The Keysight (HP / Agilent) 5343A is an automatic continuous-wave (CW) microwave frequency counter designed for highly accurate frequency measurements up to 26.5 GHz. Known for its excellent sensitivity, high resolution 11-digit LED display, and robust design, the 5343A is a standard in RF and microwave labs. With options for HP-IB (GPIB) interfacing, digital-to-analog conversion, and high-stability oven timebases, it offers excellent performance and flexibility for legacy and modern test environments.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | HP-IB (Option 011) |
| Display | Eleven-digit LED display |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 / 120 / 220 / 240 V AC (+5%, -10%), 48 to 66 Hz |
| Operating Temperature | 0 to 50 °C |
| Dimensions (W×H×D) | 425.5 mm × 132.6 mm × 498 mm |
| Weight | 8.2 kg |
Frequency Measurement Specifications
| Parameter | Value |
|---|---|
| Frequency Range Channel 1 | 500 MHz to 26.5 GHz |
| Frequency Range Channel 2 | 10 Hz to 520 MHz |
| Sensitivity Channel 1 | -33 dBm (500 MHz to 12.4 GHz), -28 dBm (12.4 GHz to 18 GHz), -23 dBm (18 GHz to 26.5 GHz) |
| Sensitivity Channel 2 | 10 mV RMS (10 Hz to 520 MHz) |
| Maximum Input Level Channel 1 | +7 dBm |
| Maximum Input Level Channel 2 | 1 V RMS (50 Ω), 125 V RMS (1 MΩ, <= 1 kHz) |
| Damage Level Channel 1 | +25 dBm |
| Input Impedance Channel 1 | 50 Ω nominal |
| Input Impedance Channel 2 | 50 Ω or 1 MΩ nominal (selectable) |
| Input Coupling Channel 1 | AC coupled |
| Input Coupling Channel 2 | AC or DC (switch selectable) |
| Acquisition Time | < 200 ms |
| Input Connector Channel 1 | APC-3.5 (male) |
| Input Connector Channel 2 | BNC (female) |
Timebase & Reference Oscillator
| Parameter | Value |
|---|---|
| Standard Reference Oscillator Frequency | 10 MHz |
| Aging Rate Monthly | < 1 x 10^-7 / month |
| Temperature Stability | < 1 x 10^-6 over 0 °C to 50 °C |
| External Reference Input Frequency | 1, 2, 5, or 10 MHz |
| External Reference Input Coupling & Termination | AC coupled, 1 kΩ nominal input impedance |
| Internal Reference Output Frequency | 10 MHz |
System Interfaces & Connectivity
| Parameter | Value |
|---|---|
| GPIB Interface Support | Optional (Option 011) |
Options & Accessories
| Parameter | Value |
|---|---|
| Digital-to-Analog Converter Option | Option 004 |
| High Stability Oven Timebase Option | Option 001 |
| Rear Panel Connector Conversion Option | Option 002 |
Power Requirements
| Parameter | Value |
|---|---|
| Power Consumption | 100 VA max |
Environmental & Compliance
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to 75 °C |
Performance
| Parameter | Value |
|---|---|
| Frequency Resolution | 1 Hz to 1 MHz (selectable in decade steps) |
| Gate Time | 1 ms to 1 s (resolution dependent) |
| Tracking Speed | 1 MHz/s (normal mode), 800 MHz/s (fast tracking mode) |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Internal Reference Output Level | > 2.4 V p-p into 50 Ω |
Standards & Interface
| Parameter | Value |
|---|---|
| GPIB Interface Capabilities | SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT1, C0 |
Input Characteristics
| Parameter | Value |
|---|---|
| Maximum Damage Level Channel 2 | +30 dBm (1 W) continuous |
| Channel 1 Low Pass Filter | 10 kHz nominal (switch-selectable) |
Reference Oscillator Options
| Parameter | Value |
|---|---|
| Oven Timebase Aging Rate (Option 001) | < 5 × 10^-10 per day (after 24-hour warm-up) |
| Oven Timebase Temperature Stability (Option 001) | < 7 × 10^-9 over 0 °C to 50 °C |
| Oven Timebase Line Voltage Stability (Option 001) | < 1 × 10^-10 for ±10% line voltage change |
Analog Output Options
| Parameter | Value |
|---|---|
| DAC Output Voltage Range (Option 004) | 0 to 0.999 V in 1 mV steps (corresponding to selected 3 digits of the display) |
Options
| Parameter | Value |
|---|---|
| Precision Input Connector Option (Option 006) | Replaces standard APC-3.5 male connector with Type N connector (decreases upper frequency limit to 18 GHz) |
| 30 GHz Frequency Range Extension Option (Option 005) | Extends Channel 2 upper limit to 30 GHz |
Electrical & Power
| Parameter | Value |
|---|---|
| Line Voltage Requirements | 100 V / 120 V / 220 V / 240 V AC (+5%, -10%; switch-selectable) |
| Line Frequency Range | 48 Hz to 66 Hz |
Technology
Heterodyne Converter
Applications
RF & Microwave Frequency Measurement