The Keysight Agilent 8114A is a high-power, programmable single-channel pulse generator designed for testing components that operate at elevated voltages or currents, such as laser diodes, infrared (IR) diodes, EEPROMs, and flash memories. It delivers fast, clean positive or negative pulses up to 100 V and 2 A at frequencies up to 15 MHz, with a highly configurable architecture that features extensive device-under-test (DUT) protection capabilities.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488.2)

Timing Characteristics

ParameterValue
Frequency Range1 Hz to 15 MHz
Period Range66.7 ns to 999 ms
Period Resolution3 digits (min 100 ps)
Pulse Width Range10 ns to 999 ms
Pulse Width Resolution3 digits (min 100 ps)
Pulse Delay Resolution3 digits (min 100 ps)
Double Pulse Delay Resolution3 digits (min 100 ps)

Output Characteristics

ParameterValue
Amplitude Range (into 50 Ω)1.00 V to 50.0 V (50 Ω source) / 1.00 V to 100 V (Low-Impedance source)
Amplitude Range (into Open Circuit)2.00 V to 100 V (50 Ω source)
Amplitude Resolution3 digits
Baseline Offset Range±25 V (with Option 001 baseline shift)
Rise Time (10% to 90%)< 12 ns
Fall Time (90% to 10%)< 12 ns
Overshoot / Preshoot / Ringing< 5% of amplitude ±100 mV
Maximum Output Current2 A
Output ProtectionUser-definable limits for voltage, current, and duty cycle
Output Connector TypeBNC

Triggering and Synchronization

ParameterValue
Trigger ModesContinuous, Triggered, Gated, Burst
External Trigger Input Frequency Range1 Hz to 15 MHz
External Trigger Input Impedance50 Ω or 10 kΩ
Sync Output LevelTTL level into 50 Ω

Operating Modes and Patterns

ParameterValue
Pulse ModesSingle pulse, Double pulse

Control and Programming

ParameterValue
GPIB InterfaceStandard, IEEE-488.2 compliant
Command Set LanguageSCPI

Output Performance

ParameterValue
Maximum Duty Cycle50% (amplitude <= 50 V); 15% (amplitude > 50 V, standard); 10% (amplitude > 50 V, Option 001)

Output Parameters

ParameterValue
Source Impedance OptionsApprox. 2 Ω (standard) / 50 Ω (with Option 002)
Maximum Amplitude Options100 V (standard) / 50 V (with Option 002) into 50 Ω

Applications

High-voltage and high-current device testing, including laser/IR diodes, EEPROMs, and flash memories