The Keysight (Agilent/HP) 53132A is a high-performance universal frequency counter offering exceptional resolution and speed for demanding benchtop and system applications. It features 12 digits per second frequency resolution, 150 ps time interval resolution, and built-in statistics and limit testing for automated analysis.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488), RS-232
DisplayVacuum Fluorescent Display (VFD)

Physical

ParameterValue
Power Supply100-120 V (50/60/400 Hz), 220-240 V (50/60 Hz), 170 VA max
Operating Temperature0 to 55
Dimensions (W×H×D)212.6 × 88.5 × 348.3
Weight3.1

Input Characteristics (Channels 1 & 2)

ParameterValue
Number of Channels2 standard (plus optional Channel 3)
Frequency RangeDC to 225 MHz
Input Impedance1 MΩ or 50 Ω (selectable)
Input CouplingAC or DC
Low Pass Filter100 kHz (switchable)

Frequency & Period Specifications

ParameterValue
Frequency Resolution12 digits per second

Time Interval Specifications

ParameterValue
Single-Shot Resolution150 ps

Timebase & Reference Oscillator

ParameterValue
Standard Timebase Frequency10 MHz

Math & Statistics

ParameterValue
Statistics OperationsAverage, Min/Max, Standard Deviation
Limit TestingAutomated limit tests

Connectivity & Interfaces

ParameterValue
Command Set CompatibilitySCPI

Physical & Power Specifications

ParameterValue
Form Factor2U half-rack

Inputs & Triggering

ParameterValue
Input Attenuation (Ch 1 & 2)x1 or x10
Maximum Input Voltage (50 Ω)5 Vrms
Maximum Input Voltage (1 MΩ)350 Vdc + AC peak (DC to 2 kHz)
Input Sensitivity (DC to 100 MHz)20 mVrms
Input Sensitivity (100 to 200 MHz)30 mVrms
Input Sensitivity (200 to 225 MHz)40 mVrms
Trigger Level Range (x1)±5.125 V
Trigger Level Resolution5 mV

Optional Channel 3

ParameterValue
Channel 3 Frequency Range (Option 030)100 MHz to 3.0 GHz
Channel 3 Frequency Range (Option 050)200 MHz to 5.0 GHz
Channel 3 Frequency Range (Option 124)200 MHz to 12.4 GHz

Performance

ParameterValue
Gate Time Range1 ms to 1000 s
Measurement Rate (via GPIB)Up to 200 measurements/s

Timebase & Reference

ParameterValue
External Reference Input Frequencies1 MHz, 5 MHz, or 10 MHz
Timebase Output10 MHz (> 1 Vpp into 50 Ω)
Timebase OptionsOption 001 (Medium Stability), Option 010 (High Stability), Option 012 (Ultra-High Stability)
External Reference Input Amplitude200 mV rms to 10 V rms
Timebase Output Amplitude> 1 V p-p into 50 Ω

Compliance

ParameterValue
Safety ComplianceIEC 1010-1, CSA C22.2 No 1010.1, UL 3111-1
EMI/EMC ComplianceCISPR-11, EN55011 Group 1 Class A

Measurement Ranges

ParameterValue
Time Interval Range1 ns to 10^5 s
Period Range4.44 ns to 10 s
Pulse Width Range5 ns to 10^5 s
Rise / Fall Time Range5 ns to 10^5 s
Phase Range-180° to 360°
Totalize Capacity0 to 10^15

Input Characteristics

ParameterValue
Trigger Level Accuracy±(15 mV + 1% of trigger level)

Option 030 / 050 / 124 (Channel 3)

ParameterValue
Channel 3 Input Impedance50 Ω (AC coupled)

Features

ParameterValue
Measurement MathScale and Offset

Power & Environmental

ParameterValue
Power Consumption170 VA max
Storage Temperature-40 °C to +71 °C
Operating HumidityUp to 95% RH at 40 °C

Technology

Frequency Counter

Applications

Test and Measurement