The Keysight Agilent HP 8560A is a high-performance portable spectrum analyzer designed for RF and microwave measurement applications. Operating over a frequency range of 50 Hz to 2.9 GHz, it combines lab-grade performance with a ruggedized, portable form factor. Its exceptional phase noise, excellent dynamic range, and selective resolution bandwidths down to 10 Hz make it ideal for demanding field and benchtop environments.

Specifications

Frequency Specifications

ParameterValue
Frequency Range50 Hz to 2.9 GHz
Minimum Span100 Hz
Maximum Span2.9 GHz
Frequency Span Range0 Hz, 100 Hz to 2.9 GHz

Bandwidth Specifications

ParameterValue
Resolution Bandwidth (RBW) Steps1, 3, 10 sequence
Video Bandwidth (VBW) Steps1, 3, 10 sequence

Amplitude Specifications

ParameterValue
Input Attenuator Range0 to 70 dB
Input Attenuator Steps10 dB
Maximum Safe RF Input Level (Continuous Wave)+30 dBm (1 W) with ≥ 10 dB attenuation
Maximum Safe RF Input Level (DC Voltage)0 V (DC coupled)

Sweep and Triggering

ParameterValue
Minimum Sweep Time (Span > 0 Hz)50 ms
Maximum Sweep Time (Span > 0 Hz)100 s
Sweep Time Range (Zero Span)50 µs to 100 s

Inputs and Outputs

ParameterValue
RF Input Connector TypeType N female
RF Input Impedance50 ohms
Calibrator Output300 MHz, -10 dBm
External Reference Input Frequency10 MHz
IF Output ConnectorBNC female
IF Output Frequency310.7 MHz
Video Output ConnectorBNC female
External Reference Input ConnectorBNC female
External Reference Output ConnectorBNC female
Sweep Output ConnectorBNC female
1st LO Output ConnectorSMA female
Earphone / Demodulator OutputInternal speaker and front panel 1/8-inch (3.5 mm) jack

User Interface and Trace Functions

ParameterValue
Number of Traces3
Marker FunctionsMarker to Peak, Delta Marker, Marker to Center

Connectivity and Control

ParameterValue
Direct Print/Plot CompatibilityHP-GL compatible printers and plotters via GPIB

Power Requirements

ParameterValue
AC Input Frequency Range47 to 440 Hz (115V), 47 to 66 Hz (230V)

Physical Characteristics

ParameterValue
Chassis Type / Form FactorPortable, ruggedized
Handle and Bumper ConfigurationTilt-handle, front and rear protective bumpers

Factory Options and Accessories

ParameterValue
Tracking Generator OptionOption 002 (300 kHz to 2.9 GHz)

Environmental

ParameterValue
Environmental ComplianceMeets MIL-T-28800C, Type III, Class 3, Style C
Maximum Operating Altitude4,570 meters (15,000 feet)
Maximum Non-Operating Altitude15,240 meters (50,000 feet)
Environmental ProfileMIL-T-28800C, Type III, Class 3, Style C

Amplitude

ParameterValue
Ref Level Calibration Uncertainty±0.3 dB at 300 MHz, -20 dBm
Residual Responses< -90 dBm (10 MHz to 2.9 GHz, 0 dB attenuation)
Image, Multiple, and Out-of-Band Responses< -70 dBc (10 MHz to 2.9 GHz)

Tracking Generator (Option 002)

ParameterValue
Tracking Generator Output Frequency Range300 kHz to 2.9 GHz (Option 002)

Options

ParameterValue
Option 0012nd IF Output (310.7 MHz)

Technology

Analog and digital RF spectrum analysis

Applications

RF transmitter testing, spurious signal search, phase noise measurement, and electromagnetic compatibility diagnostics