The Keysight (Agilent/HP) 8592L is a portable microwave spectrum analyzer designed to offer frequency-accurate RF and microwave measurements. Covering a wide frequency range up to 22 GHz, it features an easy-to-use interface with over 200 built-in functions, dual-interface options for remote control and printing, and dedicated measurement routines.

Specifications

General

ParameterValue
InterfacesOptional dual interfaces: Option 041 (GPIB and Parallel) or Option 043 (RS-232 and Parallel)
DisplayMonochrome CRT

Physical

ParameterValue
Power Supply86 to 127 V AC / 195 to 250 V AC, 47 to 66 Hz, < 150 VA
Operating Temperature0 to 55 °C
Dimensions (W×H×D)373 mm × 188 mm × 461 mm (with handle, feet, and bumpers)
Weight15.6 kg

Frequency Specifications

ParameterValue
Frequency Range9 kHz to 22 GHz
Frequency Span Range0 Hz (zero span), 50 kHz to 22 GHz
Frequency Reference Aging Rate±2 x 10^-6 / year
Frequency Reference Temperature Stability±5 x 10^-6
Sweep Time Range (Span > 0 Hz)20 ms to 100 s
Sweep Time Range (Zero Span)20 µs to 100 s
Sweep Trigger SourcesFree Run, Single, Line, Video, External

Amplitude Specifications

ParameterValue
Measurement Range-112 dBm to +30 dBm
Maximum Safe RF Input Power+30 dBm (1 Watt)
Maximum Safe DC Input Voltage0 V DC
Displayed Average Noise Level (DANL)-112 dBm @ 1 GHz (with 1 kHz RBW)
Input Attenuator Range0 to 70 dB
Input Attenuator Steps10 dB
Reference Level Range-120 dBm to +30 dBm
Log Display Scale1, 2, 5, 10 dB/division
Linear Display ScaleLinear scale (V, W, A)

Resolution and Video Bandwidths

ParameterValue
Resolution Bandwidth (RBW) Range1 kHz to 3 MHz
Resolution Bandwidth Steps1, 3, 10 sequence
Video Bandwidth (VBW) Range30 Hz to 1 MHz
Video Bandwidth Steps1, 3, 10 sequence

Front Panel Connections

ParameterValue
RF Input Connector TypeType-N female (50 ohm, standard)
RF Input Impedance50 ohms
Calibrator Output Power Level-20 dBm
Calibrator Output Frequency300 MHz
Calibrator Connector TypeBNC female

Rear Panel Connections

ParameterValue
Auxiliary Video Output0 to 1 V
Auxiliary IF Output21.4 MHz, BNC female
Sweep Output0 to 10 V ramp, BNC female
10 MHz Reference InputBNC female
10 MHz Reference OutputBNC female

Measurement and Display Functions

ParameterValue
Trace ModesClear/Write, Max Hold, Min Hold, View, Blank
Number of Measurement Traces3 (Trace A, Trace B, Trace C)
Trace Detection ModesPeak, Sample, Negative Peak
Marker FunctionsNormal, Delta, Marker to CF, Marker to Ref Level, Marker Noise, Counter
One-Button Measurement RoutinesAdjacent Channel Power (ACP), Signal Bandwidth, Third Order Intercept (TOI)
Internal Save/Recall RegistersUp to 228 states and traces

Power and Physical Specifications

ParameterValue
Dimensions (without handle/feet)325 mm W × 163 mm H × 427 mm D

Environmental and Compliance

ParameterValue
Storage Temperature Range-40 to +75 °C
Humidity Limits95% RH @ 40 °C
Altitude Limits4,572 m (operating)

Power Requirements

ParameterValue
Power Consumption< 150 VA (without options)

Performance

ParameterValue
Microwave Input TuningUnpreselected

Display

ParameterValue
Display Graticule8 x 10 division

Compliance

ParameterValue
Electromagnetic CompatibilityConforms to CISPR Publication 11 and EN 55011 Class A
Safety StandardsDesigned in accordance with IEC 348 and UL 1244

Dynamic Range

ParameterValue
Third-Order Intermodulation Distortion<-70 dBc (for two -30 dBm tones at the input mixer, >10 MHz)
Second-Harmonic Distortion (10 MHz to 2.9 GHz)<-70 dBc (for -30 dBm tone at the input mixer)
1 dB Gain Compression>-10 dBm (for mixer level, >10 MHz)
Residual Responses (150 kHz to 6.4 GHz)<-90 dBm (0 dB RF attenuation, terminated input)
Residual Responses (6.4 to 22 GHz)<-80 dBm (0 dB RF attenuation, terminated input)

Frequency

ParameterValue
Preselector Tuning RangeNone (unpreselected)
Span Accuracy (Span <= 10 MHz)±2% of span

Amplitude Accuracy

ParameterValue
Frequency Response (9 kHz to 2.9 GHz)±1.5 dB (referred to 300 MHz cal signal)
Linear Scale Fidelity±3% of Reference Level

Options

ParameterValue
Option 1D5 Frequency Reference Aging Rate<±1 x 10^-7 per year
Option 130 Resolution Bandwidth Range30 Hz to 3 MHz
Option 041 InterfaceGPIB and Parallel Printer Interface
Option 043 InterfaceRS-232 and Parallel Printer Interface

Technology

Superheterodyne Spectrum Analyzer

Applications

Microwave and RF Signal Analysis, Spectral Purity Testing, General-Purpose Laboratory Measurement