The Keysight (Agilent HP) 85033D is a precision 3.5 mm mechanical calibration kit designed for calibrating network analyzers from DC up to 6 GHz. It contains high-accuracy fixed loads, open circuits, and short circuits in both male and female genders, supporting measurements of devices with 3.5 mm and SMA connectors.

Specifications

Physical

ParameterValue
Operating Temperature15 °C to 35 °C

General Specifications

ParameterValue
Frequency RangeDC to 6 GHz
Nominal Impedance50 Ω
Connector Type3.5 mm
Compatible Connector TypeSMA
Compatible Vector Network AnalyzersAgilent/Keysight network analyzers up to 6 GHz

Kit Standards & Contents

ParameterValue
Male Open StandardIncluded
Female Open StandardIncluded
Male Short StandardIncluded
Female Short StandardIncluded
Male Fixed Load StandardIncluded
Female Fixed Load StandardIncluded

Accessories & Ordering Information

ParameterValue
Option 100 DescriptionAdds phase-matched 7 mm to 3.5 mm adapters (male and female) for use with 7 mm test port cables
Option 200 DescriptionAdds four Type-N to 3.5 mm adapters

Standard-Specific Electrical Characteristics

ParameterValue
Broadband Fixed Load Return Loss≥ 46 dB, ≥ 44 dB, or ≥ 38 dB (frequency dependent)

Environmental & Calibration Standards

ParameterValue
Calibration Reference Temperature Range20 °C to 26 °C
Storage Temperature Range-40 °C to 75 °C

Standard Definitions

ParameterValue
Male Open Offset Delay29.243 ps
Female Open Offset Delay29.243 ps
Male Short Offset Delay31.785 ps
Female Short Offset Delay31.785 ps
Standard Offset Impedance50 Ω

Environmental

ParameterValue
Operating Altitude< 4,500 m
Non-Operating Altitude< 15,240 m
Relative HumidityUp to 95% @ 40 °C (non-condensing)

Performance

ParameterValue
Fixed Load SWR (DC to 2 GHz)≤ 1.011
Fixed Load SWR (2 to 6 GHz)≤ 1.025
Fixed Load SWR (6 to 9 GHz)≤ 1.040
Fixed Load Return Loss (DC to 2 GHz)≥ 46 dB
Fixed Load Return Loss (2 to 6 GHz)≥ 38 dB
Fixed Load Return Loss (6 to 9 GHz)≥ 34 dB

Technology

Coaxial Calibration Standard

Applications

RF & Microwave Vector Network Analyzer Calibration