The HP 4140B is a high-precision picoammeter and dual DC voltage source designed for measuring extremely low currents and performing quasi-static C-V (capacitance-voltage) characterization of semiconductor devices and materials.

Specifications

General

ParameterValue
InterfacesHP-IB (IEEE-488) (SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT1, C0)
Display3.5-digit LED (current/capacitance), 4.5-digit LED (voltage)

Physical

ParameterValue
Power Supply100/120/220/240 V AC (±10%), 48-66 Hz, ~135 VA
Operating Temperature0 °C to 40 °C
Dimensions (W×H×D)426 mm × 177 mm × 498 mm
Weight13 kg

Current Measurement Performance

ParameterValue
Measurement Ranges±1.000 x 10^-12 A to ±19.99 x 10^-3 A
Maximum Resolution10^-15 A (1 fA)
DC Current Accuracy0.5%
Calculated Capacitance Range0.1 pF to 1999 pF
Capacitance Measurement RangesTwo ranges
Zero Offset FunctionAvailable
Ranging ModesAutomatic or manual

Voltage Source Specifications

ParameterValue
Voltage Source Output Range±100 V
Voltage Programming Resolution10 mV (10 V range), 100 mV (100 V range)
Voltage Ramp Rate0.001 V/s to 1.000 V/s (10 V range), 0.01 V/s to 10.00 V/s (100 V range)
Number of DC Voltage SourcesTwo programmable sources
Selectable Source Current LimitAvailable

Measurement Speed and Timing

ParameterValue
Trigger ModesINT, EXT, HOLD/MAN
Integration TimesSelectable (short, medium, or long)
Measurement FunctionsI-V, C-V, I
Quasi-Static C-V MeasurementsSupported
Hold and Step Delay FunctionProgrammable

Inputs and Outputs

ParameterValue
Input Terminal TypeTriaxial
Analog OutputAvailable for plotting I-V, C-V data

Environmental and Safety Compliance

ParameterValue
Self Test FunctionAvailable

Connectivity and Control Interfaces

ParameterValue
Parameter Selection and EntryManual or programmable

Environmental

ParameterValue
Operating Humidity≤ 70% RH (at 40 °C)
Storage Temperature-20 °C to +60 °C

Performance

ParameterValue
Warm-up Time30 minutes

Voltage Sources

ParameterValue
Voltage Source Accuracy±(0.5% of setting + 10 mV) on 10V range, ±(0.5% of setting + 100 mV) on 100V range
Voltage Source Output Impedance< 0.2 Ω
Voltage Source Maximum Current10 mA

Picoammeter (Current Measurement)

ParameterValue
Input Burden Voltage≤ 10 µV (10^-12 A to 10^-5 A ranges), ≤ 100 µV (10^-4 A to 10^-2 A ranges)

Interfaces

ParameterValue
HP-IB Interface CapabilitiesSH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT1, C0

Safety & Protection

ParameterValue
Maximum Allowable Input Voltage100 V

Technology

Semiconductor Characterization / Low-Current Measurement

Applications

Picoampere current measurements and Quasi-Static C-V characterization