The Keysight Agilent HP 54602B is a 150 MHz, 4-channel (2+2) digital storage oscilloscope from the classic 54600 series. Designed to combine the immediate response of an analog scope with the precision and storage capabilities of a digital instrument, it is a highly reliable tool for bench troubleshooting and educational laboratories.

Specifications

General

ParameterValue
InterfacesNone standard; GPIB (HP-IB), RS-232, or parallel available via optional HP 54650-series plug-in modules
Display7-inch monochrome raster CRT

Physical

ParameterValue
Power Supply100 to 240 VAC (±10%), 45 to 440 Hz, 120 VA maximum
Operating Temperature0 to +55 °C
Dimensions (W×H×D)322 mm × 172 mm × 317 mm
Weight6.2

Vertical System - Analog Channels

ParameterValue
Analog Bandwidth (Channels 1 & 2)150 MHz
Analog Bandwidth (Channels 3 & 4)250 MHz
Calculated Rise Time (Channels 1 & 2)< 2.33 ns
Calculated Rise Time (Channels 3 & 4)< 1.4 ns
Number of Input Channels4 (2 + 2 configuration)
Input Coupling (Channels 1 & 2)AC, DC, GND
Input Coupling (Channels 3 & 4)DC
Input Impedance (Channels 1 & 2)1 MΩ || ~13 pF
Input Impedance (Channels 3 & 4)1 MΩ || ~13 pF
Input Sensitivity Range (Channels 1 & 2)2 mV/div to 5 V/div (expandable to 1 mV/div with vernier)
Input Sensitivity (Channels 3 & 4)0.1 V/div and 1 V/div
Vertical Resolution8 bits
Maximum Input Voltage400 V (DC + peak AC)
DC Gain Accuracy±1.5%

Horizontal System

ParameterValue
Timebase Range2 ns/div to 5 s/div
Timebase Accuracy±0.015%
Horizontal Resolution100 ps

Acquisition System

ParameterValue
Maximum Real-Time Sample Rate20 MSa/s
Maximum Repetitive Sample Rate1.6 GSa/s
Maximum Record Length4,000 points per channel
Acquisition ModesNormal, Peak Detect, Averaging
Peak Detect Glitch CaptureCaptures glitches as narrow as 50 ns

Trigger System

ParameterValue
Trigger SourcesChannels 1, 2, 3, 4, Line
Trigger ModesAuto, Auto Level, Normal, Single
Trigger CouplingDC, AC, LF Reject, HF Reject, Noise Reject
Trigger Sensitivity (DC to 25 MHz)0.35 div or 3.5 mV
Trigger Sensitivity (25 MHz to 150 MHz)1.0 div or 10 mV
Standard Trigger TypesEdge, TV (NTSC, PAL, SECAM)

Waveform Measurements and Analysis

ParameterValue
Automatic Voltage MeasurementsVp-p, Vmax, Vmin, Vtop, Vbase, Vavg, Vrms
Automatic Time MeasurementsPeriod, Frequency, Rise time, Fall time, +Width, -Width, Duty cycle
Waveform Math OperatorsChannel 1 + Channel 2, Channel 1 - Channel 2

Display and User Interface

ParameterValue
Display Size7-inch diagonal
Graticule Format8 × 10 divisions

Connectivity and Interfaces

ParameterValue
Rear Expansion SlotAccepts HP 54650-series interface and measurement enhancement modules

Power Specifications

ParameterValue
Power Voltage Range100 to 240 VAC
Power Frequency Range45 to 440 Hz
Power Consumption220 VA maximum

Environmental and Compliance

ParameterValue
Operating Temperature Range0 to +55 °C
Non-Operating Temperature Range-40 to +70 °C
Operating Humidity95% RH at 40 °C
Operating AltitudeUp to 4,500 meters

Physical Characteristics

ParameterValue
Form FactorBenchtop

Storage

ParameterValue
Non-Volatile Setup Storage16 setups
Non-Volatile Waveform Storage4 waveforms

Triggering

ParameterValue
Trigger Holdoff Range200 ns to 13 s
Trigger HF Reject Filter50 kHz low-pass
Trigger LF Reject Filter50 kHz high-pass

Outputs

ParameterValue
Probe Adjust Output Frequency~1.2 kHz
Probe Adjust Output Voltage~5 V

Measurements

ParameterValue
Cursor MeasurementsV1, V2, ΔV, T1, T2, ΔT, 1/ΔT

Options & Accessories

ParameterValue
Compatible Expansion ModulesHP 54650A (HP-IB), HP 54652A (RS-232), HP 54657A (Measurement/Storage), HP 54655A (Test Assistant)

Auxiliary Inputs

ParameterValue
Z-Axis Input Impedance10 kΩ
Z-Axis Input LevelTTL compatible (active low blanks display)

X-Y Mode

ParameterValue
X-Y Mode BandwidthX-axis: DC to 1.8 MHz, Y-axis: DC to 150 MHz
X-Y Phase Difference±3° at 100 kHz

Vertical System

ParameterValue
Vertical Offset Range (Channels 1 & 2)±2 V (2 mV/div to 200 mV/div), ±40 V (500 mV/div to 5 V/div)
Channel-to-Channel Isolation≥ 40 dB at 100 MHz

Maintenance

ParameterValue
Recommended Calibration Interval1 year

Technology

Digital Storage Oscilloscope (DSO)

Applications

Electronic testing, laboratory measurement, and academic education