The Keysight (formerly Agilent HP) 83485A is a high-performance plug-in module featuring a 20 GHz optical channel and a 20 GHz electrical channel. Compatible with the Agilent 83480A and 86100A/B/C mainframes, it delivers high-speed waveform characterization and SONET/SDH transceiver compliance testing utilizing a switchable 4th-order Bessel-Thomson reference receiver filter.

Specifications

General

ParameterValue
InterfacesMainframe backplane interface (supports HP-IB/GPIB control via host mainframe)
DisplayNone (relies on host mainframe display)

Physical

ParameterValue
Power SupplyPowered by host mainframe
Dimensions (W×H×D)124 mm × 76 mm × 279 mm
Weight1.2 kg

System Compatibility

ParameterValue
Compatible MainframesAgilent 83480A, 86100A, 86100B, 86100C

Electrical Channel Specifications

ParameterValue
Electrical Channel Bandwidth12.4 GHz and 20 GHz (selectable)
Electrical Transition Time (12.4 GHz)28.2 ps
Electrical Transition Time (20 GHz)17.5 ps
Electrical Maximum RMS Noise (12.4 GHz)0.25 mV
Electrical Maximum RMS Noise (20 GHz)0.5 mV
Electrical Scale Factor Range1 mV/div minimum to 100 mV/div maximum (8 divisions)
Electrical DC Offset Range±500 mV
Electrical Nominal Input Impedance50 Ω
Electrical Dynamic Range±400 mV relative to channel offset
Electrical Maximum Safe Input±2 V + peak AC (+16 dBm)
Electrical Connector Type3.5 mm female

Optical Channel Specifications

ParameterValue
Optical Maximum RMS Noise (20 GHz)25 µW
Optical Scale Factor Range20 µW/div minimum to 500 µW/div maximum (8 divisions)
Optical DC Offset Range+1 to -3 mW
Optical Average Power Monitor Range-30 dBm to +3 dBm (1 µW to 2 mW)
Optical Dynamic Range2 mW
Optical Maximum Safe Input10 mW peak
Calibrated Wavelengths1310 nm and 1550 nm
Wavelength Range1200 - 1600 nm
Optical Fiber Input Type9/125 µm single-mode
Optical Connector TypeHP universal adapter

Reference Receiver Features

ParameterValue
Reference Receiver Standards ComplianceITU-TS (formerly CCITT) G.957 and Bellcore GR-253-CORE
Bessel-Thomson Filter Type4th-order
Filter Switch MechanismInternal repeatable HP microwave switch (button or HP-IB controlled)

Auxiliary Interfaces

ParameterValue
Probe Power OutputOne probe power connector

Optical Channel

ParameterValue
Photodetector TypeInGaAs
Optical Return Loss33 dB minimum
Reference Receiver Filter Rate2.488 Gb/s (OC-48/STM-16)

Environmental

ParameterValue
Storage Temperature-40 °C to +70 °C
Operating HumidityUp to 95% RH at 40 °C (non-condensing)
Operating AltitudeUp to 4,600 m (15,000 ft)

Channel Configuration

ParameterValue
Number of Channels2 (1 optical, 1 electrical)

Clock Recovery

ParameterValue
Clock Recovery Rates (Option 011)155 Mb/s, 622 Mb/s, 2488 Mb/s

Technology

Optical and Electrical Sampling Oscilloscope Module

Applications

High-speed digital transmitter characterization, transceiver compliance testing, and SONET/SDH testing