The Keysight (Agilent HP) 8713C is an RF Economy Network Analyzer designed for transmission and reflection measurements over a frequency range of 300 kHz to 3 GHz. Featuring a synthesized source with 1 Hz resolution, 100 dB of system dynamic range, and a fast sweep speed of 50 ms/sweep, the 8713C is optimized for high-throughput manufacturing and testing of RF components. It includes an integrated 3.5-inch disk drive, a built-in monochrome CRT display, VGA output, and supports HP's Instrument BASIC (IBASIC) for automated testing. Optional features include 75-ohm port impedance, fault location (SRL), AM delay measurements, and TCP/IP LAN connectivity.

Specifications

General

ParameterValue
InterfacesGPIB, RS-232, Parallel (printer), VGA, optional LAN
Display9-inch monochrome CRT
Warm-up Time30 minutes

Physical

ParameterValue
Power Supply90 to 132 V AC / 198 to 264 V AC, 47 to 66 Hz
Operating Temperature0 to 55 °C
Dimensions (W×H×D)425 mm × 177 mm × 508 mm
Weight20.5

Source Specifications

ParameterValue
Frequency Range300 kHz to 3 GHz
Frequency Resolution1 Hz
Frequency Accuracy±5 ppm
Output Power Range-60 to +10 dBm (50 Ω standard), -60 to +7 dBm (75 Ω option)
Output Power Resolution0.1 dB
Output Power Accuracy±1.0 dB
Test Port Impedance50 Ω (standard) / 75 Ω (Option 1EC)

Receiver Specifications

ParameterValue
Receiver Frequency Range300 kHz to 3 GHz
System Dynamic Range (Narrowband)100 dB
System Dynamic Range (Broadband)66 dB
Maximum Input Level+10 dBm

Measurement Characteristics

ParameterValue
Measurement FormatsLog magnitude, linear magnitude, SWR
Sweep TypesLinear frequency, list frequency, power sweep
Number of Display Channels2 channels
Sweep Time50 ms/sweep
Number of Measurement Points per Sweep3 to 1601
Internal Programming LanguageAgilent Instrument BASIC (IBASIC)

Interfaces & Connectivity

ParameterValue
Test Port ConnectorsType-N female
GPIB InterfaceIEEE-488.2 (HP-IB)
Parallel Port25-pin D-sub (Centronics) for printer
Serial Port9-pin RS-232
VGA Video Output Port15-pin D-sub (color output)
LAN Interface OptionRJ-45 (TCP/IP compliant)

System Features & Storage

ParameterValue
Internal Disk Drive3.5-inch floppy drive (LIF/DOS formats)
Limit Line TestingSupported

Power & Environmental Specifications

ParameterValue
Storage Temperature Range-40 to 70 °C

Physical Characteristics

ParameterValue
Form FactorBenchtop

Options & Accessories

ParameterValue
Option 1EC75 Ohm Port Impedance
Option 100Fault Location / SRL
Option 101AM Delay Measurement
Option 1F7LAN Interface
Option 1CMRackmount flange kit without handles
Option 1CPRackmount flange kit with handles

Receiver

ParameterValue
IF Bandwidths250 Hz, 3.5 kHz, 15 kHz

Source

ParameterValue
Harmonics<-30 dBc (at +10 dBm output power)
Non-Harmonic Spurious<-30 dBc

Markers

ParameterValue
Number of Markers8

Options

ParameterValue
Option 1E1 Attenuator Range0 to 60 dB in 10 dB steps
Option 1C2HP Instrument BASIC (IBASIC)

Rear Panel Connections

ParameterValue
Auxiliary InputBNC female, ±10 V range
External Trigger InputBNC female, TTL compatible
External Reference InputBNC female, 1, 2, 5, or 10 MHz, -2 to +10 dBm
Reference OutputBNC female, 10 MHz, +0 dBm typical
Aux I/O Port25-pin D-sub

Data Storage

ParameterValue
Internal Non-Volatile Memory Capacity512 KB

Power

ParameterValue
Maximum Power Consumption280 VA

Environmental

ParameterValue
Operating AltitudeUp to 4,500 m (15,000 ft)
Non-operating AltitudeUp to 15,240 m (50,000 ft)
Operating Humidity Range5% to 95% relative humidity (non-condensing)

Frequency Reference

ParameterValue
Standard Reference Frequency Temperature Stability±5 ppm (0 °C to 55 °C)
Standard Reference Frequency Aging Rate±5 ppm/year

Storage

ParameterValue
Disk Drive Storage Format3.5-inch MS-DOS (FAT16) or LIF

Connectivity

ParameterValue
HP-IB Command SetSCPI (Standard Commands for Programmable Instruments)

Group Delay

ParameterValue
Group Delay Aperture Range0.25% to 20% of frequency span

Technology

Vector Network Analyzer

Applications

RF component testing, filter design, transmission/reflection measurement