The Keysight (Agilent/HP) 8566B is a legendary high-performance spectrum analyzer spanning a frequency range from 100 Hz to 22 GHz, with extension capabilities up to 325 GHz via external mixers. Renowned for its exceptional frequency stability, synthesizer accuracy, and high sensitivity, the 8566B is designed for both benchtop research and automated test equipment (ATE) system applications.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | HP-IB (GPIB) standard |
| Display | Built-in CRT (part of HP 85662A IF-Display section) with digital vector graphics |
| System Components | HP 85660B RF Section and HP 85662A IF-Display Section |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 / 120 / 220 / 240 VAC (+11% / -10%), 48 to 66 Hz (Option 400 adds 400 Hz), 650 VA maximum |
| Operating Temperature | 0 to 55 °C |
Frequency Specifications
| Parameter | Value |
|---|---|
| Frequency Range (Internal Mixer) | 100 Hz to 22 GHz (preselected 2.0 GHz to 22 GHz) |
| Frequency Reference Aging Rate | < 1 × 10^-9 per day (after 30 days of continuous operation) |
| Frequency Span Range | 0 Hz (zero span), 100 Hz to 22 GHz |
Bandwidth (RBW/VBW)
| Parameter | Value |
|---|---|
| Resolution Bandwidth (RBW) Range | 10 Hz to 3 MHz |
| Resolution Bandwidth (RBW) Step Resolution | 1, 3, 10 sequence |
| Video Bandwidth (VBW) Step Resolution | 1, 3, 10 sequence |
Amplitude and Noise Specifications
| Parameter | Value |
|---|---|
| Amplitude Measurement Range | -134 dBm to +30 dBm |
| Maximum Safe Input Level (Continuous Wave AC) | +30 dBm (1 Watt) with ≥ 10 dB input attenuation |
| Maximum Safe Input Level (DC Voltage) | 0 VDC (DC coupled input) |
| Displayed Average Noise Level (DANL) @ 1 GHz | < -134 dBm (10 Hz RBW, 0 dB attenuation) |
| Input Attenuator Range | 0 dB to 70 dB |
| Input Attenuator Step Size | 10 dB |
| 1 dB Compression Point (100 Hz to 5.8 GHz) | -10 dBm at input mixer |
| 1 dB Compression Point (2.0 GHz to 22 GHz) | -5 dBm at input mixer |
| Second Harmonic Distortion (2.0 GHz to 22 GHz) | < -100 dBc (for -10 dBm mixer level, preselected) |
| Input SWR (VSWR) 5.8 GHz to 22 GHz | < 1.9:1 (with ≥ 10 dB attenuation) |
| Scale Fidelity (Linear) | ±3% of reference level |
Sweep and Triggering
| Parameter | Value |
|---|---|
| Sweep Time Range (Span > 0 Hz) | 20 ms to 1500 s |
Inputs and Outputs (Front and Rear Panel)
| Parameter | Value |
|---|---|
| RF Input Impedance | 50 Ω |
| Calibrator Output Amplitude and Frequency | -10 dBm, 100 MHz |
| External Mixer Port (IF Input) | SMA female, 321.4 MHz |
| 1st LO Output Port | SMA female, 2 to 6.2 GHz |
| IF Output Frequency and Impedance | 21.4 MHz (50 Ω) and 321.4 MHz (50 Ω) |
| Display Outputs | X, Y, Z BNC outputs (rear panel) for external monitors |
Power and Physical Specifications
| Parameter | Value |
|---|---|
| Chassis Form Factor | System configuration comprises 85660B RF Section and 85662A IF-Display Section |
Factory Options and Configurations
| Parameter | Value |
|---|---|
| Option 400 | Power Line Frequency Adaptation (allows 400 Hz operation) |
| Option 908 | Rackmount Flange Kit |
Frequency
| Parameter | Value |
|---|---|
| Internal YIG-Tuned Preselector Range | 2.0 GHz to 22.0 GHz |
| 1st LO Frequency Range | 2.0 GHz to 6.2 GHz |
Interfaces
| Parameter | Value |
|---|---|
| Direct Plot Capability | Supports direct hardcopy output to HP-IB plotters (e.g., HP 7470A, HP 7475A) without a computer |
Options
| Parameter | Value |
|---|---|
| Option 462 Description | Provides 6 dB resolution bandwidths (9 kHz and 120 kHz) for EMI measurements |
Performance
| Parameter | Value |
|---|---|
| LO Emission at RF Input | <-80 dBm (100 Hz to 2.5 GHz); <-10 dBm (2.0 to 22 GHz) with 0 dB RF attenuation |
| Maximum Safe Pulsed RF Input | +50 dBm (100 W) for pulse width < 10 µs and duty cycle < 1% (with ≥10 dB RF attenuation) |
| Image and Multiple Spurious Responses | <-70 dBc (2.0 to 18.6 GHz); <-60 dBc (18.6 to 22.0 GHz) |
Display
| Parameter | Value |
|---|---|
| On-Screen Dynamic Range | 90 dB |
| Log Scale Display Range | 1, 2, 5, 10 dB/division |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Probe Power | +15 Vdc, -12.6 Vdc; 150 mA max |
Technology
Superheterodyne Spectrum Analyzer with YIG-tuned Preselection
Applications
RF and Microwave Spectrum Analysis, Electromagnetic Compatibility (EMC) Testing, Broadband Signal Surveillance