The Keysight (Agilent/HP) 5348A is a high-performance 26.5 GHz microwave frequency counter and true power meter housed in a single, rugged half-rack instrument. Capable of making highly accurate and high-resolution measurements from 10 Hz to 26.5 GHz, the 5348A is engineered with exceptional dynamic range and linearity. This instrument is ideal for portable, field-service, and benchtop microwave applications, particularly when configured with the optional internal rechargeable battery for up to two hours of cordless operation.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488)
DisplayCustom backlit liquid crystal display (LCD)
Power Sensor Maximum Cable Length61 m (200 ft)
Save/Recall Registers9 complete instrument setups

Physical

ParameterValue
Power Supply90 - 132 V AC / 198 - 264 V AC (47 - 66 Hz / 47 - 440 Hz); 11.5 - 18 V DC; optional internal battery
Operating Temperature0 to 50 °C
Dimensions (W×H×D)213 mm × 133 mm × 402 mm
Weight9.1 kg (20 lb)
Rack Height3U (half-rack width)

Frequency Measurement (Channel 1)

ParameterValue
Frequency Range (Ch 1)10 Hz - 520 MHz
Sensitivity (Ch 1)25 mV rms
Input Impedance (Ch 1)Selectable 1 MΩ (shunted by < 70 pF) or 50 Ω nominal
Coupling (Ch 1)AC
Maximum Input Level (Ch 1)1 MΩ: 1 V rms; 50 Ω: 1.5 V rms
Damage Level (Ch 1)1 MΩ: 120 V (DC + peak AC) ≤ 1 kHz; 50 Ω: 5 V rms
Connector Type (Ch 1)BNC female

Frequency Measurement (Channel 2)

ParameterValue
Frequency Range (Ch 2)500 MHz - 26.5 GHz
Sensitivity (Ch 2)-32 dBm (500 MHz - 12.4 GHz), -27 dBm (12.4 GHz - 20 GHz), -20 dBm (20 GHz - 26.5 GHz)
Maximum Input Level (Ch 2)+7 dBm
Damage Level (Ch 2)+25 dBm (peak)
Input Impedance (Ch 2)50 Ω nominal
VSWR (Ch 2)< 2:1 typical to 18 GHz, < 3:1 typical to 26.5 GHz
Acquisition Time (Ch 2)< 125 ms
FM Tolerance (Ch 2)20 MHz p-p
AM Tolerance (Ch 2)Any modulation depth, as long as the minimum signal level is above the sensitivity limit
Connector Type (Ch 2)Precision APC-3.5 male
Tracking Speed (Ch 2)1 MHz/sec

Power Measurement

ParameterValue
Frequency Range (Power)100 kHz - 50 GHz (sensor dependent)
Power Range (Power)-70 to +44 dBm (sensor dependent)
Power Resolution0.01 dB or 0.1% (selectable)
Instrumentation Accuracy±0.02 dB (±0.5%)
Sensor CompatibilityHP 8480 series power sensors
Built-In Calibration TablesROM-stored calibration data for HP 8481A, HP 8481D, HP 8482A, HP 8485A, and HP 8487A
Power Meter LinearityWithin ±0.5% (included in instrument accuracy)

Internal Power Calibrator

ParameterValue
Calibrator Frequency50 MHz
Calibrator Power Level1.00 mW (0 dBm)
Calibrator Accuracy±1.2% (at 25 °C)

Reference Oscillator (Standard)

ParameterValue
Reference Frequency10 MHz
Aging Rate (Per Month, Standard)< 1 × 10^-7
Temperature Stability (Standard)< 1 × 10^-6 (0 °C to 50 °C)
Line Voltage Stability (Standard)< 1 × 10^-7 for a 10% change

Reference Oscillator (Option 001 Oven)

ParameterValue
Aging Rate (Per Day, Option 001)< 5 × 10^-10 (after 24-hour warm-up)
Temperature Stability (Option 001)< 7 × 10^-9 (0 °C to 50 °C)
Warm-up Time (Option 001)< 20 minutes to within 5 × 10^-9 of final value

Reference Input & Output

ParameterValue
External Reference Input Frequency1, 2, 5, and 10 MHz
External Reference Input Level0.2 V to 2.0 V rms
External Reference Impedance1 kΩ nominal
Internal Reference Output Frequency10 MHz

Measurement Functions

ParameterValue
Frequency Resolution SettingsSelectable: 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz
Self-Test DiagnosticsAutomatic internal diagnostics upon power-on

Power Requirements

ParameterValue
Line Voltage Ranges90 - 132 V AC / 198 - 264 V AC (slide switch selection)
Line Frequency Range47 - 66 Hz (all ranges); 47 - 440 Hz (90 - 132 V AC only)

Physical Characteristics

ParameterValue
Form FactorHalf-rack width, 3U height

Environmental

ParameterValue
Storage Temperature Range-40 °C to +75 °C
Operating Humidity Limit95% RH at 40 °C (non-condensing)
Maximum Operating Altitude4,600 m (15,000 ft)

Options

ParameterValue
Option 001High-stability oven timebase
Option 002Rear panel inputs
Option 002 DescriptionRear panel inputs (places Ch 1, Ch 2, Sensor, and Calibrator connectors on rear)

Inputs & Outputs

ParameterValue
Channel 1 Low-Pass Filter Cutoff100 kHz (nominal)
GPIB Interface CapabilitiesSH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT1, C0, E2

Performance

ParameterValue
Channel 2 FM Modulation RateUp to 10 MHz

Compliance

ParameterValue
EMC ComplianceCISPR 11 / EN 55011 Group 1, Class A
Safety ComplianceEN 61010-1 / IEC 1010-1

Frequency Counter

ParameterValue
Frequency Measurement Speed (HP-IB)Up to 30 readings per second

Power Meter

ParameterValue
Power Sensor Calibration Factor Range50.0% to 120.0% (in 0.1% steps)
Power Measurement Speed (HP-IB)Up to 20 readings per second

Power & Battery

ParameterValue
Battery Operating Time (Option 002)Typically 2 hours

Technology

Heterodyne / Frequency Counter & Power Meter

Applications

Microwave Frequency and Power Measurement