minimum Pulse Width1.5 µs
Voltmeter Complete Pulse Width~1.5 µs
Voltmeter Complete Output LevelTTL compatible (negative-going pulse)

Calibration

ParameterValue
Calibration StorageInternal non-volatile RAM (cal RAM) with lithium battery backup

Data Storage

ParameterValue
Reading Storage MemoryNone (direct streaming via HP-IB)

Display

ParameterValue
Maximum Display Count303,000 counts (except 300 V range)

Compliance

ParameterValue
Safety ComplianceDesigned to comply with IEC 348 and UL 1244

Technology

Multi-slope II integrating A/D conversion

Applications

Benchtop electronic testing and automated system measurement