The Keysight (formerly Agilent / HP) 435B is a classic, highly reliable analog RF and microwave power meter. It is designed to work seamlessly with the 8480 series of power sensors, offering precise measurements over a wide frequency and power range. Renowned for its low noise, automatic zeroing capability, and optional portable battery operation, the 435B remains an industry icon for analog RF power measurement.

Specifications

General

ParameterValue
InterfacesAnalog Recorder Output (rear panel BNC); no digital programming interfaces
DisplayAnalog d'Arsonval meter with dual-scale (0 to 10 & 0 to 3) and dBm scale (-10 to +3)

Physical

ParameterValue
Power Supply100/120/220/240 V AC (+5% to -10%), 48 to 66 Hz and 360 to 440 Hz, < 10 W; rechargeable battery optional (Option 001)
Operating Temperature0 to 55 °C
Dimensions (W×H×D)134 mm × 155 mm × 279 mm
Weight2.6 kg (standard, without battery)
Cabinet StyleHP System II 1/3-rack

Measurement Capabilities

ParameterValue
Number of Channels1
Frequency Range100 kHz to 110 GHz (sensor dependent)
Power Measurement Range-70 dBm to +44 dBm (sensor dependent)

Measurement Accuracy and Calibration

ParameterValue
Instrumentation Accuracy±1% of full scale on all ranges
Internal Calibration Source Frequency50 MHz
Internal Calibration Source Power Output1.00 mW (0 dBm)
Internal Calibration Source Uncertainty±0.7% (traceable to National Standards)
Internal Calibration Source ConnectorType N (female) on front panel
ZeroingManual zeroing via front-panel knob

Sensor Compatibility and Connection

ParameterValue
Compatible Sensor Families8480 Series Power Sensors
Supported Cable LengthsUp to 61 m (200 ft)
Calibration Factor Control85% to 100% in 1% steps via front-panel dial

User Interface and Display

ParameterValue
Meter Response TimeFast (approx. 100 ms) or Slow (approx. 2.5 s) switch-selectable

System Connectivity and Interfaces

ParameterValue
Recorder Output Voltage0 to 1.0 VDC into open circuit (proportional to meter deflection)
Recorder Output Impedance1 kΩ nominal

Physical and Power Specifications

ParameterValue
Power Consumption< 10 W (AC operation)
Battery Option AvailabilityAvailable as Option 001 (internal rechargeable pack)

Environmental and Compliance

ParameterValue
Storage Temperature Range-40 °C to +75 °C (excluding battery)
Relative Humidity Limit95% R.H. at 40 °C

Display

ParameterValue
Meter ScalesdBm (-10 to +3) and linear (0 to 1.0 and 0 to 3.0)
Meter Movement TypeTaut-band, mirror-backed

Power

ParameterValue
Option 001 Battery Operating TimeUp to 16 hours
Option 001 Battery Recharge Time16 hours (instrument off), 24 hours (instrument on)
AC Line Frequency Range48 to 66 Hz and 360 to 440 Hz

Inputs & Outputs

ParameterValue
Input Connector Location OptionsFront panel (standard), Rear panel (Option 002)

Reference Oscillator

ParameterValue
Reference Oscillator Impedance50 Ω (75 Ω with Option 003)
Reference Oscillator SWR< 1.05

Performance

ParameterValue
Zero Carryover< ±0.5% of full scale when zeroed on the most sensitive range

Technology

RF & Microwave Power Measurement (Analog)

Applications

Analog RF Power Measurement, Sensor Calibration, Field Testing