The Keysight (Agilent/HP) 83751B is a high-power synthesized sweeper that delivers outstanding microwave performance for general-purpose, swept test, and scalar applications. Operating across the 2.0 to 20.0 GHz range, the 83751B combines the accuracy and stability of a synthesized source with the fast sweep speed of a traditional analog sweeper.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488)
DisplayVacuum Fluorescent Display (VFD)

Physical

ParameterValue
Power Supply90 to 132 Vac (47 to 440 Hz) or 198 to 264 Vac (47 to 66 Hz), < 250 VA
Operating Temperature0 to 55 °C
Dimensions (W×H×D)425 mm × 133 mm × 503 mm
Weight16.0

Frequency Specifications

ParameterValue
Frequency Range2.0 to 20.0 GHz
Frequency Resolution (Standard)100 kHz
Frequency Resolution (Option 1E8)1 Hz
Internal Reference Aging Rate (Standard)< 2 × 10^-6 per year
Internal Reference Aging Rate (Option 1E5)< 5 × 10^-10 per day
Internal Reference Frequency10 MHz
External Reference Input Frequency10 MHz (± 5 ppm)
External Reference Input Level-3 to +20 dBm

RF Output Specifications

ParameterValue
Maximum Leveled Power+17 dBm
Minimum Leveled Power (Standard)-15 dBm
Minimum Leveled Power (Option 1E1)-85 dBm
Power Resolution0.01 dB
Power Level Accuracy±1.5 dB (typically ±0.7 dB)
Power Level Flatness±0.6 dB
Output Impedance50 ohms
Output SWR (Source Match)< 1.6:1

Spectral Purity

ParameterValue
Harmonics< -30 dBc (at +10 dBm)
Non-harmonics Spurious< -50 dBc
SSB Phase Noise (10 GHz, 10 kHz offset)< -75 dBc/Hz
SSB Phase Noise (10 GHz, 100 kHz offset)< -85 dBc/Hz

Modulation Capabilities

ParameterValue
Pulse Modulation On/Off Ratio> 80 dB
Pulse Rise/Fall Time< 15 ns
Minimum Pulse Width< 50 ns

Sweep Modes & Triggering

ParameterValue
Sweep ModesContinuous analog sweep, Stepped sweep, Power sweep, CW
Analog Sweep Time Range10 ms to 99 s
Step Sweep Time Range1 ms to 99 s per step
Sweep Trigger SourcesExternal, Internal, Single, Key, Line
Number of Frequency Markers5

Inputs & Outputs (Connectivity)

ParameterValue
RF Output Connector3.5 mm (male)
Sweep Output (Ramp Out)BNC (female), 0 to 10 V
Z-Axis Blanking OutputBNC (female)
External Trigger InputBNC (female)
External Pulse InputBNC (female)
External AM InputBNC (female)
External FM InputBNC (female)
10 MHz Reference InputBNC (female)
10 MHz Reference OutputBNC (female)

Remote Programming & Interface

ParameterValue
Programming CompatibilitySCPI, Agilent/HP 8350B Compatibility Mode

Physical & Environmental

ParameterValue
Cooling DesignForced air (fan)
Power Consumption< 250 VA max
Rackmount WidthStandard 19-inch rack (3U height)

Modulation

ParameterValue
Linear AM Sensitivity100% / V
Log AM Sensitivity10 dB / V
AM 3 dB BandwidthDC to >100 kHz
FM 3 dB Bandwidth (AC-Coupled)10 kHz to 8 MHz
FM 3 dB Bandwidth (DC-Coupled)DC to 250 kHz
Maximum FM Deviation±10 MHz
FM Sensitivity100 kHz/V, 1 MHz/V, or 10 MHz/V (selectable)
Pulse Input LevelTTL compatible
Internal Pulse Generator Period (Option 1E2)1 µs to 419 ms
Internal Pulse Generator Width (Option 1E2)100 ns to 419 ms
Internal Pulse Generator Delay (Option 1E2)100 ns to 419 ms
Internal Pulse Generator Resolution (Option 1E2)100 ns

Environmental

ParameterValue
Storage Temperature Range-40 to +75 °C
Operating Humidity RangeUp to 95% relative humidity at +40 °C
Operating AltitudeUp to 4,570 m (15,000 ft)
Non-Operating AltitudeUp to 15,240 m (50,000 ft)

Power

ParameterValue
AC Input Voltage90 - 132 VAC / 198 - 264 VAC (automatic selection)
AC Input Frequency47 - 440 Hz (90 - 132 VAC) / 47 - 66 Hz (198 - 264 VAC)
Power Sweep Range (Standard)> 20 dB
Power Sweep Resolution0.01 dB

Inputs & Outputs

ParameterValue
RF Output Connector (Option 1ED)Type-N (female)

Frequency

ParameterValue
CW Switching Time (Standard)< 70 ms
CW Switching Time (Option 1E4)< 15 ms

Sweep

ParameterValue
Dwell Time Range1 ms to 419 s
Dwell Time Resolution1 ms

Technology

Synthesized Analog Sweeper

Applications