The Keysight (Agilent) 54600B is a 100 MHz, 2-channel digital storage oscilloscope renowned for its reliable performance, analog-like feel, and high screen update rate of up to 1.5 million points per second. It is widely used in production testing, educational laboratories, and general-purpose troubleshooting.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | None standard; optional HP-IB (GPIB), RS-232, or Parallel interfaces available via plug-in expansion modules |
| Display | 7-inch monochrome CRT raster display with 500 x 256 pixel resolution |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100-240 VAC (±10%), 45-440 Hz, automatic selection, 75 W maximum power consumption |
| Operating Temperature | 0 to +55 °C |
| Dimensions (W×H×D) | 322 mm x 172 mm x 417 mm |
| Weight | 6.2 kg |
Vertical System
| Parameter | Value |
|---|---|
| Analog Bandwidth (-3 dB) | 100 MHz |
| Number of Input Channels | 2 |
| Calculated Rise Time (10% to 90%) | < 3.5 ns |
| Vertical Resolution (ADC) | 8 bits |
| Input Sensitivity Range | 2 mV/div to 5 V/div |
| Input Impedance | 1 MΩ || ~20 pF |
| Input Coupling | AC, DC, GND |
| Maximum Input Voltage | 400 V (DC + peak AC) |
| DC Gain Accuracy | ±1.5% |
| Offset Range | ±2 V (for 2 mV/div to 200 mV/div); ±40 V (for 500 mV/div to 5 V/div) |
| Bandwidth Limit | 20 MHz (selectable) |
| Channel Math | Channel 1 + Channel 2, Channel 1 - Channel 2, Invert Channel 1 and Channel 2 |
| Vertical Dynamic Range | ±8 divisions from screen center |
| Probe Attenuation Factors | 1x, 10x, 100x |
| Input Capacitance | 20 pF |
| Channel-to-Channel Isolation | 40 dB at 100 MHz |
Horizontal System
| Parameter | Value |
|---|---|
| Maximum Real-Time Sampling Rate | 20 MSa/s |
| Maximum Record Length | 4,000 points (single-channel), 2,000 points (dual-channel) |
| Time Base Range | 2 ns/div to 5 s/div |
| Time Base Accuracy | ±0.01% |
| X-Y Mode Phase Shift | < 3° at 100 kHz |
Trigger System
| Parameter | Value |
|---|---|
| Trigger Sources | Channel 1, Channel 2, Line, External |
| Trigger Modes | Auto, Auto Level, Normal, Single |
| Trigger Coupling | DC, AC, LF Reject, HF Reject, Noise Reject |
| External Trigger Input Impedance | 1 MΩ || ~20 pF |
| External Trigger Maximum Input | 400 V (DC + peak AC) |
| External Trigger Input Range | ±18 V |
| Trigger Sensitivity | DC to 10 MHz: 0.3 div or 5 mV; 10 MHz to 100 MHz: 1 div or 15 mV |
| TV Trigger Standards | NTSC, PAL, SECAM |
Acquisition System
| Parameter | Value |
|---|---|
| Maximum Waveform Capture Rate | 1,500,000 points/s |
Waveform Measurements and Analysis
| Parameter | Value |
|---|---|
| Automatic Measurements | Vpp, Vmax, Vmin, Vavg, Vrms, Vtop, Vbase, Frequency, Period, Duty Cycle, Positive Width, Negative Width, Rise Time, Fall Time |
| Cursors | V1, V2, t1, t2, ΔV, Δt, 1/Δt |
Display and User Interface
| Parameter | Value |
|---|---|
| Display Type | Monochrome CRT raster display |
| Display Size | 7-inch |
| Display Resolution | 500 x 256 pixels |
Power Specifications
| Parameter | Value |
|---|---|
| Power Supply Voltage Range | 100 to 240 VAC (±10%) |
| Power Supply Frequency Range | 45 to 440 Hz |
Environmental Specifications
| Parameter | Value |
|---|---|
| Operating Temperature Range | 0 °C to +55 °C |
| Storage Temperature Range | -40 °C to +70 °C |
| Operating Humidity Range | Up to 95% RH at 40 °C |
| Non-Operating Altitude | Up to 15,240 m (50,000 ft) |
Compliance and Standards
| Parameter | Value |
|---|---|
| Calibration Interval | 1 year |
Storage & Memory
| Parameter | Value |
|---|---|
| Front-Panel Setup Storage | 16 nonvolatile setups |
| Waveform Storage | 2 nonvolatile waveforms |
Connectivity
| Parameter | Value |
|---|---|
| Expansion Slot Compatibility | Supports HP 54650A-series interface and measurement modules |
Display
| Parameter | Value |
|---|---|
| CRT Phosphor Type | P31 |
| Display Graticule | 8 × 10 divisions (internal) |
Triggering
| Parameter | Value |
|---|---|
| Trigger Holdoff Range | 200 ns to 13 s |
Compliance
| Parameter | Value |
|---|---|
| Safety Certifications | IEC 1010-1, UL 1244, CSA C22.2 No. 231 |
| EMC Standards | CISPR 11 / EN 55011 Group 1 Class A |
Technology
Digital Storage Oscilloscope
Applications
Production testing, educational laboratories, and general-purpose electronic troubleshooting