The Keysight (Agilent) 86105C is a plug-in module for the Infiniium 86100A/B/C DCA and DCA-J sampling oscilloscope mainframes. It features a 9 GHz amplified optical channel and a 20 GHz electrical channel, offering extensive optical filter coverage for SONET/SDH and datacom/enterprise technologies up to 11.3 Gb/s in a single module.
Specifications
Physical
| Parameter | Value |
|---|---|
| Operating Temperature | 10 to +40 °C |
| Module Slots Required | 1 standard DCA module slot |
General Module Specifications
| Parameter | Value |
|---|---|
| Module Type | Optical / Electrical Plug-in Module |
| Number of Optical Channels | 1 |
| Number of Electrical Channels | 1 |
| Supported Data Rates | Up to 11.3 Gb/s (option dependent) |
Optical Channel Specifications
| Parameter | Value |
|---|---|
| Wavelength Range | 750 nm to 1600 nm |
| Optical Bandwidth (Unfiltered) | 9 GHz |
| Fiber Type Compatibility | 9/125 µm to 62.5/125 µm |
| Reference Receiver Capability | Data rates depend on installed filter options |
Electrical Channel Specifications
| Parameter | Value |
|---|---|
| Electrical Bandwidth | 12.4 GHz / 20 GHz (selectable) |
| Electrical Input Connector Type | 3.5 mm (male) |
| RMS Electrical Noise (12.4 GHz BW) | 0.25 mV |
| RMS Electrical Noise (20 GHz BW) | 0.5 mV |
| Maximum Input Signal | ± 2 Vdc |
Ordering Information and Options
| Parameter | Value |
|---|---|
| Option 100 | Combination of four low data rate filters |
| Option 200 | Multi 9G filters |
| Option 300 | Combination of low data rate filters and multi 9G filters |
| Option 110 | 155 Mb/s filter |
| Option 120 | 622 Mb/s filter |
| Option 130 | 1.063 Gb/s filter |
| Option 140 | 1.250 Gb/s filter |
| Option 150 | 2.125 Gb/s filter |
| Option 160 | 2.488 Gb/s and 2.500 Gb/s filter |
| Option 170 | 2.666 Gb/s filter |
| Option 180 | 3.125 Gb/s filter |
| Option 190 | 4.25 Gb/s filter |
| Option 193 | 5.0 Gb/s filter |
| Option 195 | 6.25 Gb/s filter |
| Option 197 | 8.5 Gb/s filter |
Optical Input
| Parameter | Value |
|---|---|
| Calibrated Wavelengths | 850 nm, 1310 nm, 1550 nm |
| Optical Connector | Keysight/Agilent Universal Optical Interface |
Electrical Specifications
| Parameter | Value |
|---|---|
| Nominal Electrical Input Impedance | 50 Ω |
| Electrical Transition Time (20 GHz Bandwidth) | < 17.5 ps (10% to 90%, typical) |
| Electrical Transition Time (12.4 GHz Bandwidth) | < 28.2 ps (10% to 90%, typical) |
Environmental
| Parameter | Value |
|---|---|
| Storage Temperature | -40 °C to +70 °C |
Technology
Sampling Oscilloscope Plug-in
Applications
SONET/SDH and datacom/enterprise transmitter compliance and waveform characterization up to 11.3 Gb/s