The Agilent / Keysight 85027C is a high-performance directional bridge designed for high-precision reflection measurements over a wide frequency range up to 18 GHz. It is engineered to operate seamlessly with HP/Agilent Scalar Network Analyzers, offering outstanding directivity and test port match for accurate microwave measurement applications.

Specifications

General

ParameterValue
InterfacesCustom multi-pin analog interface cable for Scalar Network Analyzer connection

Physical

ParameterValue
Power SupplyPowered directly by the connected Scalar Network Analyzer
Operating Temperature0 to 55 °C
Dimensions (W×H×D)124 mm × 26 mm × 147 mm
Weight0.5 kg
Integrated Cable Length1.5 m
Shipping Weight1.0 kg

Frequency and Impedance

ParameterValue
Frequency Range10 MHz - 18 GHz
Nominal Impedance50 Ω
Test Port Impedance50 Ω

RF Performance

ParameterValue
Minimum Directivity (12.4 to 18 GHz)≥ 34 dB
Minimum Directivity (0.01 to 12.4 GHz)≥ 36 dB
Typical Insertion Loss (at 18 GHz)8.0 dB
Test Port Match (0.01 to 8.4 GHz)SWR ≤ 1.15 (Equivalent Source Match)
Test Port Match (8.4 to 12.4 GHz)SWR ≤ 1.25 (Equivalent Source Match)
Test Port Match (12.4 to 18 GHz)SWR ≤ 1.43 (Equivalent Source Match)

Power and Electrical Ratings

ParameterValue
Maximum Continuous RF Input Power+23 dBm
Maximum DC Voltage (RF Port)±10 V

Connector and Port Configurations

ParameterValue
RF Input Connector TypeType-N
RF Input Connector GenderFemale
Test Port Connector TypeType-N
Test Port Connector GenderFemale

Compatibility

ParameterValue
Compatible Scalar Network Analyzers (AC & DC modes)Agilent 8757
Compatible Scalar Network Analyzers (AC mode only)Agilent 8756, Agilent 8755

Environmental

ParameterValue
Storage Temperature-40 to +75 °C

Performance

ParameterValue
Internal Detector TypeSchottky diode
Minimum Directivity (0.01 to 8.4 GHz)40 dB
Minimum Directivity (8.4 to 12.4 GHz)36 dB
Insertion Loss (Input to Test Port) at 10 MHz6.5 dB (nominal)
Insertion Loss (Input to Detector) at 10 MHz6.5 dB (nominal)
Insertion Loss (Input to Detector) at 18 GHz8.0 dB (nominal)
Variation with Frequency (Input to Test Port, 0.01 to 8.4 GHz)±0.8 dB
Variation with Frequency (Input to Test Port, 0.01 to 12.4 GHz)±1.1 dB
Variation with Frequency (Input to Test Port, 0.01 to 18 GHz)±1.5 dB
Variation with Frequency (Input to Detector, 0.01 to 8.4 GHz)±1.2 dB
Variation with Frequency (Input to Detector, 0.01 to 12.4 GHz)±1.5 dB
Variation with Frequency (Input to Detector, 0.01 to 18 GHz)±2.0 dB

Options

ParameterValue
Option 005 Test Port Connector3.5 mm (male)

Technology

RF and Microwave Directional Bridge

Applications

Reflection and transmission measurements with scalar network analyzers