The Agilent/Keysight 54846A is a high-performance 4-channel digital oscilloscope from the Infiniium 54800 Series. It features an analog bandwidth of 2.25 GHz and real-time sample rates of up to 8 GSa/s, designed to capture high-speed waveforms with precision. Equipped with a Windows-based graphical user interface, the 54846A combines the ease of use of an analog oscilloscope with the advanced analysis capabilities of a digital instrument, including options for voice control, USB test suites, and remote web control.

Specifications

General

ParameterValue
InterfacesGPIB, LAN (10/100Base-T), RS-232, Parallel (Centronics), VGA out, PS/2 mouse and keyboard
Display10.4-inch active-matrix color TFT LCD, 640 × 480 resolution
Internal StorageInternal Hard Disk Drive (HDD) and 3.5-inch 1.44 MB floppy drive
Operating SystemWindows 98 or Windows XP Professional (model/upgrade dependent)
System ProcessorIntel Celeron or Pentium III (model/upgrade dependent)
Chassis CoolingForced-air (fan)
Recommended Calibration Interval1 year

Physical

ParameterValue
Power Supply100 V to 240 V AC (±10%), 47 Hz to 440 Hz, 390 W maximum power consumption
Operating Temperature5 to 40
Dimensions (W×H×D)432 × 216 × 442
Weight14.1

Vertical System

ParameterValue
Analog Bandwidth (-3 dB)2.25 GHz
Calculated Rise Time (10% to 90%)156 ps
Number of Channels4
Input Impedance50 Ω
Input CouplingDC
Vertical Resolution8 bits (up to 12 bits with averaging)
Input Sensitivity Range2 mV/div to 1 V/div
Maximum Input Voltage5 Vrms (or ±5 V peak)
Offset Range±1.2 V (2 to 50 mV/div), ±4 V (51 to 250 mV/div), ±5 V (251 mV/div to 1 V/div)
DC Gain Accuracy±1.25% of full scale
Channel-to-Channel Isolation≥ 40 dB (DC to 1 GHz), ≥ 30 dB (1 GHz to 2 GHz), ≥ 20 dB (2 GHz to 2.25 GHz)

Horizontal System

ParameterValue
Timebase Range50 ps/div to 20 s/div
Timebase Accuracy±(0.15 ppm + 0.15 ppm/year aging)

Acquisition System

ParameterValue
Maximum Real-Time Sample Rate8 GSa/s (on 2 channels), 4 GSa/s (on 4 channels)
Standard Memory Depth64 kpts/channel (128 kpts on 2 channels)
Maximum Memory Depth1 Mpts/channel (with Option 040)
Maximum Equivalent Time Sample Rate250 GSa/s
Memory Upgrade OptionsOption 040: 1 Mpts/channel (2 Mpts on 2 channels)

Trigger System

ParameterValue
Trigger SourcesChannel 1, Channel 2, Channel 3, Channel 4, Auxiliary, Line
Trigger ModesAuto, Triggered, Single
Trigger CouplingDC, AC, Low Frequency Reject, High Frequency Reject, Noise Reject
Trigger Holdoff Range100 ns to 10 s
Internal Trigger Level Range±8 divisions from screen center
Auxiliary Trigger BandwidthDC to 100 MHz

Waveform Measurements and Analysis

ParameterValue
Automatic MeasurementsVoltage (Max, Min, P-P, Top, Base, Amp, Avg, RMS, Over/Preshoot), Time (Period, Freq, Width, Duty, Rise/Fall, Phase)
Waveform Math OperatorsAdd, Subtract, Multiply, Divide, FFT, Differentiate, Integrate
FFT Window TypesHanning, Hamming, Flat Top, Rectangular
Limit / Mask TestingOptional (with Communications Mask Test Kit)

Display and User Interface

ParameterValue
Waveform Display FormatsYT, XY
Voice Control OptionAvailable (for hands-free operation)

Environmental and Compliance

ParameterValue
Storage Temperature Range-40 °C to +70 °C
Operating HumidityUp to 95% RH (non-condensing) at +40 °C
Operating AltitudeUp to 4,600 m (15,000 ft)

Connectivity and Interfaces

ParameterValue
Logic Analyzer CompatibilityTime correlation with Agilent 16700 series logic analyzers

Display

ParameterValue
Display Size8.4 in (213 mm) diagonal
Display Resolution640 x 480 pixels (VGA)
Display TypeColor flat-panel TFT active-matrix LCD

Inputs & Outputs

ParameterValue
Probe InterfaceAutoProbe interface (supports active probes with automatic scaling and power)

Compliance & Certifications

ParameterValue
EMC ComplianceCISPR 11 / EN 55011 Class A, IEC 61000-3-2, IEC 61000-3-3
Safety StandardsIEC 61010-1 / EN 61010-1, CSA C22.2 No. 1010.1, UL 3111-1

Technology

Digital Phosphor / DSO

Applications

High-Speed Signal Integrity, Communications Test, USB Compliance, Debugging