The Keysight (Agilent) 41501A is an SMU and Pulse Generator Expander designed to extend the capabilities of the 4155A/B and 4156A/B Semiconductor Parameter Analyzers. It is not a standalone system, but rather a mainframe expansion unit that can house high-current SMUs, medium-power SMUs, and Pulse Generator Units (PGUs) to handle more complex semiconductor device characterization tasks.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | Dedicated expander interface (connects directly to 4155/4156 mainframe) |
| Display | None (status LED indicators only; displays on connected mainframe) |
| Warm-up Time | 40 minutes |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 90 V to 264 V AC, 47 Hz to 63 Hz, 350 VA max |
| Operating Temperature | 5 °C to 40 °C |
| Dimensions (W×H×D) | 338 mm × 235 mm × 450 mm |
| Weight | 17 kg (with 2 PGUs and 1 HCSMU) |
System Compatibility
| Parameter | Value |
|---|---|
| Supported Mainframes | Keysight/Agilent 4155A, 4155B, 4156A, 4156B |
System Configuration
| Parameter | Value |
|---|---|
| Slot Capacity | 4 slots (for HCSMU, MPSMU, and PGU modules) |
Pulse Generator Unit (PGU) Performance
| Parameter | Value |
|---|---|
| Pulse Generator Unit (PGU) Channels | Up to 2 channels (with Option 402 or 420) |
| PGU Amplitude Range (into High Impedance) | -40 V to +40 V |
| PGU Amplitude Range (into 50 Ω) | -20 V to +20 V |
| PGU Output Impedance | 50 Ω or low impedance |
| PGU Pulse Period Range | 1.0 µs to 9.99 s |
| PGU Pulse Period Resolution | 100 ns |
| PGU Pulse Width Range | 1.0 µs to 9.99 s |
| PGU Pulse Delay Range | 0 s to 9.99 s |
| PGU Transition Time Range (Rise/Fall) | 100 ns to 10 ms |
High Current SMU (HCSMU) Performance
| Parameter | Value |
|---|---|
| High Current SMU (HCSMU) Support | 1 channel (with Option 402, 410, 411, or 412) |
| HCSMU Maximum Current (DC) | 100 mA |
| HCSMU Maximum Current (Pulse) | 1 A |
| HCSMU Maximum Voltage (DC) | 200 V |
| HCSMU Maximum Voltage (Pulse) | 200 V |
Medium Power SMU (MPSMU) Performance
| Parameter | Value |
|---|---|
| Medium Power SMU (MPSMU) Support | Up to 2 channels (1 channel with Option 411, 2 channels with Option 412) |
| MPSMU Maximum Current | 100 mA |
| MPSMU Maximum Voltage | 100 V |
Environmental
| Parameter | Value |
|---|---|
| Operating Humidity Range | 8% to 80% RH (non-condensing) |
| Cooling Method | Forced air cooling with rear-panel fan |
| Storage Temperature Range | -20 °C to +60 °C |
| Storage Humidity Range | 5% to 90% RH (non-condensing) |
High Current SMU (HCSMU)
| Parameter | Value |
|---|---|
| HCSMU Current Measurement Resolution | 10 pA |
| HCSMU Voltage Measurement Resolution | 2 μV |
| HCSMU Connector Type | Triaxial (Force/Sense) |
| HCSMU Pulse Width Range | 100 µs to 1 ms |
| HCSMU Maximum Pulse Duty Cycle | 3% |
Medium Power SMU (MPSMU)
| Parameter | Value |
|---|---|
| MPSMU Current Measurement Resolution | 10 fA |
| MPSMU Voltage Measurement Resolution | 2 μV |
| MPSMU Connector Type | Triaxial (Force/Sense) |
Pulse Generator Unit (PGU)
| Parameter | Value |
|---|---|
| PGU Pulse Period Accuracy | ±(1% + 10 ns) |
| PGU Pulse Width Accuracy | ±(3% + 2 ns) |
| PGU Pulse Delay Accuracy | ±(3% + 2 ns) |
| PGU Connector Type | BNC |
Technology
Analog and digital semiconductor measurement expansion technology
Applications