The Keysight (formerly Agilent / HP) 83650A is a high-performance synthesized sweeper providing broadband frequency coverage from 10 MHz to 50 GHz. It combines the precise accuracy of a synthesized source with the speed and versatility of a sweep oscillator. Utilizing indirect synthesis, the 83650A achieves excellent harmonic performance and low single-sideband phase noise. It is designed for demanding applications in component test, receiver test, frequency conversion, and antenna scan pattern simulation.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | HP-IB (IEEE-488) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 / 120 / 220 / 240 VAC, 48-66 Hz (400 Hz at 100/120V); 400 VA max |
| Operating Temperature | 0 to 55 |
| Dimensions (W×H×D) | 425 × 178 × 648 |
| Weight | 27.2 |
Frequency Characteristics
| Parameter | Value |
|---|---|
| Frequency Range | 10 MHz to 50 GHz |
| Frequency Resolution | 1 kHz (standard); 1 Hz (Option 008) |
Sweep Characteristics
| Parameter | Value |
|---|---|
| Sweep Modes | Step, List, Ramp |
Output Power Characteristics
| Parameter | Value |
|---|---|
| Maximum Leveled Output Power (< 26.5 GHz) | +10 dBm |
| Maximum Leveled Output Power (26.5 to 40 GHz) | +5 dBm |
| Maximum Leveled Output Power (> 40 GHz) | +2.5 dBm |
| Power Sweep Range | -20 dBm to maximum power |
| Output Impedance | 50 Ω |
Spectral Purity
| Parameter | Value |
|---|---|
| Harmonics (2.0 to 26.5 GHz) | < -50 dBc |
| Harmonics (> 26.5 GHz) | < -40 dBc |
| Non-Harmonic Spurious Signals (> 2.0 GHz) | < -50 dBc |
Connectivity and Interfaces
| Parameter | Value |
|---|---|
| RF Output Connector | 2.4 mm (male) |
Common Options and Upgrades
| Parameter | Value |
|---|---|
| 1 Hz Frequency Resolution Option | Option 008 |
Sweep Performance
| Parameter | Value |
|---|---|
| Analog Sweep Time | 10 ms to 100 s |
Timebase & Reference
| Parameter | Value |
|---|---|
| External Reference Input Frequencies | 1, 2, 5, or 10 MHz |
| Internal Reference Output | 10 MHz (nominal 0 dBm) |
Power
| Parameter | Value |
|---|---|
| Power Consumption | 400 VA maximum |
| AC Power Input | 100/120/220/240 VAC, 50/60 Hz (400 Hz for 100/120 V) |
Environmental
| Parameter | Value |
|---|---|
| Storage Temperature | -40 °C to +75 °C |
Frequency
| Parameter | Value |
|---|---|
| Timebase Aging Rate | 5 × 10^-10 / day (after 72 hours warm-up) |
Sweep Capabilities
| Parameter | Value |
|---|---|
| Step Sweep Points | 2 to 801 points |
Leveling
| Parameter | Value |
|---|---|
| External Leveling Inputs | ALC (Detector), Power Meter |
Output Power
| Parameter | Value |
|---|---|
| Step Attenuator Range (Option 001) | 55 dB |
| Step Attenuator Resolution (Option 001) | 5 dB steps |
Technology
Indirect Synthesis
Applications