The Keysight (formerly Agilent / Hewlett-Packard) 4284A is an industry-standard precision LCR meter designed for component and material measurement. Featuring a wide 20 Hz to 1 MHz test frequency range and an exceptional basic accuracy of 0.05%, it is widely used in R&D, production, and quality assurance environments. The instrument supports versatile measurement setups with optional high-power test signals and DC bias capabilities.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488) |
| Display | Dot-matrix LCD |
| Warm-up Time | 30 minutes |
| Removable Storage | Memory card slot (JEIDA standard) for instrument states |
| Internal Setup Storage | 10 instrument states |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 / 120 / 220 / 240 V ±10%, 47 - 66 Hz, 200 VA max |
| Operating Temperature | 0 to 55 |
| Dimensions (W×H×D) | 426 × 177 × 498 |
| Weight | 15 |
Measurement Range and Accuracy
| Parameter | Value |
|---|---|
| Basic Accuracy | 0.05% |
| Impedance (|Z|, R, X) Measurement Range | 0.01 mΩ to 99.9999 MΩ |
| Admittance (|Y|, G, B) Measurement Range | 0.01 nS to 99.9999 S |
| Dissipation Factor (D) Measurement Range | 0.000001 to 9.99999 |
| Quality Factor (Q) Measurement Range | 0.01 to 99999.9 |
| Phase Angle (Theta) Measurement Range | -180.000° to 180.000° |
| Error Correction/Calibration | Open, Short, Load |
| Cable Length Compensation | 0 m, 1 m (2 m and 4 m with Option 006) |
Measurement Parameters
| Parameter | Value |
|---|---|
| Primary Measurement Parameters | |Z|, |Y|, L, C, R, G |
| Secondary Measurement Parameters | D, Q, θ, X, B, Rs (ESR), Rp |
| Display Digits/Resolution | 6 digits |
Test Signal Characteristics
| Parameter | Value |
|---|---|
| Test Frequency Range | 20 Hz to 1 MHz |
| Test Frequency Resolution | 8610 selectable frequencies |
| Test Frequency Accuracy | ±0.01% |
| Test Signal Voltage Range | 5 mVrms to 2 Vrms (up to 20 Vrms with Option 001) |
| Test Signal Current Range | 50 µArms to 20 mArms (up to 200 mArms with Option 001) |
| Output Impedance | 100 Ω (nominal) |
| DC Bias Voltage Range (Standard) | 0 V, 1.5 V, 2 V |
Options and Accessories
| Parameter | Value |
|---|---|
| High Power DC Bias Option | ±40 V DC (Option 001) |
| Power Amplifier Option | Option 001 (increases test signals to 20 Vrms / 200 mArms and bias to ±40 V) |
| Handler Interface Option | Option 201 or 202 |
| Scanner Interface Option | Option 301 |
Measurement Speed and Triggering
| Parameter | Value |
|---|---|
| Measurement Time Modes | Short, Medium, Long |
| Trigger Modes | Internal, Manual, External, Bus |
| Trigger Delay Range | 0 to 60.000 s |
| Averaging (Number of Measurements) | 1 to 256 |
Advanced Features and Functions
| Parameter | Value |
|---|---|
| List Sweep Points | 10 points |
| List Sweep Parameters | Test Frequency, Test Signal Voltage/Current, DC Bias Voltage/Current |
| Comparator / Bin Sorting | Supported |
| Number of Bins | 10 bins (9 pass, 1 fail) |
| Continuous Memory / State Storage | 10 instrument states |
Interfaces and Connectivity
| Parameter | Value |
|---|---|
| Remote Programming Interface | GPIB |
Physical and Environmental
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 to 70 °C |
Measurement Ranges
| Parameter | Value |
|---|---|
| Capacitance (C) Measurement Range | 0.01 fF to 9.99999 F |
| Inductance (L) Measurement Range | 0.01 nH to 99.9999 kH |
Measurement Features
| Parameter | Value |
|---|---|
| Equivalent Circuit Modes | Series, Parallel |
| Deviation Measurement | ∆ (absolute deviation) and ∆% (percent deviation) |
| Automatic Level Control (ALC) | Voltage and Current (Maintains constant signal level across DUT) |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Measurement Terminals | 4-terminal pair (BNC) |
Options & Accessories
| Parameter | Value |
|---|---|
| Bias Current Interface (Option 002) | Enables connection to 42841A Bias Current Source (up to 40 A DC) |
| External DC Bias Current Capability | Up to 40 A (requires Option 002, external 42841A Bias Current Source, and appropriate fixture) |
| Extended Cable Length Compensation | 2 m and 4 m lengths (requires Option 006) |
Power
| Parameter | Value |
|---|---|
| Power Consumption | 200 VA max |
Features
| Parameter | Value |
|---|---|
| Test Signal Level Monitor | Voltage (V) and Current (I) monitor across/through DUT |
Applications
Precision LCR measurement for components and materials in R&D, incoming inspection, and production.