The Keysight (Agilent/HP) 5347A is a high-performance 20 GHz Microwave Counter and Power Meter integrated into a single rugged, portable instrument. It provides precise frequency measurements from 10 Hz to 20 GHz alongside accurate power measurements using standard HP 8480 series power sensors.

Specifications

General

ParameterValue
InterfacesHP-IB (IEEE-488.2)
DisplayBacklit custom Liquid Crystal Display (LCD) with 12-digit frequency and 4-digit power readout

Physical

ParameterValue
Power Supply90 to 132 V ac or 198 to 264 V ac (auto-selecting), 47 to 66 Hz
Operating Temperature0 to 50 °C
Dimensions (W×H×D)325 mm × 133 mm × 432 mm
Weight9.1 kg (without options); 11.4 kg (with Option 002 battery)

Frequency Counter - Channel 1

ParameterValue
Channel 1 Frequency Range10 Hz to 520 MHz
Channel 1 Sensitivity25 mV rms
Channel 1 Input Impedance1 MΩ (shunted by < 70 pF) or 50 Ω nominal (selectable)
Channel 1 CouplingAC
Channel 1 Maximum Input Level5 V rms (50 Ω), 250 V peak (1 MΩ)

Frequency Counter - Channel 2 (Microwave)

ParameterValue
Channel 2 Frequency Range500 MHz to 20 GHz
Channel 2 Sensitivity (500 MHz to 12.4 GHz)-32 dBm
Channel 2 Sensitivity (12.4 GHz to 18 GHz)-27 dBm
Channel 2 Sensitivity (18 GHz to 20 GHz)-22 dBm
Channel 2 Input Impedance50 Ω nominal
Channel 2 Maximum Input Level+7 dBm
Channel 2 Damage Level+25 dBm (peak)
Channel 2 Connector TypePrecision Type N (female)
Acquisition Time< 125 ms (typical)

Power Meter

ParameterValue
Power Meter Frequency Range10 MHz to 20 GHz (sensor dependent)
Power Meter Power Range-70 dBm to +44 dBm (sensor dependent)
Compatible Power SensorsHP 8481A, HP 8481D, HP 8485A
Power Instrumentation Accuracy±0.02 dB or ±0.5%
Power Display Resolution0.01 dB (logarithmic mode) or 0.1% of full scale (linear mode)
Calibration Factor Range70% to 120% in 0.1% steps
Power Sensor Calibration Output1.00 mW (0.00 dBm) at 50 MHz
Calibration Output ConnectorType N (female), front panel

Internal Timebase

ParameterValue
Standard Timebase TypeTCXO
Standard Timebase Nominal Frequency10 MHz
Standard Timebase Temperature Stability< 1 × 10^-6 (0 to 50 °C)
Standard Timebase Aging Rate< 1 × 10^-7 per month
Option 001 High Stability Timebase Aging Rate< 5 × 10^-10 per day (after 24-hour warm-up)
Option 001 High Stability Timebase Temperature Stability< 7 × 10^-9 (0 to 50 °C)
External Reference Input Frequency1, 2, 5, or 10 MHz
External Reference Input Level0.2 V to 2.0 V rms
External Reference Input Impedance1 kΩ nominal
Internal Reference Output Frequency10 MHz
Internal Reference Output Level> 1.5 V p-p into 50 Ω

Power Requirements

ParameterValue
Option 002 Internal Battery TypeSealed Lead-Acid
Option 002 Battery Operating Time> 2 hours typical
Option 002 DC Input Range11 to 24 V dc
Power Consumption< 40 VA typical (excluding battery charging)

Environmental

ParameterValue
Storage Temperature Range-40 to 75 °C
Operating Humidity RangeUp to 95% RH at 40 °C
Safety Standards ComplianceIEC 348, EN 61010-1
Maximum Operating Altitude4,600 m (15,000 ft)
Maximum Storage Altitude15,000 m (50,000 ft)

Counter Channel 2 Specifications

ParameterValue
Channel 2 FM Tolerance20 MHz p-p (modulation frequency 100 Hz to 10 MHz)
Channel 2 AM ToleranceUp to 90% depth for modulation frequencies up to 100 kHz
Channel 2 Amplitude Discrimination10 dB (typical) within 500 MHz; 20 dB (typical) up to 20 GHz

Counter Channel 1 Specifications

ParameterValue
Channel 1 Maximum Resolution1 Hz

Power Meter Specifications

ParameterValue
Power Meter Calibrator Output SWR< 1.08
Power Meter Calibrator Output Frequency50 MHz
Power Meter Offset Range-99.99 dB to +99.99 dB

Timebase Specifications

ParameterValue
Standard Timebase Short Term Stability< 1 × 10^-9 (1 s average)
Option 001 High Stability Timebase Short Term Stability< 1 × 10^-10 (1 s average)
Option 001 High Stability Timebase Warm-up Time< 15 minutes (to within 1 × 10^-7 of final value)

Option 002 Battery Specifications

ParameterValue
Option 002 Battery Charge Time8 hours (with instrument turned off)

GPIB / Interface Specifications

ParameterValue
GPIB Interface FunctionsSH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E1

Channel 1 Input

ParameterValue
Channel 1 Connector TypeBNC female

Channel 2 Performance

ParameterValue
Channel 2 Frequency Resolution1 Hz to 1 MHz (user selectable)
Channel 2 Acquisition ModesAutomatic, Manual

Channel 2 Input

ParameterValue
Channel 2 VSWR1.5:1 (500 MHz to 12.4 GHz), 2.0:1 (12.4 GHz to 20 GHz)

Power Meter Calibrator

ParameterValue
Power Meter Calibrator Output Accuracy±1.2%
Power Meter Calibrator Output Impedance50 Ω nominal

Power Meter Performance

ParameterValue
Power Meter Calibration Factor Resolution0.1%

Interfaces

ParameterValue
Programming LanguageHP-IB

Technology

RF & Microwave Frequency Counter & Power Meter

Applications

Microwave and RF telecommunications testing, power sensor calibration, and portable frequency and power measurement

Looking for the Keysight Technologies (Agilent HP) 5347A?

Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.