The Keysight (Agilent HP) 8592A is a portable microwave spectrum analyzer operating in the 50 kHz to 22 GHz frequency range. It features a built-in preselector and comb generator, a built-in frequency counter, and options for remote computer control, making it a robust and versatile solution for legacy RF and microwave testing.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | Optional HP-IB (GPIB), RS-232, or HP-IL interfaces |
| Display | Monochrome CRT |
| Memory Capacity | Up to 50 state and trace registers (non-volatile) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 86 to 127 V AC / 195 to 253 V AC, 47 to 66 Hz, < 150 VA |
| Operating Temperature | 0 °C to +55 °C |
| Dimensions (W×H×D) | 366 mm × 213 mm × 460 mm |
| Weight | 20 kg |
Frequency Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 50 kHz to 22 GHz |
| Frequency Span Range | 0 Hz (zero span), 50 kHz to 22 GHz |
Bandwidth Specifications
| Parameter | Value |
|---|---|
| Resolution Bandwidth (RBW) Range | 1 kHz to 3 MHz |
| RBW Steps | 1, 3 sequence |
Amplitude Specifications
| Parameter | Value |
|---|---|
| Maximum Safe Input Level | +30 dBm (1 Watt) |
| Input Attenuator Range | 0 to 70 dB in 10 dB steps |
Inputs and Outputs
| Parameter | Value |
|---|---|
| RF Input Connector Type | Type-N female |
| RF Input Impedance | 50 Ω |
System and User Interface
| Parameter | Value |
|---|---|
| Internal Preselector | Built-in preselector and comb filters |
| Control Settings Storage | Store and recall of control settings and trace displays |
| Programmable Functions | Over 100 functions programmable via optional interfaces |
Sweep & Trigger
| Parameter | Value |
|---|---|
| Sweep Time Range (Span > 0) | 20 ms to 100 s |
| Sweep Time Range (Zero Span) | 20 µs to 100 s |
| Sweep Trigger Modes | Free Run, Single, Line, Video, External |
Bandwidth (RBW/VBW)
| Parameter | Value |
|---|---|
| Video Bandwidth (VBW) Range | 30 Hz to 3 MHz |
Internal Calibrator
| Parameter | Value |
|---|---|
| Calibrator Output Frequency | 300 MHz |
| Calibrator Output Amplitude | -20 dBm |
| Calibrator Output Connector | BNC female |
Amplitude Accuracy
| Parameter | Value |
|---|---|
| Linear Scale Fidelity | ±3% of reference level |
| Reference Level Range | -120 to +30 dBm |
Frequency Reference
| Parameter | Value |
|---|---|
| External Reference Input Frequency | 10 MHz |
Spurious Responses
| Parameter | Value |
|---|---|
| Third-Order Intermodulation Distortion (10 MHz to 22 GHz) | <-70 dBc (for two -30 dBm tones at input mixer) |
| LO Emission (with preselector) | <-75 dBm (at RF input) |
Environmental
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to +75 °C |
| Operating Altitude Limit | 4570 m |
| Operating Humidity Range | Up to 95% (non-condensing) |
Display
| Parameter | Value |
|---|---|
| Display Resolution (Horizontal × Vertical) | 401 × 256 points |
Inputs & Outputs
| Parameter | Value |
|---|---|
| External Trigger Input Level | TTL compatible |
RF Performance
| Parameter | Value |
|---|---|
| Preselector Tuning Range | 2.75 GHz to 22 GHz |
Technology
Superheterodyne Spectrum Analysis
Applications
RF and Microwave Spectrum Analysis